无约束三开窗主动脉弓腔内移植物精准释放系统

A System for Accurate Deployment of Unconstrained Triple-Fenestrated Aortic Arch Endografts.
PubMed 文献 · J Endovasc Ther · 2022-01-12

该研究描述了一种在全主动脉弓腔内修复术中精准旋转定向和轴向定位无约束三开窗医师改良移植物释放的系统,使用Valiant Captivia移植物在术前CT血管造影指导下制作不透射线边缘的开窗,并在31例连续患者中成功应用,所有弓部分支在释放后即刻获得系统血压且成功置入支架。

要点

  • 通过8字形不透射线标记与工作体位确保移植物旋转定向正确。
  • 左颈总动脉开窗定位在股-左颈总动脉导丝环近端以确定轴向位置。
  • 31例连续患者中所有弓部分支均成功插管并置入支架。

核心结论

  • 该系统可在全主动脉弓腔内修复术中简单可靠地实现无约束三开窗弓部移植物的精准释放。

研究设计与样本

  • 研究描述了基于患者弓部解剖重建CT血管造影图像定制开窗的释放技术,在31例连续接受全主动脉弓腔内修复术的患者中应用。

主要发现

  • 该技术在31例连续患者中成功使用,移植物释放后即刻所有弓部分支均获得系统血压,表明血流充足。
  • 所有弓部分支均成功插管并置入支架。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

PURPOSE: To achieve accurate rotational orientation and the axial position of unconstrained triple-fenestrated physician-modified endografts upon deployment in the aortic arch during total arch thoracic endovascular aortic repair (TA-TEVAR). MATERIALS AND METHODS: Following a detailed study of reconstructed computerized tomography angiography images of patients' arch anatomy, customized, sealable fenestrations with radio-opaque margins are created onsite on Valiant Captivia (Medtronic) endografts, transposing the arch branch ostial anatomic interrelationship onto the endograft precisely. Radio-opaque figure-of-8 markers, indicating the 12 o'clock (superior) position, are attached to the endograft on the surface and brought up to the surface under the endograft cover during resheathing. Resheathing without any twist in the endograft is achieved by lining up the welds in each endograft stent segment in a straight line. The fluoroscopic working view for arch endograft delivery and deployment is the left anterior oblique view that is orthogonal to the plane of the arch, which, in turn, is the right anterior oblique view in which parts of a stiff indwelling guidewire in the ascending and descending aorta precisely overlap. During introduction in the working view, the endograft delivery system is rotated in the descending thoracic aorta so that the 12 o'clock figure-of-8 markers are viewed on the edge and situated at the outer aortic curvature; continued advancement into the arch without any further rotation will ensure superior orientation of the figure-of-8 markers and, consequently, correct endograft rotational orientation. Proper axial endograft positioning requires locating the left common carotid artery (LCCA) fenestration just proximal to a taut externalized LCCA-femoral guidewire loop marking the posterior limit of the LCCA ostium. After endograft deployment during rapid cardiac pacing, the target arch branches are cannulated through their respective fenestrations using hydrophilic 0.035-inch guidewires that are externalized via distal sheaths to create femoral-arch branch (through-and-through) loops over which covered fenestrated stents are introduced and deployed. RESULTS: This technique was used successfully in 31 consecutive patients undergoing TA-TEVAR; systemic blood pressure was obtained in all arch branches immediately after endograft deployment, indicating adequate blood flow. All arch branches were successfully cannulated and stented. CONCLUSION: This system enables accurate deployment of unconstrained triple-fenestrated arch endografts simply and reliably during TA-TEVAR.

来源:J Endovasc Ther,查看原文。版权归原作者所有。