文献速递主动脉疾病颈动脉与弓上血管

应用Upstream Peripheral Go Back再入导管行原位开窗胸主动脉腔内修复术

In Situ Fenestrated Thoracic Endovascular Repair Using the Upstream Peripheral Go BackTM Re-Entry Catheter.
PubMed 文献 · J Endovasc Ther · 2023-01-09

本文报告一例74岁男性既往胸主动脉腔内修复术后出现主动脉支气管瘘及Ia型内漏,因反复咯血需紧急修复。采用Upstream Peripheral Go Back导管行原位开窗胸主动脉腔内修复术,成功创建两个开窗并实现近端密封及目标血管通畅,术后开胸确认主动脉隔绝。作者认为在急需修复且定制胸主动脉装置不可用时,Go Back导管原位开窗可作为激光开窗的替代方案。

要点

  • 74岁男性既往TEVAR后出现主动脉支气管瘘和Ia型内漏,表现为咯血、发热和恶病质。
  • 使用Go Back导管行原位开窗TEVAR,成功创建两个开窗,近端密封和目标血管通畅经CTA确认。
  • 作者认为在急需修复且定制装置不可用时,Go Back导管原位开窗是激光开窗的替代方案。

核心结论

  • 在急需修复且定制胸主动脉装置不可用的情况下,使用Go Back导管行原位开窗是激光开窗的替代方案。

研究设计与样本

  • 本文为病例报告,报告一例74岁男性既往TEVAR后出现主动脉支气管瘘和Ia型内漏,因反复咯血需紧急修复。

主要发现

  • 术前CTA显示主动脉支气管瘘和Ia型内漏。
  • 使用Upstream Peripheral Go Back导管成功创建两个开窗。
  • CTA确认近端密封和目标血管通畅。
  • 开胸手术确认主动脉隔绝。
  • 行楔形肺叶切除术联合带蒂大网膜切除术以隔绝暴露的覆膜支架。

临床怎么用

  • 作者认为在急需修复且定制胸主动脉装置不可用时,Go Back导管原位开窗可作为激光开窗的替代方案。
  • 作者在临床影响中表述,在急性TEVAR中为达到弓部近端密封常需开窗,Go Back导管原位开窗在此情况下是激光开窗和PMEGs的替代方案。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

PURPOSE: The aim is to present a case of in situ fenestrated thoracic endovascular repair, using a re-entry peripheral catheter, needing urgent repair due to recurrent episodes of hemoptysis because of an aorto-bronchial fistula after previous thoracic endovascular aortic repair (TEVAR). TECHNIQUE: A 74-year-old male with a previous TEVAR presented with hemoptysis, fever, and cachexia. An aorto-bronchial fistula and type Ia endoleak were depicted on computed tomography angiography (CTA). Due to the patient's frail general condition, a proximal TEVAR extension was planned, prior to thoracotomy for debridement, and long-term antibiotics. An in situ fenestrated TEVAR was performed to provide proximal sealing and revascularize the left common carotid and subclavian arteries. For fenestration creation, an Upstream Peripheral Go Back catheter (Bentley, Hechingen, Germany) was used. The Go Back catheter has been approved as a re-entry tool for complex peripheral arterial revascularization. In this case, both fenestrations were successfully created. Proximal seal and target vessel patency were confirmed on CTA. Thoracotomy confirmed aortic exclusion. A wedge lobectomy was performed in combination with pediculated omentectomy to exclude the exposed endograft. CONCLUSION: In urgent cases, where custom-made thoracic devices are unavailable, in situ fenestrations with the Go Back catheter are an alternative to laser fenestrations. CLINICAL IMPACT: Fenestrations are often required in acute TEVAR to achieve a proximal seal in the arch. In situ fenestrations with the Go Back catheter are an alternative to laser fenestrations and PMEGs in this setting.

来源:J Endovasc Ther,查看原文。版权归原作者所有。

引用

  1. Nana P, Le Houérou T, Gaudin A, et al. In Situ Fenestrated Thoracic Endovascular Repair Using the Upstream Peripheral Go BackTM Re-Entry Catheter. J Endovasc Ther. 2023. PMID: 36624612. doi:10.1177/15266028221144589

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。