Endonaut血管导航系统对移动C臂下腹主动脉腔内修复术中辐射暴露的影响
本研究回顾性比较了移动C臂下腹主动脉腔内修复术(EVAR)使用Endonaut导航系统前后的辐射暴露。共纳入153例患者,Endonaut组32例,对照组121例。Endonaut组剂量面积乘积(DAP)显著低于对照组(38 Gy cm2 ± 24 vs 76 Gy cm2 ± 51,P < 0.05),技术成功率无显著差异(97% vs 96%,P = 0.79)。
要点
- Endonaut组DAP显著低于对照组(38 Gy cm2 ± 24 vs 76 Gy cm2 ± 51,P < 0.05)。
- 两组空气比释动能(AK)和透视时间(FT)无显著差异。
- Endonaut组技术成功率为97%(31/32),对照组为96%(116/121)(P = 0.79)。
核心结论
- 使用Endonaut血管导航系统在移动C臂下进行EVAR可降低辐射剂量,且不影响技术成功率或手术时间。
研究设计与样本
- 回顾性分析2016年1月至2022年8月期间所有接受EVAR和/或髂支装置植入的患者,比较使用Endonaut导航系统前(对照组,n=121)与使用后(Endonaut组,n=32)的辐射数据。
主要发现
- Endonaut组DAP值显著低于对照组(38 Gy cm2 ± 24 vs 76 Gy cm2 ± 51,P < 0.05)。
- Endonaut组与对照组AK值无显著差异(196 mGy ± 114 vs 209 mGy ± 138)。
- Endonaut组与对照组透视时间无显著差异(33分钟 ± 18 vs 33分钟 ± 16)。
- Endonaut组技术成功率为97%(31/32),对照组为96%(116/121)(P = 0.79)。
- Endonaut组对比剂用量显著低于对照组(94 cc ± 41 vs 143 cc ± 66,P < 0.05)。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Fusion imaging systems have proved to reduce radiation exposure mostly in hybrid rooms but reports with mobile C-arms are few. The aim of this study was to analyze the impact of the Endonaut navigation system on radiation exposure in endovascular aneurysm repair (EVAR) performed with mobile C-arms. METHODS: All patients undergoing EVAR and/or iliac branched devices implantation between January 2016 and August 2022 were included. All procedures were performed with a mobile C-arm (Siemens Avantic or GE Elite until March 2018, Siemens Cios Alpha thereafter). The Endonaut navigation system has been used since January 2021. Two groups were, therefore, compared: before (control group [CGr]) and after the use of Endonaut. Radiation data including Dose Area Product (DAP) values, Air Kerma (AK) and fluoroscopy time (FT) were collected retrospectively. RESULTS: Overall, 153 patients were included: CGr, n = 121; Endonaut group (EnGr), n = 32. No significant difference was found between the 2 groups regarding demographic data. DAP values were significantly lower in the EnGr (38 Gy cm2 ± 24) vs. the CGr (76 Gy cm2 ± 51) (P < 0.05) despite a significantly higher number of complex procedures such as iliac branched devices (P < 0.05). AK values were not significantly different between the EnGr and the CGr (196 mGy ±114 vs. 209 mGy ±138) as well as FT (33 minutes ±18 vs. 33 minutes ±16). Technical success was 97% (31/32) in the EnGr vs. 96% (116/121) in the CGr (P = 0.79). The volume of contrast media was significantly lower in the EnGr (94 cc ± 41) vs. the CGr (143 cc ± 66) (P < 0.05). CONCLUSIONS: In this study, the use of the Endonaut angio-navigation system when performing EVAR with mobile C-arms led to a radiation dose reduction without compromising technical success or procedural time.
引用
- Malafosse C, Massiot N, Guimo F, et al. Impact of the Endonaut® Angio-Navigation System on Radiation Exposure in Endovascular Aortic Repair Performed with Mobile C-Arms. Ann Vasc Surg. 2024. PMID: 38986839. doi:10.1016/j.avsg.2024.04.013
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。