将定制开窗整合以简化急性肾内脏区主动脉原位修复
该技术报告描述了一种在预设计定制短开窗覆膜支架上采用顺行原位激光开窗(ISLF)的技术,用于急性复杂腹主动脉瘤的急诊腔内修复。支架在融合影像引导下释放,定制开窗对准肠系膜上动脉并即刻桥接,随后用分叉支架隔绝瘤体,再用大号可调弯鞘完成序贯顺行激光原位肾动脉开窗和支架植入。作者认为该技术可行,或可成为限制主动脉覆盖范围和肾内脏缺血时间的即用型解决方案。
要点
- 定制短开窗支架预载单个8 mm无支柱开窗,位于12点钟位置,对准肠系膜上动脉。
- 经双侧经皮股动脉入路,在融合影像引导下释放支架并即刻桥接肠系膜上动脉。
- 采用大号可调弯鞘完成序贯顺行激光原位肾动脉开窗和支架植入。
核心结论
- 作者认为,针对肠系膜上动脉的单血管定制短开窗支架联合顺行原位激光肾动脉开窗,在技术上可用于急性复杂腹主动脉瘤的修复,即使既往接受过肾下腹主动脉重建。
研究设计与样本
- 本文为技术报告,描述在预设计定制短开窗覆膜支架上实施顺行原位激光开窗的技术。
主要发现
- 定制支架预载单个8 mm无支柱开窗,位于12点钟位置。
- 改良预载系统可在需要时允许从远端端口进行单侧入路。
- 支架在融合影像引导下释放,定制开窗对准肠系膜上动脉开口并即刻按标准技术桥接。
- 动脉瘤随后用分叉支架隔绝。
- 使用大号可调弯鞘完成序贯顺行激光原位开窗和肾动脉支架植入。
临床怎么用
- 作者认为,该技术或可成为即用型解决方案,以限制主动脉覆盖范围和肾内脏缺血时间,即使肾水平主动脉直径狭窄的患者也可应用。
- 作者认为,该技术也可能适用于既往接受过肾下腹主动脉瘤腔内修复(EVAR)的患者。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
PURPOSE: To illustrate the technique of antegrade in situ laser fenestration (ISLF) on a predesign custom-manufactured stent-graft with single reinforced fenestration for use in emergency endovascular repair of complex abdominal aortic aneurysms (AAAs). TECHNIQUE: A short custom-made device (CMD) fenestrated graft was predesigned with a single preloaded 8 mm strut-free fenestration at 12 o'clock position. A modified preloaded system was used to allow unilateral access from the distal port if necessary. After bilateral percutaneous femoral access, the graft was deployed under fusion guidance with the CMD fenestration matching the superior mesenteric artery (SMA) origin and immediately bridged as per standard technique. The aneurysm was then excluded with a bifurcated device. A large steerable sheath was used to allow for sequential antegrade laser in situ fenestration and stenting of the renal arteries. CONCLUSIONS: Single-vessel customized short fenestrated grafts for the SMA and antegrade in situ laser renal fenestrations are technically feasible for repair of acute complex AAAs even after previous infrarenal reconstruction. It could become an off-the-shelf solution to limit aortic coverage and reno-visceral ischemia, even in patients with a narrow aortic diameter at the renal level.Clinical ImpactSingle-vessel precustomized short fenestrated grafts for the SMA combined with renal artery antegrade ISLF can be a feasible option for the acute repair of patients with complex aneurysms and a narrow aortic diameter at the reno-visceral segment. It may limit aortic coverage and reno-visceral ischemic time and also be applicable after previous infrarenal endovascular aneurysm repair (EVAR).
引用
- Berczeli M, Sonesson B, Karelis A, et al. Integration of a Custom-Made Fenestration to Simplify Acute Reno-Visceral In Situ Aortic Repair. J Endovasc Ther. 2023. PMID: 37902446. doi:10.1177/15266028231208656