复杂血管腔内主动脉修复术中应用光纤实时形状导航的初期单中心经验
单中心前瞻性研究将2017至2022年复杂开窗/分支EVAR中FORS导航病例与61例非FORS病例1:3匹配,FORS导航技术成功率85%(81/95次插管),15次(16%)无需实时透视完成;FORS导航与更短中位手术时间、透视时间及更低剂量面积乘积和空气比释动能相关(每项P≤.0001)。
要点
- FORS导航在复杂EVAR中技术成功率为85%(81/95次插管),15次(16%)未使用实时透视。
- FORS技术成功的单因素预测因素包括靶动脉狭窄<50%、钙化<50%及所尝试的靶血管(每项P<.05)。
- FORS导航与更短中位手术时间和透视时间、更低剂量面积乘积及空气比释动能相关(每项P≤.0001)。
核心结论
- 作者初期FORS经验(含学习曲线)显示前景良好,技术成功率可接受,并减少手术时间和辐射用量。
研究设计与样本
- 单中心2017至2022年前瞻性研究复杂开窗和分支EVAR,FORS导航病例按纳入靶动脉和体重指数1:3回顾性匹配非FORS病例。
主要发现
- 21例FORS导航手术匹配61例非FORS手术,共尝试95次FORS插管(87次内脏靶动脉、8次分叉门)。
- 81次插管(85%)技术成功,15次(16%)全程未使用实时透视。
- FORS技术失败原因:器械材料特性6例、器械故障2例、导丝/导管组合6例。
- FORS技术成功单因素预测因素包括靶动脉狭窄<50%、靶动脉钙化<50%及所尝试的靶血管(每项P<.05)。
- FORS导航与更短中位手术时间和透视时间及更低剂量面积乘积和空气比释动能相关(每项P≤.0001)。
关键数值与推荐
- 技术成功定义为全程FORS导航(导丝置入至交换硬导丝)成功插管靶血管。
临床怎么用
- 原文未据此给出推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVE: In the present study, we have described the technical success using Fiber Optic RealShape (FORS) endovascular guidance and its effects on the overall procedural time and radiation usage during complex endovascular aortic repair (EVAR). METHODS: Fenestrated and branched EVARs performed at a single center from 2017 to 2022 were prospectively studied. FORS-guided procedures were matched retrospectively 1:3 to non-FORS-guided procedures by the incorporated target arteries and body mass index. Technical success was defined as successful target vessel cannulation using FORS for the entirety of navigation (wire insertion to exchange for a stiff wire). The predictors of technical success were evaluated via logistic regression. The procedural times and radiation doses were compared between the matched cohorts using the Wilcoxon rank sum test. RESULTS: A total of 21 FORS-guided procedures were matched to 61 non-FORS-guided procedures. A total of 95 FORS cannulations were attempted (87 for the visceral target artery and 8 for the bifurcate gate). Technical success was achieved in 81 cannulations (85%); 15 (16%) were completed without the use of live fluoroscopy. The univariate predictors of FORS technical success included <50% target artery stenosis, <50% target artery calcification, and the target vessel attempted (P < .05 for each). FORS failures were attributed to device material properties in six cases, device failure in two cases, and the wire/catheter combination in six. The use of FORS guidance was associated with shorter median procedural and fluoroscopy times and a lower dose area product and air kerma (P ≤ .0001 for each). CONCLUSIONS: The results from our initial experience with FORS during complex EVAR, including our learning curve, has shown promise, with acceptable technical success and reductions in procedural times and radiation usage.
引用
- Finnesgard EJ, Simons JP, Jones DW, et al. Initial single-center experience using Fiber Optic RealShape guidance in complex endovascular aortic repair. J Vasc Surg. 2022. PMID: 36384183. doi:10.1016/j.jvs.2022.11.041
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。