影响2区开窗主动脉弓腔内修复术中左锁骨下动脉桥接支架压缩的因素
该单中心回顾性研究纳入54例以单开窗装置保留左锁骨下动脉的患者,评估开窗主动脉弓腔内修复术后左锁骨下动脉支架压缩的发生率及解剖与技术因素的影响。随访中9例(18%)出现支架压缩,无卒中或上肢缺血;与最近骨结构距离较短和高迂曲指数与支架压缩相关。
要点
- 单中心回顾性研究纳入54例,随访中9例(18%)出现左锁骨下动脉支架压缩,无临床后果。
- 压缩组与最近骨结构距离为11.7±8.9 mm,非压缩组为23.0±7.8 mm(p=0.003)。
- 压缩组迂曲指数为1.3±0.4,非压缩组为1.2±0.1(p=0.03),与支架压缩相关。
核心结论
- 接受开窗主动脉弓腔内修复术的患者中约五分之一发生左锁骨下动脉支架压缩,但无临床后果;与最近骨结构距离较短和高迂曲指数与支架压缩相关。
研究设计与样本
- 单中心回顾性分析2012年1月1日至2023年11月30日期间使用单开窗装置(Cook Medical, Bloomington, IN, USA)保留左锁骨下动脉的患者,共纳入54例。
主要发现
- 仅使用球囊扩张覆膜支架,18%的病例在索引术中进行了relining。
- 30天时检测到1例支架压缩,随访期间无卒中或上肢缺血诊断。
- 随访中9例(18%)出现支架压缩,平均诊断时间为索引术后18个月(四分位距[IQR]=37,范围=1-58个月)。
- 5例(56%)接受了二次relining,再干预后随访无事件。
- 压缩组与最近骨结构距离为11.7±8.9 mm,非压缩组为23.0±7.8 mm(p=0.003);压缩组迂曲指数为1.3±0.4,非压缩组为1.2±0.1(p=0.03)。
临床怎么用
- 原文未据此给出推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
INTRODUCTION: Left subclavian artery (LSA) preservation during thoracic endovascular aortic repair (TEVAR) has been related to low morbidity. This study investigated the incidence of LSA stent compression in patients managed with fenestrated endovascular arch repair (f-Arch) and evaluated the impact of anatomic and technical factors on LSA stent outcomes. METHODS: A single-center retrospective analysis of patients managed with single-fenestration devices (Cook Medical, Bloomington, IN, USA) for LSA preservation, between January 1, 2012 and November 30, 2023, was conducted. Anatomic (arch type, bovine arch, distance between the LSA and most proximal bone structure, left common carotid artery and aortic lesion, take-off angle, diameter, thrombus, calcification, dissection, tortuosity) and technical parameters (stent type, diameter, length, relining, post-dilation) were evaluated. Stent compression was any ≥50% stenosis (using center luminal line) of the stent compared with its initial diameter. Clinical outcomes included stroke and upper limb ischemia at 30 days and follow-up. Technical outcomes included stent compression and need for reintervention. RESULTS: Fifty-four cases were included. Only balloon-expandable covered stents were used, and relining during the index procedure was performed in 18%. No stroke or arm ischemia was recorded. One stent compression was detected at 30 days. During follow-up, no stroke or arm ischemia was diagnosed. Nine cases (18%) presented stent compression, with a mean time of stent-compression diagnosis at 18 months (interquartile range [IQR]=37, range=1-58 months) after the index procedure. Five (56%) underwent secondary relining. Follow-up after reintervention was uneventful. Lower distance to the nearest bone structure (compression group [CG]: 11.7±8.9 mm vs non-compression group [NCG]: 23.0±7.8 mm, p=0.003) and higher tortuosity index (CG: 1.3±0.4 vs NCG: 1.2±0.1, p=0.03) were associated with LSA stent compression. CONCLUSION: LSA stent compression in patients managed with f-Arch affected 1 in 5 cases, without clinical consequences. Distance to the nearest bone structure and higher tortuosity were associated with LSA stent compression.Clinical ImpactFenestrated endovascular arch repair for the preservation of the left subclavian artery (LSA) in patients needing landing within the aortic arch has been performed with encouraging outcomes. This analysis showed that LSA stent compression is met in 18% of patients, without though any clinical consequence. Pre-operative anatomic parameters, as lower distance to the nearest bone structure and higher tortuosity index affect negatively LSA stent performance while stent parameters seem to have no impact.
引用
- Nana P, Giordano A, Panuccio G, et al. Factors Affecting Compression of the Left Subclavian Artery Bridging Stent In Zone 2 Fenestrated Endovascular Arch Repair. J Endovasc Ther. 2024. PMID: 39097800. doi:10.1177/15266028241267753
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。