使用逼真3D打印动脉瘤模型对开窗腔内动脉瘤修复新一代桥接支架的临床前评估
本研究在搏动泵模拟人体血流动力学条件下,用四个3D打印肾旁动脉瘤和对应开窗腔内移植物比较新型BeFlared桥接支架与标准BeGraft支架。共植入64枚支架,BeFlared在移植物内近端平均直径更大、张开角度更大、手术时间更短,两者疗效与技术性能相似。
要点
- BeFlared支架在移植物内近端平均直径大于BeGraft(10.5 mm vs. 9.2 mm,p < .001)。
- BeFlared张开角度大于BeGraft(46.1° vs. 33.4°,p < .001),手术时间更短(17.0 ± 4分钟 vs. 23.9 ± 4.7分钟,p = .004)。
- 医师评估未发现失败模式,两种器械技术性能均优秀。
核心结论
- 在采用3D打印患者解剖结构的创新测试装置中,BeFlared桥接目标血管的疗效和性能与BeGraft相似。
研究设计与样本
- 临床前比较评估,使用搏动泵模拟人体血流动力学,四个3D打印肾旁动脉瘤连接台式测试装置,对应四个开窗移植物(Zenith Fenestrated;Cook,Bloomington,IN,USA)。
- 桥接支架根据患者CT扫描分析选径,包括四枚标准覆膜支架(BeGraft [BG];Bentley,Hechingen,Germany)或四枚类似覆膜支架装载于改良球囊上以实现一步植入和张开(BeFlared [BF];Bentley)。
- 每个解剖模型和对应内移植物用于四组、每组四枚桥接支架植入。
主要发现
- 共成功植入64枚支架。
- 选择性血管造影显示三例IIIc型内漏(BG两例,BF一例)和七例Ic型内漏(BG四例,BF三例),均于术中纠正。
- 与BG相比,BF在移植物内近端平均直径更大(10.5 mm vs. 9.2 mm;p < .001),张开角度更大(46.1° vs. 33.4°;p < .001)。
- BF的八次手术比BG更快(17.0 ± 4分钟 vs. 23.9 ± 4.7分钟;p = .004)。
- 医师评估未发现任何失败模式,结论为两种器械技术性能均优秀。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVE: The aim of this study was to perform a comparative pre-clinical evaluation of a new fenestrated endovascular aneurysm repair dedicated stent using perfused 3D printed patient anatomies. METHODS: The test setup included a pulsatile pump set to reproduce human haemodynamics, four 3D printed pararenal aneurysms connected to a bench test, and four corresponding fenestrated grafts (Zenith Fenestrated; Cook, Bloomington, IN, USA). Bridging stents were sized based on analysis of patients computed tomography scans and included either four standard covered stents (BeGraft [BG]; Bentley, Hechingen, Germany) or four similar covered stent grafts mounted on a modified balloon to allow stent implantation and flaring in one step (BeFlared [BF]; Bentley). Each anatomy and corresponding endograft was used for four sets of four bridging stent implantations. Selective angiography following each stent implantation was used to assess efficacy (patency and absence of type Ic or IIIc endoleak), procedure time was recorded, and cone beam computed tomography at the end of each set of four stent implantations enabled comparison of geometric layout between stents. Physician assessment evaluated technical performance and failure modes. RESULTS: Sixty four stents were successfully implanted. Selective angiograms depicted three type IIIc (two in BG, one in BF) and seven type Ic (four in BG, three in BF) endoleaks, which were all corrected intra-operatively. Compared with BG, the BF had a larger mean proximal diameter inside the endograft (10.5 mm vs. 9.2 mm; p < .001) and greater opening flaring angle (46.1° vs. 33.4°; p < .001). The eight procedures performed with the BF were quicker than BG (17.0 ± 4 minutes vs. 23.9 ± 4.7 minutes; p = .004). Physician assessment did not reveal any failure mode and concluded an excellent technical performance of both devices. CONCLUSION: The BF demonstrated similar efficacy and performance compared with the BG for bridging target vessels in an innovative test setup using 3D printed patient anatomies.