文献速递静脉血栓栓塞

人工智能引导的非专家加压超声诊断深静脉血栓路径可能减少血管实验室静脉检查

Novel Artificial Intelligence Guided Non-expert Compression Ultrasound Deep Vein Thrombosis Diagnostic Pathway May Reduce Vascular Laboratory Venous Testing.
PubMed 文献 · Eur J Vasc Endovasc Surg · 2025-05-14

本研究前瞻性纳入急诊疑似下肢深静脉血栓患者,由未经超声培训的操作者使用ThinkSono Guidance人工智能软件进行双区域近端加压超声检查,并由放射科医生远程判读。结果显示该路径对深静脉血栓诊断敏感性达100%,特异性96%,可能安全减少部分患者的静脉双功超声和D二聚体检测。

要点

  • 非超声培训操作者使用AI引导加压超声,所有扫描图像均达到诊断质量。
  • AI引导加压超声联合放射科判读诊断深静脉血栓敏感性100%,特异性96%。
  • 该路径可能使部分疑似深静脉血栓患者安全避免静脉双功超声和D二聚体检测。

核心结论

  • 本研究提示,使用ThinkSono系统可使相当比例疑似深静脉血栓患者安全避免静脉双功超声和D二聚体检测,为新型AI辅助诊断路径奠定基础。

主要发现

  • 所有扫描均达到诊断质量;AI引导加压超声联合放射科判读得出45例(85%)深静脉血栓阴性诊断。
  • 放射科医生怀疑8例患者存在深静脉血栓,双功超声确认6例,ThinkSono Guidance特异性96%,D二聚体特异性36%。
  • ThinkSono Guidance扫描和判读平均耗时6.75分钟,从扫描开始到判读的中位时间为37.5分钟。

临床怎么用

  • 在急诊疑似深静脉血栓患者中,可由未经超声培训的操作者使用AI引导加压超声进行检查,图像由放射科医生远程判读。
  • 判读为可压缩/无近端深静脉血栓且D二聚体阴性的患者可出院,其余患者转诊行静脉双功超声检查。

局限性

  • 原文未述局限性

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVE: Ultrasonography and D dimer testing are established modalities for evaluating potential lower extremity deep vein thrombosis (DVT). The ThinkSono Guidance system is an artificial intelligence (AI) based software allowing non-ultrasound trained providers to perform compression ultrasounds for evaluation by remote interpreters. This study evaluated its clinical use and potential reduction of venous duplexes and waiting times. METHODS: Patients with suspected DVT were prospectively recruited through the institution's emergency department. Patients underwent an AI guided two region proximal DVT compression examination by non-ultrasound trained providers using the ThinkSono Guidance system and D dimer testing. Ultrasound images remotely reviewed by the on call radiologist were rated for diagnostic quality; all images of sufficient quality were assessed as either "Compressible/no proximal DVT" or "Inadequate imaging/possible DVT". All patients assessed as "compressible" with negative D dimers were discharged. All other patients were sent for a venous duplex scan. Time to diagnosis, sensitivity, and specificity of ThinkSono Guidance against D dimers and full duplex scans were calculated. RESULTS: Fifty three patients (mean age 56 ± 18 years, 45% female) were scanned with ThinkSono Guidance by one of three non-ultrasound trained providers. All scans were of diagnostic quality. ThinkSono Guidance with radiologist review yielded 45 negative DVT diagnoses (85%). Seventeen of these with negative D dimers were discharged (32%) and 28 required duplex ultrasound testing per trial protocol (23 due to positive D dimers, five due to D dimer non-availability). All of these duplexes were negative (100% sensitivity). Eight patients were suspected of having DVT by the reviewing radiologist, and duplex confirmed DVT in six patients (96% ThinkSono Guidance specificity, 36% D dimer specificity). ThinkSono Guidance scans averaged 6.75 minutes for scan and review. The median time from scan initiation to review was 37.5 minutes. CONCLUSION: This study suggested that a significant proportion of patients with suspected DVT could safely avoid duplex ultrasound and D dimer testing using the ThinkSono system, setting the basis for a novel AI assisted diagnostic pathway.

来源:Eur J Vasc Endovasc Surg,查看原文。版权归原作者所有。

引用

  1. Avgerinos E, Spiliopoulos S, Psachoulia F, et al. Novel Artificial Intelligence Guided Non-expert Compression Ultrasound Deep Vein Thrombosis Diagnostic Pathway May Reduce Vascular Laboratory Venous Testing. Eur J Vasc Endovasc Surg. 2025. PMID: 40378910. doi:10.1016/j.ejvs.2025.04.070

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。