尸体模型评估大型开窗医生改良Endo-Wheat装置的实验研究
本研究在尸体模型上评估医生改良开窗Endo-Wheat装置用于血管腔内主动脉根部修复的可行性与可重复性。结论认为该装置及部署技术可行且可重复。
要点
- 冠状动脉及头臂干均未被覆盖,装置修改平均耗时19分钟。
- 左腋动脉为合适入路,部署时操控性和精确性良好。
核心结论
- Endo-Wheat装置及其部署技术在尸体模型上用于血管腔内主动脉根部修复可行且可重复。
研究设计与样本
- 实验性研究,基于急性A型主动脉夹层患者CTA分析开发标准配置,在包含心脏、升主动脉、主动脉弓和部分降胸主动脉的尸体模型上于脉动台式血流模型完成操作。
主要发现
- 冠状动脉和头臂干均未被覆盖。
- 装置修改平均耗时19分钟。
- 左腋动脉为合适入路,部署时操控性和精确性良好。
- 宏观检查未见定制覆膜支架损坏。
临床怎么用
- 对于不适合开放手术的急性A型主动脉夹层患者,该医生改良开窗Endo-Wheat装置及部署技术可作为潜在血管腔内治疗选择。
局限性
- 研究仅基于尸体模型,非活体临床环境。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Acute type A dissection are treated with open surgery. When patients are not eligible, few therapeutic solutions remain. Endovascular approaches are currently under development. The objective of this experimental study was to evaluate the feasibility and reproducibility of a physician-modified fenestrated endo-wheat device for endovascular aortic root repair, and to refine both the device and the deployment technique. METHODS: The endo-wheat module consists of a thoracic stent graft and a nitinol valve. A standard configuration was developed based on computed tomography angiography analysis from patients with acute type A aortic dissection. The stent graft was modified to cover only one valve cusp and then sutured to the valve. Visibility was enhanced using one linear and 2 lateral radiopaque markers on the covered cusp. The cadaveric models included the heart, ascending aorta, aortic arch, and a portion of the descending thoracic aorta. All procedures were performed on a pulsatile benchtop flow model. Following each procedure, the ascending aorta was opened to assess the positioning and patency of the coronary ostia. RESULTS: Twelve procedures were completed. Successful deployment of the endo-wheat module into the aortic valve and ascending aorta was achieved in all cases. Neither the coronary arteries nor the brachiocephalic trunk were covered. The average time required to modify the module was 19 min. The left axillary artery emerged as a suitable access route, providing excellent maneuverability and precision during deployment. Macroscopic inspection revealed no damage to the customized stent grafts. CONCLUSION: The endo-wheat module and its associated deployment technique are feasible and reproducible for endovascular aortic root repair in a cadaveric model.
引用
- Bacri C, Durant B, Hireche K, et al. Experimental Evaluation of a Large Fenestrated Physician-Modified Endo-Wheat Using a Cadaveric Model. Ann Vasc Surg. 2026. PMID: 41747928. doi:10.1016/j.avsg.2026.02.017
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。