新型手术软件工具优化医师改良支架开窗流程
本研究基于38例复杂腹主动脉瘤患者的开窗数据,对比EndoDraft软件辅助与手工打孔卡设计医师改良支架开窗。软件组规划时间显著缩短(63.2±21.5秒 vs 233.0±40.3秒,p<0.001),且手工方法存在可测量的纵向与周向定位误差。软件辅助设计精度更高、速度大幅提升。
要点
- EndoDraft软件辅助打孔卡设计较手工方法显著缩短规划时间(约63秒 vs 约233秒)。
- 手工打孔卡开窗存在0.4–1.0 mm量级的纵向与周向定位误差。
- 软件学习曲线短,初期与后期完成时间无显著差异。
核心结论
- 软件辅助打孔卡设计在PMEG规划中较手工技术精度更高且速度大幅提升。
研究设计与样本
- 实验研究,基于2023年5月至2024年9月择期FEVAR治疗复杂腹主动脉瘤患者的38例开窗数据,制作76张打孔卡和288个开窗。
主要发现
- 手工组准备时间233.0±40.3秒,软件辅助组63.2±21.5秒(p<0.001)。
- 手工打孔卡纵向误差:腹腔干0.8±0.6 mm,肠系膜上动脉1.0±0.6 mm,右肾动脉0.8±0.5 mm,左肾动脉1.0±0.6 mm。
- 手工打孔卡周向误差:腹腔干0.4±0.4 mm,肠系膜上动脉0.4±0.4 mm,右肾动脉0.6±0.5 mm,左肾动脉0.5±0.4 mm。
- 软件学习曲线初期与后期完成时间无显著差异(77.8±29.4秒 vs 55.8±18.5秒,p=0.15)。
临床怎么用
- EndoDraft可替代手工打孔卡制图,生成约一分钟内完成的按比例缩放模板,减少手工定位误差。
- 该工具可提高规划步骤的可重复性,降低认知负荷和准备时间,易于标准化并缩短学习曲线。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVE: Three-dimensional models are increasingly used to facilitate the positioning of fenestrations on a physician-modified endograft (PMEG) during the fenestrated endovascular aortic repair (FEVAR) of complex abdominal aortic aneurysms (CAAA). The punch card technique was developed to eliminate the 3D printing workflow while preserving the benefits of having a 3D model. The objective of the current study is to evaluate the performance of the purpose-built software EndoDraft compared to manual punch card design. METHODS: An experimental study was performed to compare software-assisted and manual punch card creation. Fenestration data of readily available procedural plans were used for patients who underwent elective FEVAR for CAAA from May 2023 to September 2024. Time needed to complete punch card planning was recorded as well as inaccuracies of the manual technique were evaluated by measuring the vertical and circumferential position of the fenestrations on the punch card. RESULTS: A total of 76 punch cards and 288 fenestrations were made based on 38 patients' fenestration data. Preparation time was 233.0 ± 40.3 seconds for the manual group, whereas 63.2 ± 21.5 seconds for the software-aided group (p<0.001). Longitudinal imprecision of the manual punch card was 0.8 ± 0.6 mm for the celiac axis (CA), 1.0 ± 0.6 mm for the superior mesenteric artery (SMA), 0.8 ± 0.5 mm for the right renal artery (RRen), whereas 1.0 ± 0.6 mm for the left renal artery (LRen). Circumferential errors of the same fenestrations were 0.4 ± 0.4 mm for the CA, 0.4 ± 0.4 mm for the SMA, 0.6 ± 0.5 mm for the RRen, and 0.5 ± 0.4 mm for the LRen. No significant difference between the completion time recorded at the beginning and at the end of the software's learning curve was detected (77.8 ± 29.4 s vs 55.8 ± 18.5 s, p=0.15). CONCLUSIONS: The software-assisted design of the punch card is associated with higher precision and a vast improvement in speed compared to the manual technique during PMEG planning for the endovascular treatment of CAAAs. The purpose-built EndoDraft tool was made freely available in an online repository offering a more streamlined workflow.Clinical ImpactEndoDraft® could meaningfully streamline the PMEG-FEVAR workflow by replacing manual punch-card drafting with rapid, software-generated, proportionally scaled templates that are produced in about one minute rather than several minutes, while reducing the small but measurable positioning errors inherent to hand drawing. For clinicians, this means a more reproducible planning step, less cognitive load and "ruler time" during preparation, and a workflow that is easier to standardize across operators with a short learning curve. The key innovation is purpose-built automation with integrated 2D/virtual 3D visualization and direct PDF output, made freely available to support wider adoption.
引用
- Juhos B, Szentiványi A, Bérczi Á, et al. A Novel Surgical Software Tool to Improve the Physician-Modified Endograft Workflow. J Endovasc Ther. 2026. PMID: 41526085. doi:10.1177/15266028251406185