文献速递主动脉疾病

模块化内分支覆膜支架重建胸主动脉腔内修复术中左锁骨下动脉的临床前评估:猪实验研究

Preclinical Evaluation of a Modular Inner-Branched Stent Graft to Reconstruct the Left Subclavian Artery in Thoracic Endovascular Aortic Repair: Experimental Study in Pigs.
PubMed 文献 · J Endovasc Ther · 2024-04-01

在14头猪中经右股动脉和右颈动脉入路植入模块化单内分支覆膜支架,技术成功率100%(14/14),1头术后5小时突然死亡,1头完成随访CTA后死亡;存活猪随访CTA示支架完整无断裂、桥接覆膜支架均通畅,造影示位置形态贴附良好、无明显内漏,组织学示生物相容性良好。

要点

  • 猪模型植入模块化单内分支覆膜支架技术成功率100%(14/14)。
  • 随访CTA示支架完整无断裂、桥接覆膜支架均通畅,无明显内漏。
  • 1头猪术后5小时突然死亡,1头完成随访CTA后死亡。

核心结论

  • 在猪模型中部署模块化单内分支覆膜支架安全可行。

研究设计与样本

  • 在14头猪中经右股动脉和右颈动脉入路植入模块化内分支覆膜支架。
  • 随访结束时对所有猪行CT血管造影(CTA)和血管造影评估支架形态特征,随后处死并采集组织行大体和组织学检查。

主要发现

  • 技术成功率为100%(14/14)。
  • 1头猪术后5小时突然死亡,1头猪在完成随访CTA后死亡。
  • 随访期间所有存活猪精神状态良好、饮食和活动正常。
  • CTA示所有支架完整无断裂,所有桥接覆膜支架均通畅。
  • 血管造影示支架位置、形态和贴附良好,未见明显内漏;组织学示支架生物相容性良好。

临床怎么用

  • 该器械是中国首个为治疗邻近左锁骨下动脉的B型主动脉夹层而设计的模块化器械。
  • 该器械为模块化两组分系统,由带逆行内分支的胸主动脉覆膜支架和桥接覆膜支架组成。
  • 作者认为,模块化设计和逆行内分支是该器械的两项重要创新,理论上可使临床医生治疗邻近左锁骨下动脉的B型主动脉夹层更容易、更安全。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

PURPOSE: In approximate 40% of thoracic endovascular aortic repair (TEVAR) procedures, the left subclavian artery (LSA) needs to be covered to obtain sufficient proximal sealing zone. To preserve the LSA during the TEVAR for type B aortic dissection (TBAD) adjacent to LSA, our team designed a modular single inner-branched stent graft. This study was performed to evaluate the safety and feasibility of deploying a modular single inner-branched stent graft in a porcine model. MATERIALS AND METHODS: Modular inner-branched stent grafts were implanted in 14 pigs via right femoral and right carotid arterial access. Computed tomography angiography (CTA) and angiography were performed in all pigs to appraise the morphological characteristics of the stent grafts at the end of follow-up. The pigs were then euthanized, and tissues were collected for gross and histological examination. RESULTS: The technical success rate was 100% (14/14). One pig suddenly died 5 hours after operation, and 1 pig died after completing the follow-up CTA. During the follow-up period, all surviving pigs showed good mental state, normal diets and activities. Computed tomography angiography examinations showed that all stent grafts were intact without fracture. All bridging covered stents were patent. Angiography showed that the position, shape, and adhesion of the stent grafts were good, and no obvious endoleaks were found. Histological examination showed that the biocompatibility of the stent grafts was good. CONCLUSIONS: This study's outcomes demonstrate that it is safe and feasible to deploy a modular single inner-branched stent graft in a porcine model.Clinical ImpactThis device is the first modular device designed to treat TBAD adjacent to LSA in China. This device is a modular two-component system consisting of a thoracic aortic stent graft with a retrograde inner branch and a bridging covered stent. The modular design and the retrograde inner branch are the two important innovations of this device. Theoretically, the device could make it easier and safer for clinicians to treat TBAD adjacent to the LSA.

来源:J Endovasc Ther,查看原文。版权归原作者所有。

引用

  1. Jia S, Zhang H, Jia X, et al. Preclinical Evaluation of a Modular Inner-Branched Stent Graft to Reconstruct the Left Subclavian Artery in Thoracic Endovascular Aortic Repair: Experimental Study in Pigs. J Endovasc Ther. 2024. PMID: 38561988. doi:10.1177/15266028241241921

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。