二氧化碳自动血管造影在血管内主动脉修复中较碘对比剂更能保护肾功能:技术与解剖细节分析
该研究回顾2016至2019年321例血管内主动脉修复(EVAR),其中72例(22.4%)使用二氧化碳(CO2)自动注射造影,249例(77.6%)使用碘对比剂(ICM)。CO2-EVAR组ICM用量显著更低(52.8 ± 6.1 vs. 88.1 ± 9.2 cc,P < 0.001),术后eGFR下降显著更少(2.3 ± 1.1 vs. 10.6 ± 5.3 mL/min,P < 0.001),但辐射剂量显著更高(总DAP 500,550.8 ± 377,394.6 vs. 332,301.8 ± 230,139.3 mGy/cm2,P = 0.001)。
要点
- CO2-EVAR组ICM用量显著低于ICM-EVAR组(52.8 ± 6.1 vs. 88.1 ± 9.2 cc,P < 0.001)。
- CO2-EVAR组术后eGFR下降显著少于ICM-EVAR组(2.3 ± 1.1 vs. 10.6 ± 5.3 mL/min,P < 0.001)。
- 近端瘤颈长度>24.5 mm可预测最低肾动脉显影(敏感性72.1%,特异性73.8%)。
核心结论
- CO2-EVAR自动注射是安全的,与ICM-EVAR相比需要更少的ICM用量,并对术后肾功能有显著获益。
- 在特定解剖情况下,ICM可能完全不需要。
- 然而辐射剂量显著更高,因此操作方案需要进一步改进。
研究设计与样本
- 回顾性分析2016年至2019年期间所有使用CO2或ICM进行的EVAR手术,共321例,其中CO2-EVAR 72例(22.4%),ICM-EVAR 249例(77.6%)。
- CO2-EVAR使用自动注射器(600 mm Hg压力;100 cc容量);在最低肾动脉显影困难或完成造影有疑问时注射少量ICM。
- 考虑临床和CT扫描术前特征,研究终点为技术成功、ICM用量和辐射剂量、术后肾功能及可能的CO2相关不良事件。
- 统计分析采用Fisher精确检验、t-Student检验、Mann-Whitney检验和ROC曲线。
主要发现
- 两组临床特征和术前肾功能相似。
- CO2-EVAR组ICM用量显著低于ICM-EVAR组(52.8 ± 6.1 vs. 88.1 ± 9.2 cc,P < 0.001)。
- CO2-EVAR组平均辐射剂量显著高于ICM-EVAR组(总DAP:500,550.8 ± 377,394.6 mGy/cm2 vs. 332,301.8 ± 230,139.3 mGy/cm2,P = 0.001)。
- CO2-EVAR组术后eGFR下降显著少于ICM-EVAR组(2.3 ± 1.1 vs. 10.6 ± 5.3 mL/min,P < 0.001)。
- CO2-EVAR中最低肾动脉正确显影50/72例(69.4%),该组近端瘤颈显著更长(中位数[IQR]:30 [14] vs. 18 [15] mm,P = 0.001)。
关键数值与推荐
- ROC曲线显示,近端瘤颈长度>24.5 mm可预测最低肾动脉显影(敏感性72.1%,特异性73.8%)。
- CO2-EVAR中有3例出现术中一过性低血压,无后果。
- 72例CO2-EVAR中16例(22.2%)使用0 cc ICM。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Contrast induced nephropathy occurs in up to 7.5% of cases in endovascular aortic repair (EVAR). Carbon dioxide (CO2) has been proposed as an alternative agent to iodinated contrast medium (ICM); however, specific protocols are not universally adopted, and the visualization of the renal arteries may be suboptimal in some cases. The aim of this study was to analyze our CO2-EVAR experience with automatic injections, in order to identify the anatomical characteristics associated with the best visualization of all the aortic vessels, with particular attention to the lowest renal artery (LoRA). METHODS: From 2016 to 2019, all EVAR performed with either CO2 or ICM were analyzed and compared. CO2-EVAR was performed using an automated injector (600 mm Hg pressure; 100 cc volume); a small amount of ICM was injected in case of difficulty in LoRA visualization or doubts at the completion angiogram. Clinical and CT-Scan preoperative characteristics were considered. The study endpoints were technical success, amount of ICM and radiation dose, postoperative renal function and possible CO2-related adverse events. Statistical analysis was by Fisher's exact, t-Student, Mann-Whitney tests and ROC curve. RESULTS: In the considered period, 321 EVAR procedures, 72 (22.4%) with CO2 and 249 (77.6%) with ICM, were performed. The 2 groups were similar for clinical characteristics and preoperative renal function. ICM was injected in a significantly lower amount in the CO2-EVAR group (52.8 ± 6.1 vs. 88.1 ±9.2 cc, P < 0.001), which received a significantly higher mean radiation dose (Total DAP: 500,550.8 ± 377,394.6 mGy/cm2 CO2-EVAR vs. 332,301.8 ±230,139.3 mGy/cm2 ICM-EVAR, P = 0.001). Postoperative eGFR decreased significantly less in the CO2-EVAR (2.3 ± 1.1 mL/min) compared with the ICM-EVAR group (10.6 ±5.3 mL/min), P < 0.001. LoRA was correctly visualized in 50/72 (69.4%) cases of CO2-EVAR, which had a significantly longer proximal neck (Median [IQR]: 30 [14] vs. 18 [15] mm, P = 0.001). At ROC curve, a proximal neck length >24.5 mm was predictive of LoRA visualization (72.1% sensitivity, 73.8% specificity). Three CO2-EVAR cases had intraoperative transient hypotension with no consequences. Sixteen/72 (22.2%) CO2-EVAR procedures were performed using 0 cc of ICM. CONCLUSIONS: CO2-EVAR by automated injections is safe and requires a lower amount of ICM if compared with ICM-EVAR, with a consequent significant benefit on postoperative renal function. If specific anatomical situations are present, ICM may be completely unnecessary. The radiation dose is however significantly higher, therefore procedural protocols need further refinements.