文献速递主动脉疾病内脏/肾动脉与血管畸形

开窗与分支腔内动脉瘤修复术中血管内超声检测桥接支架不稳定性的多中心研究:274条靶血管

Intravascular Ultrasound in the Detection of Bridging Stent Graft Instability During Fenestrated and Branched Endovascular Aneurysm Repair Procedures: A Multicentre Study on 274 Target Vessels.
PubMed 文献 · Eur J Vasc Endovasc Surg · 2023-09-12

这项前瞻性观察性多中心研究纳入4个主动脉中心80例接受开窗与分支腔内动脉瘤修复术的患者,评估血管内超声检测桥接支架不稳定的安全性和有效性。血管内超声评估技术成功率为91.9%(274/298),无血管内超声相关不良事件;7条靶血管(2.5%)经血管内超声发现桥接支架不稳定并立即修正。作者认为血管内超声是安全且有潜在价值的辅助影像技术,需更大队列的长期研究验证。

要点

  • 血管内超声评估技术成功率为91.9%(274/298),无血管内超声相关不良事件。
  • 7条靶血管(2.5%)经血管内超声发现桥接支架不稳定并立即修正。
  • 术后首次CTA中7/268条靶血管(2.6%)因桥接支架不稳定需再干预。

核心结论

  • 本研究提示血管内超声是术中检测靶内脏血管不稳定的安全且有潜在价值的辅助影像技术。
  • 作者提出需更大队列的长期研究来验证这些有希望的结果,并从效率、辐射暴露、手术时间和费用方面比较血管内超声与替代技术。

研究设计与样本

  • 前瞻性观察性多中心研究,纳入4个主动脉中心2019年1月至2021年9月间接受开窗与分支腔内动脉瘤修复术的80例患者(69%为男性;中位年龄72岁,四分位距59、77岁)。
  • 手术类型包括70例分支腔内动脉瘤修复术(21例现成支架、49例定制支架)、8例开窗腔内动脉瘤修复术(定制支架)和2例开窗与分支腔内动脉瘤修复术(定制支架),共300条潜在靶内脏血管。
  • 2条靶内脏血管(0.7%)未置入支架并被排除分析;7条(2.3%)因血管内超声导管无法进入、17条(5.7%)因血管内超声导管故障未能检查。
  • 所有手术均在临时杂交手术室完成。

主要发现

  • 血管内超声评估技术成功率为91.9%(274/298),无血管内超声相关不良事件。
  • 7条靶血管(2.5%)经血管内超声显示桥接支架不稳定,并立即进行修正。
  • 在274条经血管内超声评估的靶血管中,268条有术后至少30天的首次CTA;其中7条(2.6%)因桥接支架不稳定需再干预。
  • 血管内超声检查结果与术中血管造影的比较、以及术后CTA发现而血管内超声未检出的发生率,原文未报告具体数值。

关键数值与推荐

  • 靶内脏血管不稳定定义为需要额外操作或再干预的任何分支或开窗问题。
  • 任何血管内超声评估发现狭窄、扭结或任何靶内脏血管几何问题均视为分支不稳定。

临床怎么用

  • 本研究提示血管内超声可作为术中检测靶内脏血管不稳定的安全且有潜在价值的辅助影像技术;作者提出需更大队列的长期研究验证并与其他技术比较效率和费用等。

局限性

  • 作者指出需更大队列的长期研究来验证这些有希望的结果。
  • 作者指出需从效率、辐射暴露、手术时间和费用方面比较血管内超声与替代技术。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVE: The use of intravascular ultrasound (IVUS) reduces contrast medium use and radiation exposure during conventional endovascular aneurysm repair (EVAR). The aim of this study was to evaluate the safety and efficacy of IVUS in detecting bridging stent graft (bSG) instability during fenestrated and branched EVAR (F/B-EVAR). METHODS: This was a prospective observational multicentre study. The following outcomes were evaluated: (1) technical success of the IVUS in each bSG, (2) IVUS findings compared with intra-operative angiography, (3) incidence of post-operative computed tomography angiography (CTA) findings not detected with IVUS, and (4) absence of IVUS related adverse events. Target visceral vessel (TVV) instability was defined as any branch or fenestration issues requiring an additional manoeuvre or re-intervention. Any IVUS assessment that detected stenosis, kinking, or any geometric TVV issue was considered to be branch instability. All procedures were performed in ad hoc hybrid rooms. RESULTS: Eighty patients (69% males; median age 72 years; interquartile range 59, 77 years) from four aortic centres treated with F/B-EVAR between January 2019 and September 2021 were included: 70 BEVAR (21 off the shelf; 49 custom made), eight FEVAR (custom made), and two F/B-EVAR (custom made), for a total of 300 potential TVVs. Two TVVs (0.7%) were left unstented and excluded from the analysis. The TVVs could not be accessed with the IVUS catheter in seven cases (2.3%). Furthermore, 17 (5.7%) TVVs could not be examined due to a malfunction of the IVUS catheter. The technical success of the IVUS assessment was 91.9% (274/298), with no IVUS related adverse events. Seven TVVs (2.5%) showed signs of bSG instability by means of IVUS, leading to immediate revisions. The first post-operative CTA at least 30 days after the index procedure was available in 268 of the 274 TVVs originally assessed by IVUS. In seven of the 268 TVVs (2.6%) a re-intervention became necessary due to bSG instability. CONCLUSION: This study suggests that IVUS is a safe and potentially valuable adjunctive imaging technology for intra-operative detection of TVV instability. Further long term investigations on larger cohorts are required to validate these promising results and to compare IVUS with alternative technologies in terms of efficiency, radiation exposure, procedure time, and costs.

来源:Eur J Vasc Endovasc Surg,查看原文。版权归原作者所有。

引用

  1. Asciutto G, Ibrahim A, Leone N, et al. Intravascular Ultrasound in the Detection of Bridging Stent Graft Instability During Fenestrated and Branched Endovascular Aneurysm Repair Procedures: A Multicentre Study on 274 Target Vessels. Eur J Vasc Endovasc Surg. 2023. PMID: 37704100. doi:10.1016/j.ejvs.2023.09.009