胸降主动脉远端严重迂曲影响胸主动脉腔内修复术中远端释放的准确性
该回顾性研究分析了2008年10月至2021年5月在本机构接受胸主动脉腔内修复术(TEVAR)的426例患者,其中56例尝试在腹腔干或肠系膜上动脉上方释放支架。基于大弯与直线比值(G/S比值)将患者分为严重迂曲组(n=21)和轻度迂曲组(n=35),比较手术和临床结局。结果显示严重迂曲与支架远端释放位置偏移和倾斜角度增大相关。
要点
- G/S比值是影响支架释放偏移的解剖因素,cutoff值为1.15。
- 严重迂曲组支架远端距目标血管更远(中位数10.0 mm vs 3.0 mm,p=0.015)。
- 严重迂曲组支架远端边缘倾斜角更大(中位数21.4度 vs 9.5度,p<0.01)。
核心结论
- 胸降主动脉远端严重迂曲与支架远端释放位置偏移和倾斜相关。
研究设计与样本
- 回顾性 reviewing 2008年10月至2021年5月在本机构接受TEVAR的426例患者的术前和术后CT扫描,其中56例尝试在腹腔干或肠系膜上动脉上方释放支架。
主要发现
- 支架释放偏移发生21例,所有解剖变量中仅G/S比值在偏移病例中显著更大(p=0.049)。
- 严重迂曲组支架远端距目标血管中位数10.0 mm(IQR 2.5-19.5)vs 轻度迂曲组3.0 mm(IQR 0-8.0),p=0.015。
- 严重迂曲组支架远端边缘倾斜角中位数21.4度(IQR 15.8-24.5)vs 轻度迂曲组9.5度(IQR 5.5-12.5),p<0.01。
- 两组原发性Ib型内漏发生率(p=0.454)、继发性Ib型内漏发生率(p=1.0)和远端再干预率(p=0.276)相当。
关键数值与推荐
- G/S比值cutoff值为1.15,通过受试者工作特征曲线分析确定。
临床怎么用
- 原文未据此给出推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
PURPOSE: An accurate distal deployment is essential for successful thoracic endovascular aortic repair (TEVAR) of a paradiaphragmatic aortic aneurysm. This study aimed to investigate the anatomical and intraoperative factors that affect the accuracy of distal deployment during TEVAR. METHODS: We conducted a retrospective review of preoperative and postoperative computed tomography scans of 426 patients undergoing TEVAR at our institution between October 2008 and May 2021, of which the stent-graft was attempted to be deployed just above the celiac axis or the superior mesenteric artery in 56 patients. Based on the anatomical factors related to the malposition (deployed >10 mm away from the target vessel) and the greater curve to the straight-line ratio (G/S ratio), the patients were categorized as severe tortuosity (n=21) and mild tortuosity (n=35) groups to compare the operative and clinical outcomes. RESULT: Stent-graft malpositioning occurred in 21 cases. Among all anatomical variables, only the G/S ratio was significantly larger in the malpositioned cases (p=0.049). A cutoff G/S ratio value of 1.15 was determined using the receiver operating curve analysis. In the severe tortuosity group, the distal end of the stent-graft was significantly farther (median: 10.0 [interquartile range (IQR): 2.5-19.5] mm vs 3.0 [0-8.0] mm; p=0.015) from the target vessel, and the tilt angle of the stent-graft's distal edge was larger (median: 21.4 [IQR: 15.8-24.5] vs 9.5 [5.5-12.5] degree; p<0.01) than that in the mild tortuosity group. Both groups were comparable for the incidence of a primary type Ib endoleak (p=0.454), a secondary type Ib endoleak (p=1.0), and the rate of distal reintervention (p=0.276). CONCLUSION: Severe tortuosity in the distal descending thoracic aorta is associated with a malpositioned and tilted distal end of the stent-graft. CLINICAL IMPACT: Thoracic endovascular aortic repair (TEVAR) for paradiaphragmatic thoracic aortic aneurysms requires accurate distal landing. In this paper, a retrospective CT analysis revealed that the greater curve to the straight-line ratio (G/S ratio) was associated to affects the malposition of the stent graft, defined as being deployed more than 10 mm away from the target vessel. Further, a comparative analysis based on the G/S ratio demonstrated that severe aortic tortuosity was associated with a more distal and tilted deployment of the stent graft.
引用
- Sato T, Banno H, Ikeda S, et al. Severe Tortuosity of the Distal Descending Thoracic Aorta Affects the Accuracy of Distal Deployment During a Thoracic Endovascular Aortic Repair. J Endovasc Ther. 2022. PMID: 36458820. doi:10.1177/15266028221141023