文献速递主动脉疾病

编辑推荐——血管腔内动脉瘤修复术中术者估计辐射剂量

Editor's Choice - Estimated Radiation Dose to the Operator During Endovascular Aneurysm Repair.
PubMed 文献 · Eur J Vasc Endovasc Surg · 2023-04-01

本研究利用辐射剂量结构化报告中的详细曝光信息,通过蒙特卡洛方法计算kerma面积乘积(PKA)与术者器官剂量的转换系数,并估算81例肾下腹主动脉瘤腔内修复术(EVAR)的术者剂量。无屏蔽时中位有效剂量为113 μSv,结肠和胃的中位器官剂量最高;使用0.25 mm铅屏蔽躯干和上腿可降低有效剂量约6倍,加用天花板和床旁屏蔽可实现25至50倍降低。

要点

  • 无屏蔽时EVAR术者中位有效剂量为113 μSv(IQR 71,252 μSv)。
  • 最高中位器官剂量为结肠154 μSv(IQR 81,343)和胃133 μSv(IQR 76,307)。
  • 0.25 mm铅屏蔽躯干和上腿可降低有效剂量约6倍,加用天花板和床旁屏蔽可达25至50倍降低。

核心结论

  • 模型提示,优化使用屏蔽可将术者剂量降至相当于一至两天自然本底暴露的水平,并远低于法定剂量限值。

研究设计与样本

  • 利用蒙特卡洛方法计算91个束角和7种典型X射线能谱下PKA与术者器官剂量的转换系数,并编写计算机程序根据结构化报告中的每次曝光选择相应转换系数乘以PKA,用于估算81例EVAR术者的剂量。

主要发现

  • 无任何屏蔽时,中位估计有效剂量为113 μSv(IQR 71,252 μSv)。
  • 最高中位器官剂量为结肠154 μSv(IQR 81,343)和胃133 μSv(IQR 76,307)。
  • 使用0.25 mm铅覆盖躯干和上腿提供最小屏蔽时,有效剂量降低约6倍。
  • 加用天花板和床旁屏蔽后,剂量可实现25至50倍降低。
  • 估计剂量在主要束流直接背离术者时最高。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVE: To estimate operator organ doses from fluoroscopically guided infrarenal endovascular aneurysm repair (EVAR) procedures, using the detailed exposure information contained in radiation dose structured reports. METHODS: Conversion factors relating kerma area product (PKA) to primary operator organ doses were calculated using Monte Carlo methods for 91 beam angles and seven x-ray spectra typical of clinical practice. A computer program was written, which selects the appropriate conversion factor for each exposure listed in a structured report and multiplies it by the respective PKA. This system was used to estimate operator doses for 81 EVAR procedures for which structured reports were available. The impact of different shielding scenarios and variations in operator position was also investigated. RESULTS: Without any shielding, the median estimated effective dose was 113 μSv (interquartile range [IQR] 71, 252 μSv). The highest median organ doses were for the colon (154 μSv, IQR 81, 343) and stomach (133 μSv, IQR 76, 307). These dose estimates represent all exposures, including fluoroscopy and non-fluoroscopic digital acquisitions. With minimal shielding provided by 0.25 mm of Pb covering the torso and upper legs, the effective dose was reduced by a factor of around 6. With additional shielding from ceiling and table shields, a 25 to 50 fold reduction in dose is achievable. Estimated doses were highest where the primary beam was pointed directly away from the operator. CONCLUSION: The models suggest that with optimal use of shielding, operator doses can be reduced to levels equivalent to one to two days of natural background exposure and well below statutory dose limits.