近红外光谱与残端压力检测颈动脉内膜剥脱术中脑缺血的对比:前瞻性多中心观察研究
该前瞻性多中心观察研究在2家瑞典医院纳入293例局麻和66例全麻下颈动脉内膜剥脱术患者,比较近红外光谱(NIRS)与残端压力(SP)检测夹闭期脑缺血的价值。结果显示同侧rSO2相对变化≤-9%预测脑缺血的敏感度为91%、特异度为82%,阴性预测值99%;SP<50 mm Hg敏感度78%、特异度60%。结论认为NIRS可减少不必要的转流且敏感度高于SP,并能监测转流通畅性。
要点
- 同侧rSO2相对变化≤-9%检测夹闭期神经症状的阴性预测值为99%。
- ΔrSO2≤-9%预测脑缺血敏感度91%、特异度82%,优于SP<50 mm Hg的78%和60%。
- NIRS可减少不必要的转流,并能监测转流通畅性。
核心结论
- NIRS可减少不必要的转流,且检测脑缺血的敏感度高于残端压力,并能监测转流通畅性。
研究设计与样本
- 前瞻性多中心观察研究,纳入2家瑞典医院接受颈动脉内膜剥脱术的患者,其中局麻293例、全麻66例。
- 所有患者报告残端压力、不同时间点rSO2及术中神经症状,采用ROC分析确定SP和rSO2的最佳截断点。
主要发现
- 夹闭期出现神经症状者较未出现者ΔrSO2更低(-14%±6% vs. -4±5%,P<0.001),SP也更低(36±21 mm Hg vs. 56±19 mm Hg,P<0.001)。
- 同侧rSO2相对变化≤-9%检测夹闭期神经症状的阴性预测值为99%。
- ΔrSO2≤-9%预测脑缺血的敏感度为91%(95%CI:75-98%),特异度为82%(95%CI:77-87%)。
- SP<50 mm Hg预测脑缺血的敏感度为78%(95%CI:58-91%),特异度为60%(95%CI:54-67%)。
关键数值与推荐
- 同侧rSO2相对变化≤-9%作为检测夹闭期神经症状的阈值,阴性预测值99%。
- ΔrSO2≤-9%预测脑缺血的敏感度91%(95%CI:75-98%)、特异度82%(95%CI:77-87%)。
- SP<50 mm Hg预测脑缺血的敏感度78%(95%CI:58-91%)、特异度60%(95%CI:54-67%)。
临床怎么用
- 全麻下颈动脉内膜剥脱术中可考虑使用NIRS监测rSO2以辅助判断是否需要转流,减少不必要的转流。
- NIRS还可用于监测转流通畅性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Determining the need for shunt use during carotid endarterectomy (CEA) under general anesthesia (GA) can be based on different methods, and one of the most common is stump pressure (SP). Previous studies have found favorable results of near infrared spectroscopy (NIRS). METHODS: Patients undergoing CEA in 2 swedish hospitals were prospectively included. SP, rSO2 at different timepoints as well as neurological symptoms during surgery were reported for all patients. Receiver operating curve (ROC) analysis was used to determine the optimal cutoff points for SP and rSO2. RESULTS: A total of 293 patients underwent CEA under local anesthesia (LA) and 66 under GA in 2 hospitals. Thirty two patients operated under LA had neurological symptoms. A relative change in ipsilateral rSO2 ≤ -9% had a negative predictive value (NPV) of 99% in detecting neurological symptoms during clamping. Patients who developed neurological symptoms during clamping had lower ΔrSO2 (-14% ± 6% vs. -4 ± 5%, P < 0.001) and lower SP (36 ± 21 mm Hg vs. 56 ± 19 mm Hg, P < 0.001) compared to those who did not. A ΔrSO2 of ≤ -9% had a sensitivity of 91% (95% confidence interval (CI): 75-98%) and a specificity of 82% (95% CI: 77-87%) in predicting cerebral ischemia, whereas a SP < 50 mm Hg had a sensitivity of 78% (95% CI: 58-91%) and a specificity of 60% (95% CI: 54-67%). CONCLUSION: This study found that NIRS could reduce unnecessary shunting while still having a higher sensitivity compared to SP. Additionally, it can also monitor shunt patency.
引用
- Lindberg M, Gillgren P, Gunnarsson K, et al. Near Infrared Spectroscopy Versus Stump Pressure in Detecting Cerebral Ischemia During Carotid Endarterectomy: A Prospective Multicenter Observational Study. Ann Vasc Surg. 2025. PMID: 41205747. doi:10.1016/j.avsg.2025.10.016