文献速递主动脉疾病

使用Cook Zenith Alpha腹主动脉移植物行腔内动脉瘤修复术后肢体移植物闭塞

Limb graft occlusion after endovascular aneurysm repair with the Cook Zenith Alpha abdominal graft.
PubMed 文献 · J Vasc Surg · 2022-10-26

单中心回顾性研究纳入241例使用低剖面Zenith Alpha腹主动脉移植物行腔内动脉瘤修复术的患者,术后3年每肢体肢体移植物闭塞累积发生率为7%(95%CI 5%-9%),共27例患者(11%)的33条肢体(7%)发生闭塞。说明书及既往文献提示的危险因素均未显示与肢体移植物闭塞正相关,而重度钙化的髂动脉显示保护性关联。

要点

  • 术后3年每肢体肢体移植物闭塞累积发生率为7%(95%CI 5%-9%)。
  • 说明书及既往文献提示的危险因素均未显示与肢体移植物闭塞正相关。
  • 重度钙化髂总动脉和髂外动脉较非钙化血管显示与肢体移植物闭塞降低相关。

核心结论

  • 使用Zenith Alpha移植物行腔内动脉瘤修复术后,每肢体肢体移植物闭塞累积发生率至术后3年为7%。
  • 说明书或当前文献提示的分析危险因素均未显示与肢体移植物闭塞正相关。

研究设计与样本

  • 单中心回顾性研究,纳入2015年10月1日至2018年9月30日接受低剖面Zenith Alpha移植物治疗的241例患者。
  • 所有计算机断层扫描血管造影使用三维软件分析,数据提取自电子病历直至研究期结束(2020年12月31日)。
  • 评估肢体移植物闭塞及闭塞相关再干预的累积发生率,并在特定时间点行回归分析评估可能危险因素。
  • 分析的危险因素包括主动脉和髂动脉直径、移植物组件 oversizing、髂动脉扭曲度和钙化、移植物组件重叠、同侧和对侧腿近端对齐以及髂外动脉密封区。
  • 在肢体移植物闭塞患者中评估髂动脉钙化和扭曲度的阅片者一致性。

主要发现

  • 27例患者(11%)的33条肢体(7%)发生闭塞。
  • 术后3年每肢体肢体移植物闭塞累积发生率为7%(95%CI 5%-9%)。
  • 既往描述的危险因素经回归分析未发现与肢体移植物闭塞正相关。
  • 术后3年重度钙化髂总动脉较非钙化血管与肢体移植物闭塞降低相关(风险降低17%[95%CI -27%至-7%];P=0.001)。
  • 术后3年重度钙化髂外动脉较非钙化血管与肢体移植物闭塞降低相关(风险降低15%[95%CI -26%至-5%];P=0.005)。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVE: Prior reports of the low profile Zenith Alpha abdominal graft (Cook Medical Inc, Bloomington, IN) have shown impaired limb graft patency to be the primary causes of reintervention. Special notices from the manufacturer have indicated certain instructions for use (IFU) violations as the main reasons for these complications. In the present study, we assessed the incidence of limb graft occlusion (LGO) and analyzed the effects of the detailed anatomic risk factors for LGO highlighted in the IFU and previously reported studies. METHODS: A retrospective study was performed of 241 patients treated with the low profile Zenith Alpha at a single institution from October 1, 2015 to September 30, 2018. All computed tomography angiograms were analyzed using three-dimensional software. Data were extracted from the electronic medical records until the end of the study period (December 31, 2020). The cumulative incidence of LGO and LGO-related reinterventions were assessed. A regression analysis was performed to evaluate the possible risk factors associated with the development of LGO at specified time points. These included aortic and iliac diameters, graft component oversizing, iliac tortuosity and calcification, overlap of graft components, proximal alignment of ipsilateral and contralateral legs, and sealing zone in the external iliac artery. Reader agreement of iliac calcification and tortuosity was assessed in patients with LGO. RESULTS: A total of 33 limbs (7%) in 27 patients (11%) had become occluded. The cumulative incidence of LGO was 7% (95% confidence interval [CI], 5%-9%) per limb up to 3 years postoperatively. The previously described risk factors for LGO were studied using regression analysis; however, no positive association with LGO was identified. Heavily calcified common iliac arteries (CIAs) and external iliac arteries were protective against LGO compared with noncalcified vessels up to 3 years postoperatively (decreased risk, 17% [95% CI, -27% to -7%]; P = .001; and 15% [95% CI, -26 to -5]; P = .005, respectively). The reader agreement of iliac calcification and tortuosity showed substantial agreement (CIA intrareader kappa = 0.75; CIA interreader kappa = 0.62) and almost perfect agreement (intrareader kappa = 0.85; interreader kappa = 0.84), respectively. CONCLUSIONS: The cumulative incidence of LGO after endovascular aneurysm repair with the Zenith Alpha graft was 7% per limb up to 3 years postoperatively. None of the analyzed risk factors suggested by the IFUs or current literature were positively associated with LGO.

来源:J Vasc Surg,查看原文。版权归原作者所有。

引用

  1. Broda M, Eiberg J, Taudorf M, et al. Limb graft occlusion after endovascular aneurysm repair with the Cook Zenith Alpha abdominal graft. J Vasc Surg. 2022. PMID: 36306934. doi:10.1016/j.jvs.2022.10.026

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。