持续性血小板减少提示复杂腔内动脉瘤修复后内漏
本研究基于医师发起的IDE研究,评估开窗分支腔内修复(F-BEVAR)术后持续性血小板减少与高压内漏的相关性。185例患者中26例(14%)发生高压内漏,术后第4天平均血小板<100,000/μL者高压内漏风险显著升高(OR 12.3, 95%CI 3.5-43.0),此类患者中35%存在需再干预的高压内漏。
要点
- F-BEVAR术后第4天血小板<100,000/μL提示高压内漏风险增加。
- 高压内漏患者术后早期血小板计数显著低于低压内漏及无内漏者。
- 血小板<100,000/μL的患者中35%存在需再干预的高压内漏。
核心结论
- F-BEVAR术后发生高压内漏的患者术后血小板计数显著更低。
- 术后第4天血小板<100,000/μL应考虑主动脉影像学检查。
研究设计与样本
- 医师发起的IDE研究,评估医师改良和Cook制造的F-BEVAR患者。
- 共纳入185例患者,分析术后6个月内CT发现的内漏。
主要发现
- 185例中26例(14%)发生高压内漏,30例(16%)发生低压内漏。
- 术后第1至4天,高压内漏组血小板<100,000/μL的比例为42%,低压内漏组27%,无内漏组10%(p<0.001)。
- 调整人口学和合并症后,高压内漏组术后第4天平均血小板<100,000/μL的比值比为12.3(95%CI 3.5-43.0)。
- 术后第4天血小板<100,000/μL的患者中35%存在需再干预的高压内漏。
关键数值与推荐
- 每日平均血小板计数<100,000/μL作为切点。
- 高压内漏定义为I型或III型内漏,低压内漏定义为II型内漏。
临床怎么用
- F-BEVAR术后第4天血小板<100,000/μL时应考虑主动脉影像学检查以排查高压内漏。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVES: High-pressure endoleaks (type I and III) after fenestrated-branched endovascular aneurysm repair (F-BEVAR) pose ongoing rupture risk. Early identification of such endoleaks may facilitate appropriate reintervention. Our objective was to determine if persistent thrombocytopenia after F-BEVAR correlates with high-pressure endoleaks in the early postoperative period. METHODS: Patients in a physician-sponsored investigational device exemption study of physician-modified and Cook-manufactured fenestrated/branched endografts were evaluated. Endoleaks identified on computed tomography within the first 6 months of follow-up were classified as high-pressure (type I or III) or low-pressure (type II). Bivariable and multivariable analyses were performed to assess if thrombocytopenia was associated with the presence of endoleaks. RESULTS: Among 185 patients, 26 (14%) developed high-pressure endoleaks and 30 (16%) developed low-pressure endoleaks. Demographics, procedure indications, and preoperative medications were similar across endoleak groups. To maximize sensitivity and clinical utility, we selected a daily average platelet count <100,000/μL as a cut point. The proportion of high-pressure endoleak patients with platelet counts <100,000/μL was significantly greater compared to low-pressure and no endoleak patients on postoperative day 1 (POD#1) through POD#4 (high-pressure 42%, low-pressure 27%, no endoleak 10%; p<0.001). After adjusting for demographics and comorbidities, patients with high-pressure endoleak were over 12 times more likely to have average platelets <100,000/μL on POD#4 compared to patients with no endoleak (OR 12.3, 95% CI 3.5-43.0). CONCLUSIONS: F-BEVAR patients who developed high-pressure endoleak had significantly lower platelet counts postoperatively. Aortic imaging should be considered in patients with platelet counts <100,000/μL at POD#4 following F-BEVAR, as 35% of such patients will have a high-pressure endoleak requiring reintervention.Clinical ImpactPersistent thrombocytopenia following F-BEVAR may be an early indication of a high pressure endoleak. Imaging is recommended for patients with platelet count <100,000/μL on POD 4 as 35% of such patients in this study had a high pressure endoleak.
引用
- Long A, Zettervall SL, Sweet MP, et al. Persistent Thrombocytopenia Indicates Endoleak After Complex Endovascular Aneurysm Repair. J Endovasc Ther. 2025. PMID: 40583330. doi:10.1177/15266028251351146
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。