数字孪生技术预测胸主动脉腔内修复干预:概念验证
结果显示eOA%每环均低于10%,平均4%,近端和远端关键环平均低于3.8%,每次分析约10小时,准确性优于文献报道值。
要点
- TEVAR数字孪生可模拟患者特异性支架植入,每环eOA%均低于10%。
- 近端和远端支架环误差平均低于3.8%,对密封性评估关键。
- 仿真分析约10小时完成,准确性优于文献报道值。
核心结论
- TEVAR数字孪生经术后影像验证具有可信度,可用于预测患者特异性短期支架几何形态。
研究设计与样本
- 采用水平集技术分割患者特异性解剖,生成对应支架模型数字孪生,并行显式有限元数值模拟。
- 以支架环开口面积百分比误差(eOA%)为定量指标,比较模拟与术后影像重建的支架构型。
主要发现
- 近端和远端支架环误差尤其低,平均低于3.8%。
- 每次分析约10小时完成,准确性优于文献报道值。
关键数值与推荐
- eOA%每环均低于10%,平均4%。
- 近端和远端支架环eOA%平均低于3.8%。
- 每次仿真分析约10小时。
临床怎么用
- 数字孪生可支持患者特异性术前规划,优化手术策略。
- 通过预测支架术后几何形态,可能有助于减少并发症并改善TEVAR结局。
局限性
- 验证限于短期支架几何形态,未评估临床结局。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVE: Thoracic endovascular aortic repair (TEVAR) has become the preferred treatment for thoracic aortic pathologies, with current planning based on geometric assessment, although it neglects device-aorta interaction. This manuscript reports the use of a validated high fidelity finite element methodology to replicate the TEVAR procedure. The study aimed to establish the credibility of the TEVAR digital twin for clinical use, providing a reliable tool for pre-procedural planning in patient specific cases. METHODS: Pre- and post-operative computed tomography angiography (CTA) images of patients who underwent TEVAR using the Valiant Captivia stent graft were analysed retrospectively. To ensure the accuracy and reliability of the TEVAR digital twin, the simulation results were rigorously validated against post-operative image reconstructions. The patient specific anatomy was segmented using the level set technique from pre-operative CTA images, and a digital twin of the corresponding stent graft model was generated. Explicit finite element numerical simulations were conducted to replicate the stent graft implantation. Simulation validation analysis compared the simulated stent configuration with the stent reconstructed from post-operative images, using the stent ring opening area percentage error (eOA%) as a quantitative metric. RESULTS: Twenty patients were included. The simulations replicated stent graft implantation in typical patient specific anatomies, with eOA% consistently below 10% per stent ring, averaging 4%. Notably low errors were observed in the proximal and distal stent rings, critical for ensuring sealing (< 3.8% average). These analyses, completed in about 10 hours each, showed improved accuracy compared with values reported in the literature. CONCLUSION: The credibility of a TEVAR digital twin is essential for its trustworthiness and integration into clinical practice. The findings, achieved in a short time, validate the use of numerical simulations for predicting short term stent graft post-operative geometry in patient specific anatomies. By enabling personalised planning, digital twins could help optimise procedure strategies, reduce complications, and improve TEVAR outcomes.
引用
- Ramella A, De Campo G, Barati S, et al. Prediction of Thoracic Endovascular Aortic Repair Intervention: Proof of Concept using Digital Twin Technology. Eur J Vasc Endovasc Surg. 2026. PMID: 41520772. doi:10.1016/j.ejvs.2026.01.008
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。