文献速递主动脉疾病

电磁术中定位系统(IOPS)作为血管内主动脉手术三维成像辅助的安全性与可行性研究

Electromagnetic Intraoperative Positioning System (IOPS) Is Safe and Effective as a 3D Imaging Adjunct in Endovascular Aortic Surgery-A Safety and Feasibility Study.
PubMed 文献 · Ann Vasc Surg · 2026-03-10

结果显示技术成功率100%,10天随访无归因于IOPS的严重或非严重不良事件。

要点

  • IOPS在TEVAR/FEVAR/EVAR中提供辅助三维导航,技术成功率达100%。
  • 10天随访无归因于IOPS的严重或非严重不良事件。
  • 80%病例使用单根传感器导丝,90%病例使用单根传感器导管。

核心结论

  • 电磁IOPS作为辅助三维引导在肾下及开窗腹主动脉瘤血管内修复中安全有效。
  • IOPS生成的三维解剖图在所有TEVAR/FEVAR/EVAR靶目标中实现准确主动脉导航。

研究设计与样本

  • 术前CTA生成三维解剖模型并与术中锥形束CT融合,术中IOPS辅助二维透视导航传感器导丝和导管。

主要发现

  • 患者平均年龄75岁(SD=8.2),80%为男性,83%为白人,66.7%有冠心病史,90%有高血压史。
  • IOPS辅助下主动脉导航及分支血管插管的技术成功率为100%。
  • 10天随访中归因于IOPS系统的严重不良事件和非严重不良事件均为零。
  • 80%的病例使用单根传感器导丝,90%的病例使用单根传感器导管。

关键数值与推荐

  • 技术成功定义为能够导航肾下、内脏及降胸主动脉并插管必要的主动脉分支血管,插管成功经透视确认。

临床怎么用

  • IOPS可作为TEVAR/FEVAR/EVAR中二维透视的辅助三维导航工具,用于导丝和导管操作及分支血管插管。

局限性

  • 需未来研究探讨IOPS降低辐射暴露、对比剂用量及手术时间的潜力。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Navigation of the aorta and its branch vessels using standard of care 2D X-ray fluoroscopy in thoracic, fenestrated, and infrarenal endovascular aortic aneurysm repair (TEVAR/FEVAR/EVAR) has multiple disadvantages including exposure to ionizing radiation, contrast agent administration, and limited visualization of complex anatomy. The objective of this study is to demonstrate the safety and efficacy of the electromagnetic-based intraoperative positioning system (IOPS) in providing accurate 3D vessel mapping and guidance for wire and catheter navigation as an adjunct to fluoroscopy during TEVAR/FEVAR/EVAR. METHODS: Thirty patients with aortic aneurysms suitable for TEVAR/FEVAR/EVAR were enrolled across 2 sites in the United States from 2020 to 2022. Prior to repair, patients underwent computed tomography angiography to generate a 3D anatomical model that was fused with intraoperative cone beam computed tomography. The intervention was performed utilizing IOPS as a 3D guidance system for sensorized wire and catheter manipulation as an adjunct to 2D fluoroscopy. Technical success was defined as the ability to navigate the infrarenal, visceral, and descending thoracic aorta as well as cannulate any necessary aortic branch vessels. Successful cannulation was confirmed using fluoroscopy. Patients were monitored for serious adverse events (SAEs) and nonserious adverse events (non-SAEs) attributable to IOPS intraoperatively and postoperatively. Demographic and patient history data were collected. RESULTS: Patients were predominantly White (83%) males (80%) with a history of coronary artery disease (66.7%) and hypertension (90%). The mean age was 75 years (standard deviation [SD] = 8.2). Interventions included EVAR, FEVAR, and TEVAR procedures. Physicians selected the most suitable graft for treatment, with grafts manufactured by Gore, Endologix, Terumo, Medtronic, and Cook. The fenestrated grafts utilized consisted of commercially available devices, custom made devices, and physician-modified endografts. Technical success was achieved in 100% of patients with IOPS providing adjunctive 3D guidance for wire and catheter manipulation during aortic navigation and branch vessel cannulation. There were zero SAEs and non-SAEs attributable to the IOPS system at the 10-day follow-up. A single sensorized guidewire was used in 80% of cases, and a single sensorized catheter was used in 90% of cases. CONCLUSION: The electromagnetic-based (IOPS is safe and effective in providing adjunctive 3D guidance for wire and catheter manipulation in infrarenal and fenestrated endovascular abdominal aortic aneurysm repair in humans. The 3D anatomical maps generated allowed for accurate aortic navigation in all targets during TEVAR/FEVAR/EVAR. Future research is needed to investigate the potential for IOPS to reduce radiation exposure to patients and operators, reduce contrast usage, and overall reduce operative time.

来源:Ann Vasc Surg,查看原文。版权归原作者所有。

引用

  1. Hoell NG, Erdely A, Goel V, et al. Electromagnetic Intraoperative Positioning System (IOPS) Is Safe and Effective as a 3D Imaging Adjunct in Endovascular Aortic Surgery-A Safety and Feasibility Study. Ann Vasc Surg. 2026. PMID: 41819359. doi:10.1016/j.avsg.2026.02.041

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。