OptionELITE 腔静脉滤器置入术的临床与影像学结局
回顾性分析6家附属医院2013年1月1日至2019年4月19日间594例患者共603次OptionELITE下腔静脉滤器置入操作,602次技术成功(99.8%);191例尝试回收,189枚成功回收(99%),回收距置入4.3±6.3个月(中位2.5个月);随访期间新发或加重下腔静脉滤器远端深静脉血栓57例(9.5%),新发症状性非致死性肺栓塞19例(3.2%),症状性肺栓塞及滤器相关并发症发生率低。
要点
- 603次OptionELITE滤器置入中602次技术成功(99.8%)。
- 191例尝试回收中189枚成功回收(99%),置入至回收4.3±6.3个月。
- 随访中新发或加重滤器远端深静脉血栓57例(9.5%),新发症状性非致死性肺栓塞19例(3.2%)。
核心结论
- OptionELITE下腔静脉滤器置入后症状性肺栓塞及滤器相关并发症发生率低,尝试回收时成功率为99%。
研究设计与样本
- 回顾性分析6家附属医院2013年1月1日至2019年4月19日间置入OptionELITE下腔静脉滤器的594例患者共603次操作,数据来自放射信息系统、影像数据库和电子病历。
主要发现
- 603次操作中602次技术成功(99.8%);594例患者中356例表现为急性肺栓塞和/或深静脉血栓。
- 191例尝试回收,189枚滤器成功回收,回收距置入4.3±6.3个月(中位2.5个月)。
- 未行回收的414枚滤器随访14.8±17.7个月(中位6.9个月)。
- 随访期间57例(9.5%)发生新发或加重的滤器远端深静脉血栓,19例(3.2%)发生新发症状性非致死性肺栓塞。
- 1枚滤器移位并断裂,另1枚滤器断裂并丢失1根支撑杆;399例影像资料充分者中12例(3%)可疑穿孔、17例(4.3%)明确穿孔,13例(3.3%)穿透至周围结构;332例影像资料充分者中12例(3.6%)发生明显下腔静脉血栓,4例(1.2%)发生下腔静脉闭塞。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVE: We sought to determine the clinical and imaging outcomes after placement of the OptionELITE inferior vena cava (IVC) filter (Argon Medical Devices, Plano, TX). METHODS: The clinical characteristics and imaging findings of patients who had undergone OptionELITE filter placement at six affiliated hospitals between January 1, 2013 and April 19, 2019, were analyzed. Data were obtained from the Radiology Information System, imaging database, and electronic medical records. The patients were followed up until 1 month after filter removal, death, or the first clinic visit and imaging study after June 1, 2019. RESULTS: A total of 603 filter placement procedures in 594 patients were evaluated. Of the 603 procedures, 602 were technically successful (99.8%). Of the 594 patients, 356 had presented with acute pulmonary embolism and/or deep vein thrombosis. A total of 189 filters were retrieved from the 191 patients for whom retrieval had been attempted at 4.3 ± 6.3 months (median, 2.5 months) after placement. The 414 filters without removal procedures had been followed up for 14.8 ± 17.7 months (median, 6.9 months). During the follow-up period, 57 patients (9.5%) had developed new or worsening deep vein thrombosis caudal to the filter and 19 (3.2%) had experienced new symptomatic nonfatal pulmonary embolism after filter placement. One filter had migrated and fractured, and another filter had fractured and a strut was lost. Perforation was demonstrated equivocally in 12 (3%) and definitely in 17 (4.3%) of 399 patients with adequate imaging studies available, with penetration into surrounding structures in 13 patients (3.3%). Of 332 patients with adequate imaging studies, 12 (3.6%) had developed substantial IVC thrombosis and 4 (1.2%) had developed IVC occlusion. CONCLUSIONS: The rates of symptomatic pulmonary embolism and filter-related complications were low after OptionELITE IVC filter placement, and filter retrieval, when attempted, was 99% successful.