模块化Endo-Bentall覆膜支架系统用于A型主动脉夹层的解剖可行性
本研究纳入152例A型主动脉夹层患者,基于CT血管造影测量评估模块化Endo-Bentall覆膜支架系统的解剖可行性。应用排除标准后66.4%的患者解剖上适合该支架系统,85.1%的患者适合长度为70 mm、80 mm或90 mm的支架。作者认为约三分之二的A型主动脉夹层患者解剖上适合血管内修复,需进一步动物研究优化器械设计。
要点
- 应用排除标准后66.4%的A型主动脉夹层患者解剖上适合模块化Endo-Bentall支架系统。
- 85.1%的患者适合长度为70 mm、80 mm或90 mm的支架。
- 根据窦管交界与头臂干直径差异,需同时准备顺向和逆向锥形支架。
核心结论
- 采用模块化Endo-Bentall覆膜支架设计,约三分之二的A型主动脉夹层患者解剖上适合血管内修复。
- 需进一步动物研究以优化器械设计。
研究设计与样本
- 纳入152例A型主动脉夹层患者进行影像测量与分析。
- 所有CT血管造影图像在3mensio工作站10.2版(3mensio Medical Imaging B.V.)上采用中心线法评估。
- 比较各平面直径与长度与模块化Endo-Bentall覆膜支架系统拟议解剖标准的符合情况。
主要发现
- 患者以男性为主(67.1%),中位年龄56.5岁(四分位距50.0-65.0岁)。
- 91.5%的主动脉夹层向近端延伸至窦管交界,仅8.6%局限于管状升主动脉。
- 主动脉瓣环周长派生直径中位数为24.1 mm。
- 窦管交界和头臂干处主动脉最大直径中位数分别为44.6 mm和43.5 mm。
- 左冠状动脉、右冠状动脉和主动脉窦高度中位数分别为12.7 mm、16.7 mm和28.4 mm。
关键数值与推荐
- 应用排除标准后,66.4%的患者解剖上适合模块化Endo-Bentall覆膜支架系统。
- 85.1%的患者适合长度为70 mm、80 mm或90 mm的支架。
- 根据窦管交界与头臂干之间直径差异,需同时准备顺向和逆向锥形支架。
临床怎么用
- 原文未据此给出推荐。
局限性
- 需进一步动物研究以优化器械设计。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVE: The implementation of thoracic endovascular aortic repair in patients with type A aortic dissection has been strictly constrained due to the pulsatile movement and distensibility and the insufficient length of landing zones on ascending aorta. The most prevalent anatomical limitation is the insufficient length of proximal and distal landing zones. We propose a modularly designed Endo-Bentall stent graft system to broaden the scope of thoracic endovascular aortic repair in the ascending aorta by covering intimal tears in the aortic root and ascending aorta and reconstructing coronary arteries. This study was conducted to assess the anatomical feasibility of a novel stent graft design. METHODS: In this study, we included 152 patients with type A aortic dissection for image measurement and analysis. All computed tomography angiography images were assessed on a 3mensio Workstation version 10.2 (3mensio Medical Imaging B.V.) utilizing the centerline method. We compared the diameters and lengths at various planes in relation to the proposed anatomical criteria for the modular Endo-Bentall stent graft system. RESULTS: The patients were predominantly male (67.1%), with a median age of 56.5 years (interquartile range, 50.0-65.0 years). Among all aortic dissections, 91.5% extended proximally to the sinotubular junction, whereas only 8.6% were restricted to the tubular ascending aorta. The median perimeter-derived diameter of the aortic annulus was 24.1 mm. The median maximum aortic diameter at the sinotubular junction and brachiocephalic trunk were 44.6 mm and 43.5 mm, respectively. The median height of the left coronary artery, right coronary artery, and sinus of Valsalva were 12.7 mm, 16.7 mm, and 28.4 mm, respectively. After applying exclusion criteria, 66.4% of all patients were anatomically eligible for the modular Endo-Bentall stent graft system. A total of 85.1% of patients were suitable for stent grafts with lengths of 70 mm, 80 mm, or 90 mm. Both antegradely and retrogradely tapered stent grafts were required, according to the diameter differences between the STJ and brachiocephalic trunk. CONCLUSIONS: Utilizing the modular Endo-Bentall stent-graft design, approximately two-thirds of patients with type A aortic dissection are anatomically eligible for endovascular repair. Further animal studies are required to optimize the device design.