StoneSense自动目标识别在铥光纤激光碎石术中的多模型体外评估
该体外研究评估了集成于铥光纤激光系统的StoneSense自动目标识别(ATR)功能,发现ATR将激光激发限制在约2.3-2.7 mm的结石-光纤距离内,在持续灌流模型中减少激光激活时间与能量输出、提高消融效率并降低温升,但延长总手术时间。
要点
- ATR将激光激发限制在约2.3-2.7 mm结石-光纤距离,不同LED强度或频率间无显著差异。
- 在灌流模型中,ATR减少激光激活时间与能量输出,提高消融效率并降低温升。
- 模拟RIRS模型中ATR增加消融质量与效率,但延长手术时间。
核心结论
- StoneSense ATR将TFL激发限制在特定光学范围内,体外提高消融效率并减少能量浪费和温升,但延长手术时间。
研究设计与样本
- 采用BegoStone体模的体外研究,包含三个实验设置:Z轴平台测定最大允许激发距离、灌流比色皿模型和单术者模拟RIRS模型。
主要发现
- ATR将激光激发限制在约2.3-2.7 mm结石-光纤距离,不同LED强度(50% vs 100%)或脉冲频率(1 J, 10 Hz vs 20 Hz)间无统计学显著差异。
- 灌流模型中,ATR激活组较未激活组激光激活时间减少(均值[SD] 146[18] vs 179[19] s;P=0.020),能量输出减少(1.46[0.18] vs 1.79[0.20] kJ;P=0.021)。
- 灌流模型中,ATR激活组消融效率提高(1.77[0.25] vs 1.42[0.13] mg/J;P=0.022),温升降低(4.2[0.3] vs 5.1[0.5]°C;P=0.001),总手术时间轻度增加。
- 模拟RIRS模型中,ATR增加消融质量和效率(P<0.05),手术时间延长。
临床怎么用
- ATR引导碎石可能提高精准度和热安全性,但需权衡手术时间延长。
局限性
- 研究为体外模型,结论需临床进一步验证。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
Objectives To determine the maximal stone-to-fibre firing distance permitted by StoneSense™ (IPG Medical, Marlborough, MA, USA) automatic target recognition (ATR) integrated into a thulium fibre laser (TFL) system, and to evaluate its effects on ablation efficiency, energy delivery, and thermal dispersion in complementary in vitro models. Materials and Methods An IPG Medical TFL prototype with integrated ATR was evaluated using BegoStone (Bego GmbH & Co. KG, Bremen, Germany) phantoms in three experimental settings. First, maximal ATR-permitted firing distance was determined using a stabilised Z-stage under varying light-emitting diode (LED) intensities (50% vs 100%) and laser frequencies (1 J, 10 Hz vs 20 Hz). Second, in a cuvette-mesh model with continuous irrigation and thermocouple monitoring, lithotripsy was performed with ATR activated and inactivated to assess active laser time, energy delivery, ablation efficiency, procedural duration, and temperature change. Third, a single-surgeon simulated retrograde intrarenal surgery (RIRS) model evaluated fragmentation over 120 s of laser emission. Results The ATR restricted laser firing to a narrow stone-to-fibre distance of approximately 2.3–2.7 mm, with no statistically significant differences observed between the tested LED intensities or pulse frequencies. In the cuvette model, ATR activation reduced active laser time (mean [SD] 146 [18] vs 179 [19] s; P = 0.020) and energy delivery (mean [SD] 1.46 [0.18] vs 1.79 [0.20] kJ; P = 0.021), improved ablation efficiency (mean [SD] 1.77 [0.25] vs 1.42 [0.13] mg/J; P = 0.022), and reduced thermal rise (mean [SD] 4.2 [0.3] vs 5.1 [0.5]°C; P = 0.001), with a modest increase in total procedure time. In the simulated RIRS model, ATR increased ablated mass and efficiency (P < 0.05), with longer procedural duration. Conclusions The StoneSense ATR constrains TFL activation to a defined optical range and improves ablation efficiency while reducing energy waste and thermal rise in vitro. Although associated with increased procedural duration, ATR-guided lithotripsy may enhance precision and thermal safety.
引用
- BJU International:Evaluation of StoneSense™ automatic target recognition in thulium fibre laser lithotripsy: a multi‐model in vitro study
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。