标准化医师改良覆膜支架开窗与定制器械的台架力学性能对比
本研究在台架上比较标准化医师改良覆膜支架(sPMEG)开窗与定制器械(CMD)开窗的力学行为,评估支架释放和喇叭口扩张后开窗完整性及疲劳测试耐久性。二十个开窗(各十例)显示两者形态和结构完整性一致,8 mm sPMEG开窗扩张后直径和面积增加更大但不影响贴附,疲劳测试无结构缺陷。
要点
- sPMEG开窗在支架释放和扩张后保持圆形和结构完整。
- 8 mm sPMEG开窗扩张后直径和面积增加大于CMD,但不影响形态和贴附。
- 疲劳测试后两组开窗均无结构缺陷。
核心结论
- sPMEG开窗在支架释放和喇叭口扩张后表现与CMD开窗相似,并确保长期完整性。
- 台架测试支持sPMEG开窗在紧急复杂动脉瘤修复中的潜在临床应用。
研究设计与样本
- 体外台架研究,比较sPMEG与CMD开窗的力学性能。
- sPMEG通过电灼孔制作开窗,并用0.018「导丝和5.0锁定缝线加固。
主要发现
- 显微镜评估显示所有开窗在支架释放和扩张前后均保持圆形及织物、缝线和加固环的结构完整性。
- 与CMD相比,6 mm开窗扩张后直径和表面积的中位增加无统计学意义(p = .41)。
- 8 mm sPMEG开窗扩张后直径和表面积增加大于CMD(p = .046),但未影响形态或开窗与支架贴附。
- 疲劳测试后两组均未发现结构缺陷。
关键数值与推荐
- 6 mm开窗扩张后直径和表面积中位增加与CMD比较p = .41。
- 8 mm sPMEG开窗扩张后直径和表面积增加与CMD比较p = .046。
临床怎么用
- 台架结果提示sPMEG开窗可作为紧急复杂腹主动脉瘤修复中CMD的潜在替代。
- 临床应用中需注意8 mm sPMEG开窗扩张后尺寸变化较大,但形态和贴附仍可接受。
局限性
- 研究限于体外台架条件,结论需临床验证。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVE: This study aimed to evaluate the mechanical behaviour of a standardised fenestrated physician modified endograft (sPMEG) compared with a custom made device (CMD). The structural integrity of standardised fenestrations following bridging stent deployment and flaring as well as durability after fatigue tests were assessed. METHODS: The sPMEG was prepared by fenestrating a Valiant Captivia stent graft system (Medtronic) by creating an electrocautery hole, reinforced with a 0.018" guidewire secured by a 5.0 locking suture. Mechanical performance was assessed with microscopic and radiographic imaging and compared with a CMD fenestration sheet (Cook Zenith Fenestrated, Cook Medical, Bloomington, IL, USA). Integrity was evaluated by measuring fenestration diameter, area, and circularity changes before and after stent deployment and flaring. Additionally, fabric integrity, reinforcement ring stability, suture line integrity, and fenestration to stent apposition were compared. For fatigue testing, the sPMEG was deployed in a three dimensional printed tubular model under fluoroscopic guidance, followed by deployment and flaring of VBX stents (W.L. Gore & Associates, Inc. Newark, DE, USA). The sPMEG fenestrations were exposed to 7.3 million cycles using a controlled benchtop machine. RESULTS: Twenty fenestrations were evaluated (ten sPMEG and ten CMD). In all cases, microscopic evaluation showed a consistent circular shape and structural integrity of the fabric, suture line, and reinforcement ring both before and after stenting and flaring. Compared with the CMD, the median increase in fenestration diameter and surface area after stent flaring was not statistically significant for the 6 mm fenestration (p = .41) but was greater for the 8 mm sPMEG (p = .046) without compromising shape or fenestration to stent apposition. Fatigue testing revealed no structural defects in either group. CONCLUSION: SPMEG fenestrations perform similarly to CMD fenestrations after stenting and flaring, while also ensuring long term integrity. Although limited by in vitro conditions, these on bench tests support the potential clinical application of sPMEG fenestration in urgent complex aneurysm repair.
引用
- Piazza M, Torsello G, Spertino A, et al. Fenestrations of a Standardised Physician Modified Endograft for the Treatment of Complex Abdominal Aortic Aneurysm Demonstrate Similar on Bench Mechanical Performance Compared with a Manufactured Custom Made Device. Eur J Vasc Endovasc Surg. 2025. PMID: 41033639. doi:10.1016/j.ejvs.2025.09.057
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。