Gore Excluder顺应性腔内移植物在肾下主动脉颈的几何分析:EXCeL注册研究一年结果
EXCeL前瞻性注册研究纳入4家医院2018—2022年植入Gore Excluder顺应性腔内移植物(CEXC)且有30天CT血管造影(CTA)的患者,评估30天与1年的移植物贴附(最短贴附长度SAL)与位置(最短织物距离SFD)。87例有30天CTA,其中56例有1年CTA;30天中位SAL为21.2 mm,1年SAL显著增加1.1 mm(p=.042),SFD轻微增加0.3 mm(p=.019),主动脉几何形态未见显著变化。
要点
- 30天中位SAL为21.2 mm,13例(15%)SAL<10 mm,1例SAL为0 mm并出现Ia型内漏。
- 1年SAL较30天显著增加1.1 mm(p=.042),SFD增加0.3 mm(p=.019)。
- 1例发生移植物移位(SFD增加≥5 mm);术后移植物倾斜中位数为15.8°。
核心结论
- 多数患者CEXC植入位置接近肾动脉,30天获得充分(≥10 mm)贴附,1年时贴附轻度增加;术后移植物倾斜相对较低,植入后主动脉几何形态保持不变,作者认为可能与其高顺应性有关。
研究设计与样本
- EXCeL为前瞻性注册研究,从4家医院选取2018—2022年间接受CEXC治疗且有30天CTA的患者。
- 87例有30天CTA,其中56例有1年CTA;在30天与1年CTA上评估SAL与SFD,并比较术前与术后CTA的肾下主动脉最大曲度。
主要发现
- 术前中位(四分位间距[IQR])肾下颈长为22 mm(IQR 15, 32),肾下成角为52°(IQR 31, 72)。
- 全部纳入患者30天中位SAL为21.2 mm(IQR 14.0, 29.3);13例(15%)30天SAL<10 mm,1例SAL为0 mm并出现Ia型内漏。
- 符合与不符合使用说明(IFU)的患者之间SAL无显著差异;1年SAL显著增加1.1 mm(IQR -2.3, 4.7;p=.042)。
- 1年时SAL在7例(13%)中减小、13例(23%)中增加、36例(64%)中保持稳定;30天中位SFD为2.0 mm(IQR 0.5, 3.6),1年轻微增加0.3 mm(IQR -0.5, 1.8;p=.019)。
- 1例出现移植物移位(SFD增加≥5 mm);中位移植物倾斜为15.8°(IQR 9.7, 21.4);术前肾下最大曲度为36 m-1(IQR 26, 56),其后未见显著变化。
关键数值与推荐
- 术后充分贴附定义为SAL≥10 mm;30天时13例(15%)SAL<10 mm,1例SAL为0 mm并出现Ia型内漏。
- 移植物移位定义为SFD增加≥5 mm,1例出现该情况。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVE: The Gore Excluder Conformable Endoprosthesis (CEXC) is designed to treat challenging infrarenal anatomy because of its active angulation control, repositionability, and enhanced conformability. This study evaluated 30 day and one year position and apposition of the CEXC in the infrarenal neck. METHODS: Patients treated with the CEXC between 2018 and 2022 with an available 30 day computed tomography angiogram (CTA) were selected from four hospitals in a prospective registry. Endograft apposition (shortest apposition length [SAL]) and position (shortest fabric distance [SFD]) were assessed on the 30 day and one year CTAs. Maximum infrarenal aortic curvature was compared between the pre- and post-operative CTAs to evaluate conformability of the CEXC. RESULTS: There were 87 patients with a 30 day CTA, and for 56 of these patients the one year CTA was available. Median (interquartile range [IQR]) pre-operative neck length was 22 mm (IQR 15, 32) and infrarenal angulation was 52° (IQR 31, 72). Median SAL was 21.2 mm (IQR 14.0, 29.3) at 30 days for all included patients. The SAL in 13 patients (15%) was < 10 mm at 30 days, and one patient had a SAL of 0 mm and a type Ia endoleak. There was no significant difference in SAL between patients within and outside instructions for use. The SAL significantly increased by 1.1 mm (IQR -2.3, 4.7; p = .042) at one year. The SAL decreased in seven patients (13%), increased in 13 patients (23%), and remained stable in 36 patients (64%). Median SFD was 2.0 mm (IQR 0.5, 3.6) at 30 days, which slightly increased by 0.3 mm (IQR -0.5, 1.8; p = .019) at one year. One patient showed migration (SFD increase ≥ 5 mm). Median endograft tilt was 15.8° (IQR 9.7, 21.4). Pre-operative maximum infrarenal curvature was 36 m-1 (IQR 26, 56) and did not significantly change thereafter. CONCLUSION: In most patients, the CEXC was implanted close to the renal arteries, and sufficient (≥ 10 mm) post-operative apposition was achieved at 30 days, which slightly increased at one year. Post-operative endograft tilt was relatively low, and aortic geometry remained unchanged after implantation of the CEXC, probably due to its high conformability.
引用
- Zuidema R, van Sambeek MRHM, Zwetsloot J, et al. Geometric Analysis of the Gore Excluder Conformable Endoprosthesis in the Infrarenal Aortic Neck: One Year Results of the EXCeL Registry. Eur J Vasc Endovasc Surg. 2024. PMID: 38670221. doi:10.1016/j.ejvs.2024.04.026