临时沟槽内漏减少急诊原位激光开窗修复内脏旁主动脉瘤时的肾内脏缺血:技术要点

Temporary Gutter Endoleak to Reduce Renovisceral Ischemia During Urgent In Situ Laser Fenestrated Paravisceral Aortic Aneurysm Repair: Technical Note.
PubMed 文献 · J Endovasc Ther · 2025-04-22

本文描述两种临时沟槽内漏策略,用于在急诊原位激光开窗(ISLF)修复内脏旁主动脉瘤(pAA)时确保肾内脏灌注。两例患者(79岁和83岁)均使用Cook Zenith Alpha覆膜支架,分别经肠系膜上动脉内的8F鞘或覆膜支架与主动脉壁之间的5 mm×200 mm球囊创建临时沟槽内漏;内脏动脉覆盖后造影证实经沟槽内漏灌注肾内脏动脉,随后完成ISLF和桥接支架置入,两例均无围手术期并发症。作者认为该技术可能减少肾内脏缺血并提高ISLF修复的安全性。

要点

  • 两例症状性pAA患者(79岁和83岁)接受ISLF修复,术中采用临时沟槽内漏维持肾内脏灌注。
  • 病例1经肠系膜上动脉内8F鞘、病例2经5 mm×200 mm球囊在覆膜支架与主动脉壁间创建沟槽内漏。
  • 内脏动脉覆盖后造影证实经沟槽内漏灌注肾内脏动脉,两例均无围手术期并发症。

核心结论

  • 作者认为,上述在急诊pAA的ISLF修复中创建沟槽内漏的技术可能减少肾内脏缺血并提高ISLF技术的安全性。

研究设计与样本

  • 技术要点,报告两例症状性内脏旁主动脉瘤患者(79岁和83岁)的急诊原位激光开窗修复。

主要发现

  • 两例均使用Cook Zenith Alpha覆膜支架,所有内脏血管预置支架作为引导标记。
  • 病例1在覆膜支架释放前将Cook Flexor 8F鞘置于肠系膜上动脉,病例2将5 mm×200 mm Armada球囊置于覆膜支架与主动脉壁之间。
  • 内脏动脉覆盖后造影证实经有意创建的沟槽内漏灌注肾内脏动脉。
  • 在移除鞘或球囊以停止沟槽内漏之前,完成肠系膜上动脉和肾动脉的ISLF及桥接支架置入。
  • 两例患者均未发生任何类型的围手术期并发症。

临床怎么用

  • 作者认为该沟槽内漏技术可能减少肾内脏缺血并提高ISLF修复的安全性。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

PURPOSE: Emergent repair of paravisceral aortic aneurysms (pAAs) with in situ laser fenestration (ISLF) technique is associated with renovisceral ischemia. We describe two strategies of temporary gutter endoleak creation to ensure renovisceral perfusion during ISLF. TECHNIQUE: Two patients (79- and 83-year-old) presented with symptomatic pAA. A Cook Zenith Alpha endograft (Cook Medical LLC, Bloomington, IN, USA) was used in both cases. All visceral vessels were prestented to serve as guide markers. Prior to the deployment of the endograft, a Cook Flexor 8F-sheath (Cook Medical LLC, Bloomington, IN, USA) was placed in the superior mesenteric artery (SMA) (patient #1), and a 5 mm × 200 mm angioplasty Armada balloon (Abbott, Green Oaks, Illinois, USA) was placed between the endograft and the aortic wall (patient #2). Angiograms after visceral coverage confirmed perfusion of the renovisceral arteries through the intentional gutter endoleaks. Thereafter, ISLF and bridging stenting for SMA and the renal arteries were performed before the removal of the sheath or balloon to stop the gutter endoleak. Both patients did not experience any kind of perioperative complications. CONCLUSIONS: The above-described techniques for gutter endoleak creation during emergent pAA repair with ISLF can potentially reduce reno visceral ischemia and increase the ISLF technique's safety.Clinical ImpactThe ISLF technique for pAA repair typically requires coverage of the visceral arteries with unpredictable ischemia time until laser fenestrations are established. The gutter endoleak technique reduces the renovisceral ischemia and potentially increases the safety of the ISLF repair.

来源:J Endovasc Ther,查看原文。版权归原作者所有。