文献速递主动脉疾病颈动脉与弓上血管

新型微孔覆膜支架重建猪主动脉弓上分支并隔绝胸主动脉瘤的安全性与有效性

Safety and Efficacy of a New Microporous Stent-Graft for Reconstructing Porcine Supra-Aortic Arch Branches and Excluding Thoracic Aortic Aneurysm.
PubMed 文献 · Ann Vasc Surg · 2025-09-30

该研究在6头三元大白猪降主动脉建立动脉瘤模型并置入微孔覆膜支架隔绝动脉瘤,另在6头猪主动脉弓置入支架重建弓上分支。术后DSA示动脉瘤均未显影,支架通畅,无内漏、断裂或移位,右颈内动脉收缩期峰值流速平均下降20%(P<0.0001)。

要点

  • 微孔覆膜支架可改变血流动力学使主动脉瘤完全血栓化。
  • 微孔可在安全期内维持主动脉弓上分支的血供。
  • 扩大微孔可快速重建弓上分支,单支与双支重建时间分别为4.5±0.5 min和11.5±1.9 min。

核心结论

  • 微孔覆膜支架通过改变血流动力学使主动脉瘤完全血栓化,微孔可在安全期内维持弓上分支血供,扩大微孔可快速重建弓上分支且操作简便。
  • 随访期间12头猪均无感染、跛行、截瘫、脑梗死或死亡,DSA示支架均通畅,无内漏或支架断裂/移位。

研究设计与样本

  • 动物实验研究:6头三元大白猪降主动脉建立动脉瘤模型并置入微孔覆膜支架隔绝动脉瘤。
  • 另6头三元大白猪于主动脉弓置入微孔覆膜支架重建弓上分支,记录重建时间并用彩色多普勒超声测量右颈内动脉收缩期峰值流速。

主要发现

  • 置入微孔覆膜支架后DSA即刻见动脉瘤淡化,术后6-8天随访DSA示动脉瘤未显影。
  • 右颈内动脉收缩期峰值流速支架置入前为81.5±7.3 cm/sec,置入后为65.1±7.6 cm/sec,平均下降20%(P<0.0001)。
  • 单支与双支弓上分支重建所需时间分别为4.5±0.5 min和11.5±1.9 min。

临床怎么用

  • 该技术或可用于主动脉弓部病变的腔内治疗中隔绝动脉瘤并重建弓上分支血供。
  • 微孔设计在维持分支血供的同时可能促进瘤腔内血栓形成,扩大微孔有助于加快分支重建。

局限性

  • 研究为猪动物实验,样本量小(共12头),随访时间短(DSA随访6-8天),结论外推至人体需进一步验证。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: To evaluate the safety and efficacy of a new microporous stent-graft for reconstructing porcine supra-aortic arch branches and excluding porcine thoracic aortic aneurysm. METHODS: Aneurysm models were established in the descending aorta of 6 ternary large white pigs; simultaneously, microporous stent-grafts were placed to exclude the aneurysms. Digital subtraction angiography (DSA) was performed 7 ± 1 days postoperatively to assess the aneurysms. Microporous stent-grafts were placed in the aortic arch, and the supra-aortic arch branches were reconstructed in an additional 6 ternary large white pigs. The time required to reconstruct the supra-aortic arch branches was recorded. The peak systolic velocity in the right internal carotid artery was measured before and after stent placement using color Doppler ultrasound. RESULTS: The aneurysms were noted to fade on DSA immediately after placement of microporous stent-grafts. Follow-up DSA at 6-8 days postoperatively showed no visualization of the aneurysms. The peak systolic velocity in the right internal carotid artery was 81.5 ± 7.3 cm/sec before stenting and 65.1 ± 7.6 cm/sec after stenting, a mean decrease of 20% (P < 0.0001). The time required to reconstruct a single and double branch was 4.5 ± 0.5 min and 11.5 ± 1.9 min, respectively. During the follow-up period, all 12 pigs showed no infection, lameness, paraplegia, cerebral infarction, or death. DSA showed that all the stent-grafts were patent; furthermore, there was no endoleak or fracture/migration of the stents. CONCLUSION: The microporous stent-graft allows complete thrombosis of the aortic aneurysm by altering its hemodynamics, and the micropores maintain the blood supply to the supra-aortic arch branches for a safe period. Enlarging the micropores of the stent-graft allows rapid reconstruction of the supra-aortic arch branches; moreover, the procedure is easy to perform and the technology is easily accessible.

来源:Ann Vasc Surg,查看原文。版权归原作者所有。

引用

  1. Wan Q, Zhang X. Safety and Efficacy of a New Microporous Stent-Graft for Reconstructing Porcine Supra-Aortic Arch Branches and Excluding Thoracic Aortic Aneurysm. Ann Vasc Surg. 2025. PMID: 41038502. doi:10.1016/j.avsg.2025.09.041

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。