通过植入新型一体式外双分支覆膜支架完全腔内重建犬升主动脉、主动脉弓及弓上血管
研究评估新型一体式外双分支覆膜支架系统在犬体内重建升主动脉、主动脉弓及弓上血管的可行性与安全性。10只成年杂种犬全麻下经输送系统植入支架,随访3个月后CTA显示支架及自体血管通畅、内表面覆盖新生内膜、升主动脉无逆行性主动脉夹层。
要点
- 新型一体式外双分支覆膜支架系统可在犬体内成功植入升主动脉和主动脉弓。
- 平均手术时间142.7±13.7分钟,平均透视时间20.73±2.22分钟。
- 随访期间无脑栓塞症状和切口感染,CTA示支架及自体血管通畅。
核心结论
- 该动物研究证明一体式外双分支覆膜支架系统可用于重建犬升主动脉、主动脉弓及弓上血管。
研究设计与样本
- 动物实验:10只成年杂种犬全身麻醉下经输送系统将一体式外双分支覆膜支架植入升主动脉、主动脉弓及弓上血管,随访3个月。
主要发现
- 平均手术时间142.7±13.7分钟,平均透视时间20.73±2.22分钟,平均造影剂剂量95.9±8.7 mL。
- 术中三通道导管成功在主动脉内平行导丝,10枚分支支架全部成功植入犬升主动脉和主动脉弓。
- 随访期间无脑栓塞症状和切口感染。
- CTA和标本显示分支支架及自体血管通畅,支架内表面覆盖新生内膜,升主动脉无逆行性主动脉夹层。
临床怎么用
- 作者认为该一体式外双分支覆膜支架系统可应用于治疗累及升主动脉中远端、主动脉弓及近端降主动脉的主动脉病变。
局限性
- 原文未报告局限性
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
PURPOSE: We evaluated the feasibility and safety of using a new unibody outer double-branched stent-graft system to reconstruct the canine ascending aorta, aortic arch, and supra-aortic vessels. MATERIALS AND METHODS: The outer-branched stent-graft was a unibody design. The branched stent-graft consisted of a main stent-graft and 2 branches. The introducer system included a tri-channel catheter, 2 detachable sleeves, a front fixing device, a constraining wire, and a curved outer sheath. The branched stent-graft was loaded into the introducer system. Ten adult mongrel dogs underwent general anesthesia, and the branched stent-grafts were deployed into the canine ascending aorta, aortic arch, and supra-aortic vessels by the introducer system. All animals were followed up for 3 months. At the end of the follow-up period, computed tomographic angiography (CTA) was performed to observe the patency of the branched stent-grafts. RESULTS: The mean operation time was 142.7±13.7 minutes. The mean fluoroscopy time was 20.73±2.22 minutes. The mean dosage of contrast agent was 95.9±8.7 mL. During the operation, the tri-channel catheters successfully paralleled the wires in the aorta. All 10 branched stent-grafts were successfully implanted into the canine ascending aorta and aortic arch. There were no symptoms of cerebral embolization and no incision infection during the follow-up period. Computed tomographic angiography and specimens showed that the branched stent-grafts and native vessels were patent, the inner surfaces of the branched stent-grafts were covered by neointima, and there was no retrograde aortic dissection in the ascending aorta. CONCLUSIONS: This animal research demonstrated that the unibody outer double-branched stent-graft system could be applied to reconstruct the canine ascending aorta, aortic arch, and supra-aortic vessels. CLINICAL IMPACT: Thoracic endovascular aortic repair has been the main treatment method for aortic aneurysms or dissections involving the descending thoracic aorta. However, the aortic arch and ascending aorta remain the last segments of the aorta without a validated and routinely used endovascular option. In this research, we designed a new unibody outer branched stent-graft system to reconstruct the distal ascending aorta, aortic arch and supra-aortic vessels. The unibody outer branched stent-graft system could be applied to treat aortic pathologies which involve the middle and distal proximal ascending aorta, aortic arch and proximal descending aorta.
引用
- Li W, Li Q, Li J, et al. Complete Endovascular Reconstruction of the Canine Ascending Aorta, Aortic Arch, and Supra-Aortic Vessels by Implanting a New Unibody Outer Double-Branched Stent-Graft. J Endovasc Ther. 2022. PMID: 36444636. doi:10.1177/15266028221139194
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。