文献速递静脉血栓栓塞

两种下腔静脉滤器取出成功率和费用的比较

Comparison of success and cost after retrieval of two inferior vena cava filters.
PubMed 文献 · J Vasc Surg Venous Lymphat Disord · 2023-09-12

单中心研究比较Denali与Option下腔静脉滤器的取出特征、并发症和总费用。Denali滤器首次尝试取出成功率更高(94% vs 77%),手术和透视时间更短,滤器显著倾斜和钩嵌入下腔静脉壁更少。滤器钩穿透下腔静脉壁显著降低取出成功率(41% vs 99%)。

要点

  • Denali滤器首次尝试取出成功率94%,Option Elite为77%(P = .019)。
  • Denali滤器手术时间29分钟 vs 63分钟,透视时间7分钟 vs 25分钟(均P < .001)。
  • 滤器钩嵌入下腔静脉壁时取出成功率41% vs 99%(P < .001),倾斜≥15°对成功率无显著影响。

核心结论

  • 研究结果提示,尽管两种滤器设计相似,Denali滤器在取出便利性方面优于Option滤器。
  • Option滤器发生滤器倾斜和钩嵌入下腔静脉壁显著更多,这些因素可能使取出更困难,并导致手术和透视时间更长。

研究设计与样本

  • 单中心回顾性研究,纳入2016年3月至2021年10月期间在该中心尝试取出Denali或Option下腔静脉滤器的所有患者。
  • 共87例患者,取出52枚Denali滤器和35枚Option Elite滤器。
  • 纳入标准为有影像学检查可评估滤器放置、倾斜存在或程度和/或滤器钩嵌入下腔静脉壁。

主要发现

  • Denali滤器首次尝试取出成功率94% vs Option Elite 77%(P = .019)。
  • Denali滤器手术时间29分钟 vs Option Elite 63分钟(P < .001),透视时间7分钟 vs 25分钟(P < .001)。
  • Denali滤器取出时显著倾斜(≥15°)比例12% vs Option Elite 29%(P < .001)。
  • Denali滤器钩嵌入下腔静脉壁比例6% vs Option Elite 40%(P < .001)。
  • 滤器倾斜≥15°对取出成功率无显著影响(无倾斜或倾斜<15° vs 倾斜≥15°:98% vs 100%;P = .58)。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVE: The objective of this study was to help guide inferior vena cava (IVC) filter choices by better understanding the retrieval characteristics, complications, and total costs between two commonly used IVC filters. METHODS: All patients who underwent retrieval or attempted retrieval of Denali (Bard Peripheral Vascular) or Option (Argon Medical Devices) IVC filters were identified between March 2016 and October 2021 at a single tertiary care center. Those with imaging studies that permitted evaluation of filter placement, presence or degree of tilt, and/or hooking of the filter into the IVC wall were included in the present study. Filter retrieval success, number of attempts, use of advanced techniques, and fluoroscopy and procedural times were recorded and compared between the two filters. RESULTS: A total of 87 patients presented for retrieval of 52 Denali and 35 Option Elite filters during the study period. Denali filters were more likely to be successfully retrieved at the first attempt (94% vs 77%; P = .019). The procedural and fluoroscopy times were shorter for Denali filters (29 minutes vs 63 minutes [P < .001] and 7 minutes vs 25 minutes [P < .001], respectively). Denali filters were less likely to be significantly tilted (≥15○) at retrieval (12% vs 29%; P < .001) or to have the filter hook embedded in the IVC wall (6% vs 40%; P < .001). Tilting of the filter of ≥15○ had no significant effects on the retrieval success rate (no tilt or tilt <15○ vs tilt of ≥15○: 98% vs 100%; P = .58). In contrast, filter hook penetration into the IVC wall significantly reduced successful recovery (41% vs 99%; P < .001). CONCLUSIONS: The findings from this study suggest that although the filter designs are similar, a benefit exists in the ease of retrievability of the Denali over the Option filter. We found that tilting and hooking of the filter in the IVC wall occurred significantly more with the Option filter. These factors likely made retrieval more difficult and contributed to the longer procedure and fluoroscopy times.

来源:J Vasc Surg Venous Lymphat Disord,查看原文。版权归原作者所有。

引用

  1. Bhinder J, O'Brien-Irr M, Chang M, et al. Comparison of success and cost after retrieval of two inferior vena cava filters. J Vasc Surg Venous Lymphat Disord. 2023. PMID: 37708934. doi:10.1016/j.jvsv.2023.09.001

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。