新型血管内抓取器用于困难下腔静脉滤器取出的研发
本研究研发了一种新型可关节活动的血管内抓取器,用于取出各种构型的永久性和可取出型下腔静脉滤器。体外实验显示抓取器取出居中永久滤器的时间显著短于圈套器(29秒 vs 79秒,P=0.009),动物实验中全部滤器成功取出且无不良事件,总手术时间和透视时间均短于匹配的患者组。
要点
- 体外实验中,抓取器取出居中永久滤器的时间显著短于圈套器(29秒 vs 79秒,P=0.009)。
- 动物模型中全部倾斜下腔静脉滤器经右颈静脉用抓取器成功取出,无不良事件。
- 动物模型总手术时间(21.2 vs 43.5分钟,P=0.01)和透视时间(4.3 vs 10分钟,P=0.044)均短于患者组。
核心结论
- 新型血管内抓取器可有效取出不同构型的不同类型下腔静脉滤器,体外模型中与圈套器相比结果有利。
- 体内实验显示与匹配的患者取出过程相比,抓取器取出更有效。
研究设计与样本
- 体外实验将下腔静脉滤器固定于柔性管内壁呈居中或倾斜构型,用高对比度背光相机模拟二维透视投影,共四组比较:可取出滤器居中、可取出滤器倾斜、永久滤器居中、永久滤器倾斜,每组重复五次,与圈套器对照。
- 体内实验采用猪模型,经右颈静脉入路用抓取器取出六个倾斜的肾下下腔静脉滤器,并与患者手术进行总手术时间、取出时间和透视时间的比较分析。
主要发现
- 体外实验中,实验组1、2、4的取出时间与对照组相当。
- 体外实验中,抓取器取出居中永久滤器(组3)的时间显著短于对照组(29秒 vs 79秒,P=0.009)。
- 动物模型中所有下腔静脉滤器均用抓取器取出,无不良事件。
- 动物模型总手术时间(21.2 vs 43.5分钟,P=0.01)和透视时间(4.3 vs 10分钟,P=0.044)显著短于患者组。
- 患者组中16.7%的取出需要先进血管内技术,一个下腔静脉滤器未能取出(成功率=91.7%),而动物模型全部成功取出且未使用额外工具。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVE: Although inferior vena cava (IVC) filters are commonly retrieved using a snare, lateral tilt and fibrosis around struts can complicate the procedure and sometimes require the use of off-label devices. We describe the development of a novel articulating endovascular grasper designed to remove permanent and retrievable IVC filters in any configuration. METHODS: For in vitro testing, the IVC filters were anchored to the inner wall of a flexible tube in a centered or tilted configuration. A high-contrast backlit camera view simulated the two-dimensional fluoroscopy projection during retrieval. The time from the retrieval device introduction into the camera field to complete filter retrieval was measured in seconds. The control experiment involved temporary IVC filter retrieval with a snare. There were four comparative groups: (1) retrievable filter in centered configuration; (2) retrievable filter in tilted configuration; (3) permanent filter in centered configuration; and (4) permanent filter in tilted configuration. Every experiment was repeated five times, with median retrieval time compared with the control group. For in vivo testing in a porcine model, six tilted infrarenal IVC filters were retrieved with grasper via right jugular approach. Comparison analysis between animal and patient procedures was performed for the following variables: total procedure time, the retrieval time, and fluoroscopy time. RESULTS: The in vitro experiments showed comparable retrieval times between the experimental groups 1, 2, and 4 and the control. However, grasper removal of a centered permanent filter (group 3) required significantly less time than in the control (29 vs 79 seconds; P = .009). In the animal model, all IVC filters were retrieved using the grasper with no adverse events. The total procedure time (21.2 vs 43.5 minutes; P = .01) and the fluoroscopy time (4.3 vs 10 minutes; P = .044) were significantly shorter in the animal model compared with the patient group. Moreover, in the patient group, 16.7% of retrievals required advanced endovascular techniques, and one IVC filter could not be retrieved (success rate = 91.7%), whereas all the IVC filters were successfully retrieved in the animal model without the use of additional tools. CONCLUSIONS: The novel endovascular grasper is effective in retrieving different types of IVC filters in different configurations and compared favorably with the snare in the in vitro model. In vivo experiments demonstrated more effective retrieval when compared with matched patient retrievals.