文献速递主动脉疾病

X线引导血管内手术的蒙特卡洛剂量学验证

Monte Carlo Dosimetry Validation for X-Ray Guided Endovascular Procedures.
PubMed 文献 · Ann Vasc Surg · 2023-09-17

本研究在EVAR患者中,用光激发光被动剂量计测量剑突、耻骨联合及双侧髂嵴4个皮肤解剖点的剂量,并与GATE和GGEMS蒙特卡洛模拟剂量比较。GATE法相关系数r=0.97(95%CI 0.96-0.98),GGEMS法r=0.96(95%CI 0.94-0.97),平均相对误差分别为5±5%和6±5%。

要点

  • GATE模拟剂量与实测剂量高度相关(r=0.97,95%CI 0.96-0.98)。
  • GGEMS模拟剂量与实测剂量高度相关(r=0.96,95%CI 0.94-0.97)。
  • GGEMS执行时间6秒,快于GATE的5小时。

核心结论

  • 在当前影像设置下,GATE和GGEMS蒙特卡洛模拟平台有潜力模拟EVAR术中的三维剂量分布。

研究设计与样本

  • 临床研究,纳入接受EVAR治疗的患者,在剑突、耻骨联合、右侧髂嵴和左侧髂嵴4个皮肤解剖点用光激发光被动剂量计测量剂量,并与GATE和GGEMS蒙特卡洛模拟的对应剂量比较。

主要发现

  • GATE法模拟剂量与实测剂量高度相关(P<0.0001;r=0.97;95%CI 0.96-0.98)。
  • GGEMS法模拟剂量与实测剂量高度相关(P<0.0001;r=0.96;95%CI 0.94-0.97)。
  • 所有测量的平均相对误差在GATE组为5±5%,在GGEMS组为6±5%。
  • GGEMS执行时间为6秒,GATE蒙特卡洛模拟为5小时,GGEMS更快。

临床怎么用

  • 原文未据此给出临床推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Endovascular treatment is continuously gaining ground in vascular surgery procedures. However, current patient radiation dose estimation does not take into account the exact patient morphology and organs' composition. Monte Carlo (MC) simulation can accurately estimate the dose by recreating the irradiation process generated during X-ray-guided interventions. This study aimed to validate the MC simulation models by comparing simulated and measured dose distributions in endovascular aortic aneurysm repair (EVAR) procedures. METHODS: We conducted a clinical study in patients treated for EVAR. Patient dose measurements were taken with passive dosimeters using Optically Stimulated Luminescence technology in 4 specific anatomical points on the skin: xiphoid process, pubic symphysis, right and left iliac crest. Dose measurements were compared to the corresponding simulated doses with the Geant4 Application for Emission Tomography (GATE) and GPU Geant4-based Monte Carlo Simulations (GGEMS) MC simulations softwares. The MC simulation took as input the computed tomography scan of the patient and the parameters of the imaging system (orientation angles, tube voltage, and aluminum filtration) and gives as output the three-dimensional (3D) dose map for each patient and angulation. RESULTS: A good agreement with real doses was found for doses simulated by the MC GATE method (P < 0.0001; r = 0.97; 95% confidence interval [CI] [0.96-0.98]), as well as for doses simulated by the GGEMS method (P < 0.0001; r = 0.96; 95% CI [0.94-0.97]). The mean relative error for all measurements was 5 ± 5% in the MC GATE group and 6 ± 5% in the GGEMS group. Process execution on GGEMS (6 sec) was faster than the GATE MC simulation (5 hr). CONCLUSION: Considering the current imaging settings, this study shows the potential of using the GATE and GGEMS MC simulations platforms to model the 3D dose distributions during EVAR procedures.

来源:Ann Vasc Surg,查看原文。版权归原作者所有。

引用

  1. Nasr B, Villa M, Benoit D, et al. Monte Carlo Dosimetry Validation for X-Ray Guided Endovascular Procedures. Ann Vasc Surg. 2023. PMID: 37717818. doi:10.1016/j.avsg.2023.07.104

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。