Alto主动脉腔内修复装置对术后大动脉僵硬度的影响
该回顾性观察研究纳入116例腹主动脉或髂动脉瘤EVAR患者,比较Alto、ENDURANT II与EXCLUDER三种装置术后脉搏波速度变化。ENDURANT和EXCLUDER组术后PWV显著升高,Alto组无显著变化;多因素分析显示ΔmBP、缺血性心脏病史和外骨架设计独立关联ΔPWV,且装置类型与ΔmBP存在显著交互作用。
要点
- Alto装置术后PWV无显著升高,未明显增加大动脉僵硬度。
- ENDURANT与EXCLUDER组术后PWV显著升高。
- 外骨架设计、ΔmBP和缺血性心脏病史独立关联ΔPWV。
核心结论
- Alto装置与术后大动脉僵硬度增加无显著关联。
- 血压变化与PWV的关系因装置设计而异,提示EVAR后装置特异性血流动力学反应。
研究设计与样本
- 纳入单一装置平台治疗者,分为Alto组(n=28)、ENDURANT组(n=53)和EXCLUDER组(n=35),共116例。
- EVAR前后无创测量脉搏波速度,主要终点为术后PWV变化(ΔPWV)。
主要发现
- ENDURANT和EXCLUDER组术后PWV显著升高,Alto组术后PWV无显著变化。
- 多因素分析中,ΔmBP、缺血性心脏病史和外骨架设计独立关联ΔPWV。
- 装置类型与ΔmBP存在显著交互作用,提示不同装置设计下PWV对血压变化的反应不同。
临床怎么用
- 对于关注术后大动脉僵硬度增加的患者,可考虑选用无连续金属骨架的Alto装置。
- 术后血压变化对PWV的影响因装置设计而异,需结合装置类型解读PWV随访结果。
局限性
- 为回顾性观察研究,可能存在选择偏倚。
- 各组样本量不均衡,Alto组样本量较小。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Large artery stiffness (LAS) adversely affects multiple organs. Endovascular aneurysm repair (EVAR) is associated with a postoperative increase in LAS; however, LAS responses may differ according to stent graft design. Ovation Alto is a novel abdominal EVAR device characterized by a nonscaffolded main body without a continuous metallic framework. We evaluated the impact of the Alto device on postoperative LAS compared with that of representative exoskeleton EVAR devices. METHODS: In this retrospective observational study, consecutive patients who underwent EVAR for abdominal or iliac artery aneurysms between January 2017 and October 2025 were screened. We included patients treated with a single-device platform using the Alto, ENDURANT II, or EXCLUDER. Pulse wave velocity (PWV) was measured noninvasively before and after EVAR, and the postoperative change in PWV (ΔPWV) was the primary end point. RESULTS: In total, 116 patients were included (Alto, n = 28; ENDURANT, n = 53; and EXCLUDER, n = 35). PWV increased significantly after EVAR in patients treated with ENDURANT and EXCLUDER, whereas no significant postoperative changes in PWV were observed in the Alto group. In the multivariable analysis, change in mean blood pressure (ΔmBP), a history of ischemic heart disease, and exoskeleton design were independently associated with ΔPWV. In addition, a significant interaction between device type and ΔmBP was observed, indicating differential PWV responses to blood pressure changes according to device design. CONCLUSION: The Alto device was not significantly associated with an increase in LAS. In addition, the relationship between blood pressure changes and PWV differed by device design, suggesting a device-specific hemodynamic response after EVAR.
引用
- Hori Y, Sato K, Akiyoshi K, et al. Effect of the Alto Endovascular Aneurysm Repair Device on Postoperative Large Artery Stiffness. Ann Vasc Surg. 2026. PMID: 41796670. doi:10.1016/j.avsg.2026.02.031