间歇性跛行患者运动经皮氧分压测量的重测信度与最小可检测变化
本研究纳入28例跛行下肢动脉疾病患者,间隔1个月内完成两次跑步机运动经皮氧分压测量,评估氧分压下降值(DROP)的重测信度。近端与远端组内相关系数分别为0.66与0.65,全部DROP值的测量标准误为8 mm Hg,最小可检测变化为23 mm Hg。作者认为该测量信度中等,个体干预后DROP变化≥23 mm Hg才可判定为显著变化。
要点
- 跛行下肢动脉疾病患者运动经皮氧分压测量的重测信度为中等。
- 近端与远端DROP的组内相关系数分别为0.66与0.65。
- 全部DROP值的测量标准误为8 mm Hg,最小可检测变化为23 mm Hg。
核心结论
- 在最大步行距离≤300m的下肢动脉疾病患者中,运动经皮氧分压测量DROP值具有中等的重测信度。
- 个体干预后DROP改善或恶化≥23 mm Hg,才有95%把握认为变化显著。
- 作者认为在评估跛行患者治疗效果时应考虑该测量。
研究设计与样本
- 数据来自在1个月间隔内完成两次跑步机测试的患者。
- 在双侧臀部(近端)和双侧小腿(远端)测量氧分压下降值(DROP)。
- 通过两次记录经皮氧分压测量评估重测重现性,以测量标准误和组内相关系数表示。
- 最小可检测变化按公式MDC=SEM×1.96×√2计算。
主要发现
- 共纳入28例下肢动脉疾病患者,年龄61±9岁。
- 近端组内相关系数为0.66 [0.50, 0.79],远端为0.65 [0.49, 0.79]。
- 近端DROP的测量标准误为7 [6, 9] mm Hg,远端为9 [8, 11] mm Hg。
- 全部(近端和远端)DROP值的测量标准误为8 [7, 10] mm Hg,DROP的最小可检测变化为23 mm Hg。
临床怎么用
- 作者认为,在评估跛行患者治疗效果时,应考虑运动经皮氧分压测量DROP值。
- 作者指出,个体干预后DROP改善或恶化≥23 mm Hg,才有95%把握认为变化显著。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Exercise transcutaneous oxygen pressure measurement (Exercise-TcPO2) can be used to diagnose Lower Extremity Artery Disease (LEAD) and allows the quantification of limb ischemia during exercise on treadmill. Exercise-TcPO2 test-retest reliability in patients with LEAD and severe walking impairment is unknown. The aim of this study was to evaluate the test-retest reliability, standard error of measurement (SEM), and Minimal Detectable Change (MDC) of exercise-TcPO2 in patients with claudication. METHODS: Data were collected from patients that performed 2 treadmill tests within a 1-month interval. Delta from Rest of Oxygen Pressure (DROP) values were measured at both buttocks (proximal) and both calves (distal). Test-retest reproducibility was assessed by recording transcutaneous oximetry measurements twice and expressed as SEM and intra-class correlation coefficients. MDC was calculated using the formula MDC = SEM x 1.96 x √ 2. RESULTS: Twenty eight LEAD patients (61 ± 9 years old) were included. Intra-class correlation coefficients were 0.66 [0.50, 0.79] and 0.65 [0.49, 0.79] for the proximal and distal levels, respectively. The SEM of DROP at the proximal and distal levels were 7 [6, 9] mm Hg and 9 [8, 11] mm Hg, respectively. The SEM for all (proximal and distal) DROP values was 8 [7, 10] mm Hg and the MDC of DROP was 23 mm Hg. CONCLUSIONS: Exercise-TcPO2 with measurement of DROP values has a moderate test-retest reliability in LEAD patients with a maximal walking distance ≤ 300m. For an individual, an improvement or deterioration in DROP of ≥ 23 mm Hg after an intervention would be required to be 95% confident that the change is significant. It should be considered in evaluating the impact of treatment in patients with claudication.