更宽的抓握:EndoAnchor在EVAR术中增加近端密封面
一项回顾性单中心研究(2015-2022年)纳入221例腹主动脉瘤患者,比较单用Endurant支架(190例)与联合Heli-FX EndoAnchor(31例)的EVAR术后实际达到的密封区(RASZ)。尽管EndoAnchor组近端密封区解剖更不利,其RASZ面积和长度仍显著增加;在24对匹配患者中结果一致。作者认为EndoAnchor可改善RASZ,常规评估RASZ可能优化治疗决策。
要点
- EndoAnchor组近端密封区更短(11 vs 17 mm)且肾下角更大(45° vs 18°)。
- EndoAnchor组RASZ面积增加48.3 mm²、长度增加5.8 mm,均优于单用Endurant组。
- 匹配的24对患者中,EndoAnchor组RASZ面积和长度仍显著优于单用组。
核心结论
- 与单用相同支架行EVAR相比,EndoAnchor显著改善RASZ长度和表面积。
- 作者认为常规进行器械特异性RASZ评估可能优化治疗算法,并为更个体化、持久的血管腔内解决方案铺路。
研究设计与样本
- 回顾性单中心研究(2015-2022年),221例腹主动脉瘤患者接受Endurant支架置入,单用组190例,联合Heli-FX EndoAnchor组31例。
- 术前和术后CT血管造影采用标准化测量技术(闭合多边形面积、中心腔线距离)量化目标预期密封区(TASZ)和RASZ。
- 主要终点为RASZ长度和面积相对于TASZ的变化;次要终点包括1A型内漏和任何内漏发生率、总体再干预。
- 统计分析包括未调整和调整的Cox比例风险回归模型、ROC曲线和倾向性评分匹配。
主要发现
- Heli-FX组近端密封区更不利:TASZ长度更短(11 vs 17 mm;P=0.003),肾下角更大(45° vs 18°;P<0.001)。
- 尽管解剖不利,Heli-FX组RASZ面积较标准EVAR组增加48.3 mm²(P=0.001),长度增加5.8 mm(P=0.003)。
- ROC分析显示各组预测1A型内漏的RASZ截断值不同:Heli-FX组为27.7 mm²和4.5 mm,Endurant组为49.5 mm²和9.5 mm。
- 在24对TASZ特征匹配的患者中,Heli-FX组RASZ面积增加42 mm²而单用组减少31 mm²(P=0.003),长度增加5 mm而单用组减少3 mm(P=0.003)。
关键数值与推荐
- Heli-FX组TASZ长度11 mm vs 单用组17 mm(P=0.003)。
- Heli-FX组肾下角45° vs 单用组18°(P<0.001)。
- Heli-FX组RASZ面积增加48.3 mm²(P=0.001),长度增加5.8 mm(P=0.003)。
- 匹配队列中Heli-FX组RASZ面积变化+42 mm² vs 单用组-31 mm²(P=0.003),长度变化+5 mm vs -3 mm(P=0.003)。
- 预测1A型内漏的RASZ截断值:Heli-FX组27.7 mm²和4.5 mm,Endurant组49.5 mm²和9.5 mm。
临床怎么用
- 作者认为常规进行器械特异性RASZ评估可能优化治疗算法,并为更个体化、持久的血管腔内解决方案铺路。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: To evaluate how the Heli-FX EndoAnchor system (Medtronic Vascular, Santa Rosa, CA, USA) influences the Real Achieved Sealing Zone (RASZ) compared to Endovascular Aneurysm Repair (EVAR) with Endurant (Medtronic Vascular, Santa Rosa, CA, USA) alone in a matched and unmatched real-world cohort. METHODS: In a retrospective, single-center study (2015-2022), 221 Abdominal Aortic Aneurysm patients underwent Endurant stent graft (Medtronic Vascular, Santa Rosa, CA, USA) placement either alone (n = 190) or with adjunctive Heli-FX EndoAnchors (n = 31). Preoperative and postoperative computed tomography angiography scans were evaluated using a standardized measurement technique (closed-polygon area, center lumen line distances) to quantify the Target Anticipated Sealing Zone (TASZ) and the RASZ. Primary endpoints were changes in RASZ length and area compared to TASZ. Secondary endpoints included type 1A and any endoleak rates, overall reinterventions. Statistical analysis included unadjusted and adjusted Cox proportional hazard regression models, receiver operating characteristic (ROC) curves, and propensity score matching. RESULTS: Heli-FX patients presented significantly more "hostile" proximal sealing zones, featuring shorter TASZ length (11 vs. 17 mm; P = 0.003), higher infrarenal angulation (45° vs. 18°; P < 0.001). Despite this hostile anatomy, the Heli-FX group exhibited a substantial increase in RASZ area (+48.3 mm2; P = 0.001) and length (+5.8 mm; P = 0.003) compared to standard EVAR patients. ROC analysis indicated distinct RASZ cutoffs for predicting type 1A endoleak in each group (27.7 mm2 and 4.5 mm for Heli-FX vs. 49.5 mm2 and 9.5 mm for Endurant). In the 24 TASZ-features-matched couples, Heli-FX patients showed increased RASZ area (+42 vs -31 mm2 over TASZ; P = 0.003), and length (+5 vs -3 mm over TASZ; P = 0.003). CONCLUSION: EndoAnchors substantially improve RASZ length and surface area compared to EVAR with the same stent graft alone. Routine device-specific RASZ assessment could refine treatment algorithms and pave the way for more personalized and durable endovascular solutions.