影响多分支成品支架治疗胸腹主动脉瘤内脏适用性的因素

Factors Affecting the Visceral Suitability of a Multibranched off-the-Shelf Endograft for the Treatment of Thoraco-Abdominal Aortic Aneurysms.
PubMed 文献 · Ann Vasc Surg · 2021-11-04

回顾性分析2015年1月至2019年12月54例胸腹主动脉瘤患者的CTA,评估Zenith T-branch在内脏区的解剖适用性。33.3%的患者因腹腔干与肠系膜上动脉共干或内脏/肾动脉直径受限而不适用;支架居中放置时分支长度标准满足率97.9%,偏心放置降至93.7%。约67%患者内脏区适用,目标动脉直径是最主要限制因素。

要点

  • 54例胸腹主动脉瘤CTA中约67%内脏区适用Zenith T-branch支架。
  • 33.3%患者因内脏/肾动脉直径或腹腔干与肠系膜上动脉共干受限。
  • 支架偏心放置使分支长度标准适用率由97.9%降至93.7%。

核心结论

  • Zenith T-branch系统在内脏区约67%的胸腹主动脉瘤患者中适用,目标动脉直径是最主要的限制标准。
  • 支架在主动脉腔内的居中位置显著影响分支长度距离标准。

研究设计与样本

  • 回顾性分析2015年1月至2019年12月54例胸腹主动脉瘤患者的计算机断层扫描血管造影(CTA),通过多平面重建评估Zenith T-branch在内脏区的解剖适用性。

主要发现

  • 33.3%的患者因肠系膜上动脉与腹腔干共干或1条及以上内脏/肾动脉直径受限而不适用。
  • 当支架置于肠系膜上动脉分支偏离角为向左5度的位置时,除1例(97.9%)外所有患者满足分支偏离角标准。
  • 当支架置于主动脉中心时,除1例(97.9%)外所有患者满足分支长度标准;偏心放置使适用性降至93.7%。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: To assess the factors affecting visceral suitability of the use of the Zenith T-branchTM system in a group of patients with thoraco-adbominal aortic aneurysms (TAAAs). METHODS: Computer tomography angiography (CTA) of patients who presented a TAAA from 01/2015 to 12/2019 were retrospectively examined. Multi-Planar Reconstructions were performed on CTA images to assess the anatomic suitability of the Zenith T-branch in the visceral district. In particular, the branch deviation angle (BDA), and the branch-length were computed for each target vessel. RESULTS: Fifty-four CTA were examined. In 33.3% of these patients the presence of either a common origin of the superior mesenteric artery and the celiac trunk, or the diameter of 1 or more visceral/renal artery limited the visceral suitability of the device. All patients except 1 (97.9%) fitted the BDA criterion when the graft was placed in a position in which the BDA for the SMA was 5 degrees to the left. The branch-length criteria was met in all patients, except for 1 (97.9%), when the graft was placed in the center of the aorta. The eccentrical placement of the endograft decreased the suitability to 93.7%. CONCLUSIONS: The Zenith T-branch system can be suitable in the visceral district for about 67% of patients. The target artery diameter was the most limiting criterion. The central location of the graft within the aortic lumen significantly affected the branch-length distance criteria.