对比增强断层三维超声与超声图像融合测量颈动脉斑块体积(负荷)的可行性与准确性
本研究在颈动脉内膜剥脱术(CEA)前用断层超声(tUS)、对比增强断层超声(CEtUS)及两者融合图像测量颈动脉斑块体积(CPV),以Archimedes技术测得标本体积为参照。129例标本平均CPV为0.75±0.43 cm³;tUS、CEtUS和融合图像所测CPV与标本体积差异均无临床意义,观察者内/间差异极小。结论认为tUS可准确测量CPV,CEtUS在需要精确测量时提高准确性。
要点
- 129例CEA标本平均CPV为0.75±0.43 cm³(范围0.10-2.47 cm³)。
- tUS所测CPV与标本体积差值为0.13±0.24 cm³,CEtUS为-0.01±0.21 cm³,融合为-0.08±0.20 cm³。
- tUS测量CPV的观察者内和观察者间差异极小。
核心结论
- tUS可准确测量CPV,观察者内/间一致性极好。
- 如需精确CPV测量用于研究,CEtUS可提高准确性;仅tUS足以用于人群筛查。
研究设计与样本
- 连续纳入接受CEA的患者,术前即刻行tUS和CEtUS检查。
- 由受过培训的血管科学家用tUS、CEtUS及两者融合图像测量CPV。
- 用Archimedes技术测量内膜剥脱标本的精确体积作为参照。
主要发现
- 129例内膜剥脱标本的平均CPV为0.75±0.43 cm³(范围0.10-2.47 cm³)。
- tUS(n=114)所测平均CPV为0.87±0.51 cm³。
- CEtUS(n=104)所测平均CPV为0.75±0.45 cm³。
- 融合图像(n=95)所测平均CPV为0.83±0.49 cm³。
- 标本体积与tUS所测CPV差值为0.13±0.24 cm³,与CEtUS差值为-0.01±0.21 cm³,与融合图像差值为-0.08±0.20 cm³,均无临床意义。
临床怎么用
- 作者认为tUS可准确测量CPV,观察者内/间一致性极好。
- 作者认为如需精确CPV测量用于研究,CEtUS可提高准确性;仅tUS足以用于人群筛查。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Stenosis severity has been the indication for carotid endarterectomy (CEA) for 4 decades, but the annual stroke risk in asymptomatic carotid stenosis >70% is under 2%. Atherosclerotic volume has emerged as a risk factor for future stroke, but needs to be measured noninvasively. Tomographic ultrasound (tUS) is a novel technology that assembles 3D images in seconds. We evaluated accuracy of measuring Carotid Plaque Volume (CPV) with tUS in patients undergoing CEA. METHOD: Consecutive patients were imaged immediately before CEA by tUS and contrast-enhanced tUS (CEtUS). CPV was measured using tUS, CEtUS, and a fused images incorporating both tUS and CEtUS by trained vascular scientists. Precise volume of the endarterectomy specimen was measured using Archimedes technique. RESULTS: Mean ± sd (range) CPV in 129 endarterectomy specimens was 0.75 ± 0.43 cm3 (0.10-2.47 cm3). Mean ± sd CPV measured by tUS (n = 114) was 0.87 ± 0.51 cm3, CEtUS (n = 104) was 0.75 ± 0.45 cm3 and with fusion (n = 95) was 0.83 ± 0.49 cm3. Differences between specimen volume and CPV measured by tUS (0.13 ± 0.24 cm3), CEtUS (-0.01 ± 0.21 cm3) or fusion (-0.08 ± 0.20) were clinically insignificant. Intra-/interobserver differences were minimal. CONCLUSIONS: tUS accurately measures CPV with excellent intra-/interobserver agreement. CEtUS improves accuracy if precise CPV measurement is needed for research but tUS alone would be sufficient for population screening.