文献速递主动脉疾病内脏/肾动脉与血管畸形

可转向电磁追踪导管在猪模型中的导航性能与影像融合的比较

A Steerable and Electromagnetically Tracked Catheter: Navigation Performance Compared With Image Fusion in a Swine Model.
PubMed 文献 · J Endovasc Ther · 2022-09-17

该临床前猪研究在3头猪中对双侧肾动脉、肠系膜上动脉和腹腔干尝试插管,7名术者先使用三维影像导航联合电磁追踪、再使用常规影像融合引导。两种方法各尝试72次插管,首次尝试成功率均为79%(57次),插管率无差异(p=1),操作时间相似;允许多次尝试时影像融合成功率达97%。结论认为电磁追踪导航性能与影像融合相似,需进一步在主动脉疾病患者中验证。

要点

  • 在3头猪中,电磁追踪与影像融合两种导航方法各尝试72次插管,首次尝试成功率均为79%(57次)。
  • 两种方法的插管率无差异(p=1),操作时间相似。
  • 允许多次尝试时,影像融合引导的插管成功率为97%。

核心结论

  • 本研究证明可转向电磁追踪导管联合三维影像导航具有可行性,电磁追踪的导航性能与影像融合相似,插管率和操作时间无统计学显著差异。
  • 需要在主动脉疾病患者中进一步研究以证明其效用。

研究设计与样本

  • 临床前猪研究,在3头猪中对双侧肾动脉、肠系膜上动脉和腹腔干使用7F可转向电磁追踪导管尝试插管。
  • 7名术者先使用三维影像导航联合电磁追踪、再使用常规影像融合引导进行插管。
  • 记录插管成功率、操作时间和辐射暴露;电磁追踪位置数据通过基于血管的配准算法注册到术前三维图像。
  • 由于缺乏电磁追踪导丝,需要超过1次尝试的插管仅使用影像融合进行。

主要发现

  • 两种方法各尝试72次插管,首次尝试成功率均为79%(57次)。
  • 插管率无差异(p=1),操作时间相似。
  • 允许多次尝试时,影像融合引导的插管成功率为97%。

临床怎么用

  • 作者认为电磁追踪联合新型可转向导管可减少辐射和对比剂剂量,同时在血管内主动脉手术中提供三维可视化和灵活导航。
  • 作者指出需进一步研究以在主动脉疾病患者中证明其效用。

局限性

  • 由于缺乏电磁追踪导丝,需要超过1次尝试的插管仅使用影像融合进行。
  • 需进一步研究以在主动脉疾病患者中证明其效用。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

PURPOSE: Cannulation of visceral vessels is necessary during fenestrated and branched endovascular aortic repair. In an attempt to reduce the associated radiation and contrast dose, an electromagnetically (EM) trackable and manually steerable catheter has been developed. The purpose of this preclinical swine study was to evaluate the cannulation performance and compare the cannulation performance using either EM tracking or image fusion as navigation tools. MATERIALS AND METHODS: Both renal arteries, the superior mesenteric artery, and the celiac trunk were attempted to be cannulated using a 7F steerable, EM trackable catheter in 3 pigs. Seven operators attempted cannulation using first 3-dimensional (3D) image navigation with EM tracking and then conventional image fusion guidance. The rate of successful cannulation was recorded, as well as procedure time and radiation exposure. Due to the lack of an EM trackable guidewire, cannulations that required more than 1 attempt were attempted only with image fusion. The EM tracking position data were registered to preoperative 3D images using a vessel-based registration algorithm. RESULTS: A total of 72 cannulations were attempted with both methods, and 79% (57) were successful on the first attempt for both techniques. There was no difference in cannulation rate (p=1), and time-use was similar. Successful cannulation with image fusion was achieved in 97% of cases when multiple attempts were allowed. CONCLUSION: This study demonstrated the feasibility of a steerable and EM trackable catheter with 3D image navigation. Navigation performance with EM tracking was similar to image fusion, without statistically significant differences in cannulation rates and procedure times. Further studies are needed to demonstrate this utility in patients with aortic disease. CLINICAL IMPACT: Electromagnetic tracking in combination with a novel steerable catheter reduces radiation and contrast media doses while providing three-dimensional visualization and agile navigation during endovascular aortic procedures.

来源:J Endovasc Ther,查看原文。版权归原作者所有。

引用

  1. Nypan E, Tangen GA, Brekken R, et al. A Steerable and Electromagnetically Tracked Catheter: Navigation Performance Compared With Image Fusion in a Swine Model. J Endovasc Ther. 2022. PMID: 36121010. doi:10.1177/15266028221123434

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。