文献速递主动脉疾病

使用人工智能(PRAEVAorta 2)对CT血管造影扫描中完全自动化的肾下最大主动脉瘤直径测量进行外部验证

External Validation of Fully-Automated Infrarenal Maximum Aortic Aneurysm Diameter Measurements in Computed Tomography Angiography Scans Using Artificial Intelligence (PRAEVAorta 2).
PubMed 文献 · J Endovasc Ther · 2024-11-13

本研究为多中心回顾性验证研究,基于ZEPHYR注册前瞻性收集的871例CT血管造影(来自347例接受EVAR的患者),比较PRAEVAorta 2全自动测量与国际公认核心实验室测量的肾下最大主动脉瘤直径。结果显示所有扫描的组内相关系数(ICC)为0.972(95%CI=0.968-0.976),95.4%的测量绝对差异在临床可接受的5 mm范围内,但10%的扫描存在明显分割错误。作者认为该人工智能软件可作为研究和临床实践的重要辅助工具,但需要经验丰富的观察者审查报告以识别错误分割。

要点

  • 全自动测量与核心实验室测量高度相关(r=0.97),ICC为0.972(95%CI=0.968-0.976)。
  • 95.4%的测量绝对差异在临床可接受的5 mm范围内。
  • 10%的扫描存在明显分割错误,需经验丰富的观察者审查报告。

核心结论

  • 在大多数病例中,PRAEVAorta 2可全自动准确测量最大腹主动脉瘤直径,表明人工智能软件可作为研究和临床实践的重要辅助工具。
  • 然而,需要经验丰富的观察者对生成的报告进行严格审查并谨慎使用,以识别错误的分割。

研究设计与样本

  • 多中心回顾性验证研究,使用ZEPHYR注册前瞻性收集的数据。
  • 数据集包括2016年至2019年间在德国、比利时和荷兰接受Zenith Alpha腔内覆膜支架(Cook Medical)行EVAR的347例患者的871次CT血管造影扫描。
  • 将PRAEVAorta 2(Nurea,法国波尔多)的自动测量与国际公认核心实验室(Syntactx,美国纽约州纽约市)的测量进行比较;核心实验室审查员为经过培训并使用既定测量标准的测量技术员,由血管外科医生和放射科医生监督。

主要发现

  • 所有扫描的测量相关性极好(r=0.97),组内相关系数(ICC)为0.972(95%CI=0.968-0.976)。
  • 总计95.4%的测量绝对差异在临床可接受的5 mm范围内。
  • 总计10%的扫描显示明显分割错误,主要由于未能检测到血管节段(肾动脉、主动脉分叉)或错误检测AAA外边界(十二指肠、下腔静脉、主动脉分支),这些扫描被排除在分析之外。

临床怎么用

  • 作者认为人工智能软件可作为研究和临床实践的重要辅助工具。
  • 作者建议经验丰富的观察者对生成的报告进行严格审查并谨慎使用,以识别错误的分割。

局限性

  • 10%的扫描显示明显分割错误,主要由于未能检测到血管节段(肾动脉、主动脉分叉)或错误检测AAA外边界(十二指肠、下腔静脉、主动脉分支),这些扫描被排除在分析之外。
  • 需要经验丰富的观察者对生成的报告进行严格审查并谨慎使用,以识别错误的分割。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

PURPOSE: This study investigates the accuracy of fully-automated maximum aortic diameter measurements in abdominal aortic aneurysm (AAA) patients using artificial intelligence software (PRAEVAorta 2, Nurea, Bordeaux, France). MATERIALS AND METHODS: This is a multicenter, retrospective validation study using prospectively collected data from the Zenith alpha for aneurysm Repair Registry (ZEPHYR). Automated measurements of PRAEVAorta 2 are compared with measurements of an internationally recognized core laboratory (Syntactx, New York, New York State). The reviewers at the core laboratory were measurement technologists trained to and utilizing established measurement standards, overseen by vascular surgeons and radiologists. The data set comprised 871 computed tomography angiography scans from the ZEPHYR registry with 347 patients who underwent endovascular aneurysm repair (EVAR) with the Zenith Alpha Endovascular Abdominal Graft (Cook Medical, Bloomington, Indiana) in Germany, Belgium, and The Netherlands between 2016 and 2019. RESULTS: The analysis demonstrated excellent correlation of the measurements (r=0.97) with an intraclass correlation (ICC) of 0.972 (95% confidence interval [CI]=0.968-0.976) across all scans. For preoperative computed tomography (CT) scans, ICC was 0.953 (95% CI=0.941-0.963), and for postoperative scans, ICC was 0.979 (95% CI=0.975-0,983), respectively. In total, 95.4% of measurements were within the clinically acceptable range of 5 mm in absolute difference. In total, 10% of scans demonstrated obvious segmentation errors, mainly due to failure in detecting vessel segments (renal arteries, aortic bifurcation) or due to mis-detecting the outer border of the AAA (duodenum, inferior vena cava, aortic branches) and were excluded from the analysis. CONCLUSION: In this study, the maximum AAA diameter could be accurately measured fully-automatically by PRAEVAorta 2 (Nurea) in most cases demonstrating that artificial intelligence (AI) software could serve as an important adjunct for research and clinical practice. However, critical review of the generated reports by an experienced observer and cautious use is warranted to identify flawed segmentations. CLINICAL IMPACT: This multicenter, retrospective validation study assessed the accuracy of fully-automated maximum infrarenal aortic aneurysm diameter measurements. It was demonstrated, that 95.4% of measurements were within the clinically acceptable range of 5 mm in absolute difference, positioning the software as a potential adjunct for clinical practice and research. It is also highlighted however, that critical review of the measurements is obligatory, due to a 10% segmentation error rate.

来源:J Endovasc Ther,查看原文。版权归原作者所有。

引用

  1. Hatzl J, Uhl C, Barb A, et al. External Validation of Fully-Automated Infrarenal Maximum Aortic Aneurysm Diameter Measurements in Computed Tomography Angiography Scans Using Artificial Intelligence (PRAEVAorta 2). J Endovasc Ther. 2024. PMID: 39534983. doi:10.1177/15266028241295563

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。