文献速递主动脉疾病

Ovation内移植物植入术中「主动脉球囊塑形」扩展移植物用于不良瘤颈解剖的适用范围

"Aortic Balloon Molding" during Ovation Endograft Implantation Expands Graft Use for Hostile Neck Anatomy.
PubMed 文献 · Ann Vasc Surg · 2022-07-06

该回顾性研究纳入2019年3月至2020年12月在本机构接受Ovation内移植物行EVAR的43例患者,其中26例(60.5%)采用主动脉球囊塑形(ABM)技术。ABM组更多为反向锥形瘤颈、显著环形血栓或钙化及超出Ovation解剖适应证,但需颈相关辅助手术的比例更低(7.7% vs 23.5%,P=0.1),两组1年免于Ia型内漏或移位均为100%。作者认为ABM是安全有效的辅助技术,需进一步研究评估长期结果。

要点

  • ABM组26例(60.5%),更多为反向锥形瘤颈、显著环形血栓或钙化及超Ovation解剖适应证(均P=0.1)。
  • ABM组需颈相关辅助手术比例7.7% vs 非ABM组23.5%(P=0.1),技术成功率100%。
  • 完成造影仅非ABM组1例Ia型内漏,两组1年免于Ia型内漏或移位均为100%。

核心结论

  • ABM是使用Ovation支架移植物治疗不良瘤颈解剖AAA的安全有效辅助技术,可能有助于实现最佳尺寸选择以获得充分封闭。
  • 作者认为需进一步研究评估该技术的长期结果。

研究设计与样本

  • 回顾性研究,纳入2019年3月至2020年12月在本机构接受Ovation内移植物EVAR的全部患者43例。
  • ABM技术为移植物预置鞘后,在聚合物注入期间于密封环处充盈顺应性主动脉球囊。
  • 不良瘤颈解剖定义为成角>60°、反向锥形构型、≥50%环形血栓或钙化中的任何或全部。
  • 终点包括颈相关辅助手术、技术成功、完成造影Ia型内漏及1年免于Ia型内漏和移位。

主要发现

  • ABM组反向锥形瘤颈61.5% vs 非ABM组41.2%(P=0.1)。
  • ABM组显著环形血栓或钙化23.1% vs 非ABM组5.9%(P=0.1)。
  • ABM组超出Ovation解剖适应证65.4% vs 非ABM组41.2%(P=0.1)。
  • 技术成功100%,ABM组需颈相关辅助手术7.7% vs 非ABM组23.5%(P=0.1)。
  • 完成造影仅非ABM组1例Ia型内漏,1年免于Ia型内漏或移位两组均为100%。

关键数值与推荐

  • 不良瘤颈解剖标准:成角>60°、反向锥形构型、≥50%环形血栓或钙化中的任何或全部。
  • ABM组反向锥形瘤颈61.5% vs 非ABM组41.2%(P=0.1)。
  • ABM组显著环形血栓或钙化23.1% vs 非ABM组5.9%(P=0.1)。
  • ABM组超出Ovation解剖适应证65.4% vs 非ABM组41.2%(P=0.1)。
  • ABM组需颈相关辅助手术7.7% vs 非ABM组23.5%(P=0.1)。

临床怎么用

  • 作者认为ABM是使用Ovation支架移植物治疗不良瘤颈解剖AAA的安全有效辅助技术。
  • 作者认为ABM可能允许最佳内移植物尺寸选择以在复杂主动脉解剖中实现充分封闭。
  • 作者认为需进一步研究评估该技术的长期结果。

局限性

  • 原文为回顾性研究。
  • 样本量43例,平均随访7.9±6个月。
  • 作者认为需进一步研究评估该技术的长期结果。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Challenging aortoiliac anatomy such as short neck and narrow access vessels is responsible for endovascular repair of abdominal aortic aneurysm (EVAR) ineligibility in up to 50% of cases. The Ovation stent graft helped widen the range of abdominal aortic aneurysms (AAAs) suitable for EVAR thanks to its low-profile delivery system and polymer-filled sealing rings. However, its advantages are offset by a tight sizing chart that can lead to increased risk of type Ia endoleak or endograft infolding from under- or oversizing, respectively. We sought to assess the safety and efficacy of a novel endovascular technique developed to expand the use of the Ovation endograft while avoiding these issues. METHODS: We conducted a retrospective review of all patients who underwent EVAR with the Ovation endograft at our institution between March 2019 and December 2020. "Aortic Balloon Molding" or ABM is a novel endovascular technique in which the graft is pre-cannulated and a compliant aortic balloon is inflated at the site of the graft's sealing rings during polymer administration. The technique was preferentially performed in patients with hostile neck anatomy (HNA) defined as any or all of angulation >60°, reverse taper configuration, ≥50% circumferential thrombus, or calcification. Patients undergoing traditional deployment were compared to those in whom ABM was performed. End points included neck-related adjunctive procedures, technical success, type Ia endoleak at completion angiogram, and 1-year freedom from type Ia endoleak and migration. RESULTS: A total of 43 patients were included in the study, of which 26 (60.5%) were treated with the ABM technique. Mean follow-up was 7.9 ± 6 months. Patients in the ABM group were more likely to have a reverse taper neck (61.5% vs. 41.2%, P = 0.1), have significant circumferential thrombus or calcium (23.1% vs. 5.9%, P = 0.1), and be treated outside of the Ovation indications for use regarding anatomic characteristics (65.4% vs. 41.2%, P = 0.1). Technical success was achieved in 100% of cases. However, patients in the ABM group were less likely to require a neck-related adjunctive procedure (7.7% vs. 23.5%, P = 0.1). Only 1 type Ia endoleak was observed at completion angiogram in a patient treated without ABM. At 1 year, freedom from type Ia endoleak or migration was 100% for both groups. CONCLUSIONS: ABM proves to be a safe and effective adjunctive technique for the treatment of AAAs with HNA using the Ovation stent graft. This may allow optimal endograft sizing to achieve adequate seal in complex aortic anatomies. Further research is warranted to evaluate the long-term outcomes of this technique.

来源:Ann Vasc Surg,查看原文。版权归原作者所有。

引用

  1. Jensen R, Mathlouthi A, Al-Nouri O, et al. "Aortic Balloon Molding" during Ovation Endograft Implantation Expands Graft Use for Hostile Neck Anatomy. Ann Vasc Surg. 2022. PMID: 35803461. doi:10.1016/j.avsg.2022.05.042

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。