可调节经闭孔男性吊带系统(ATOMS)感染:致病微生物与临床特征
多中心回顾性研究纳入902例接受ATOMS治疗的患者,其中24例(2.7%)因感染取出装置。感染多由革兰阴性菌(50%)和革兰阳性球菌(37.5%)引起,最常见为肠球菌和变形杆菌(各16.7%)。尽管感染,70.8%患者对植入满意,75%接受了二次装置。
要点
- ATOMS感染率为2.7%,肠球菌和变形杆菌最常见。
- 29.2%感染发生于术后3个月内,25%伴阴囊端口侵蚀。
- 感染后70.8%满意,75%可再植入装置。
核心结论
- ATOMS感染是令人困扰的并发症,部分与阴囊端口侵蚀和/或实质性尿路感染相关。
- 肠球菌和变形杆菌是最常见的致病微生物,可能影响手术预防用药选择。
研究设计与样本
- 多中心回顾性研究,评估9家学术机构902例ATOMS治疗患者中因感染取出装置的病例。
- 分析临床和微生物学数据。
主要发现
- 24例(2.7%)出现感染,中位年龄73±7岁,从植入到取出的中位间隔11±26.5个月。
- 7例(29.2%)在术后3个月内诊断感染,6例(25%)存在阴囊端口侵蚀,2例(8.3%)有睾丸实质性感染全身症状。
- 假体周围液体培养阳性20例(83.3%):革兰阴性菌12例(50%),革兰阳性球菌9例(37.5%),酵母菌1例(4.2%)。
- 最常见分离菌为肠球菌和变形杆菌(各16.7%),其次为假单胞菌和表皮葡萄球菌(各12.5%),耐甲氧西林金黄色葡萄球菌仅1例(4.2%)。
- 17例(70.8%)对植入满意,18例(75%)在取出后中位9.7±12.6个月接受二次装置(11例重复ATOMS,7例人工尿道括约肌)。
临床怎么用
- ATOMS感染时需考虑肠球菌和变形杆菌覆盖,可能影响手术预防抗生素选择。
- 感染后部分患者仍可满意并接受二次装置植入。
局限性
- 回顾性设计,且非感染原因取出的ATOMS缺乏微生物培养数据。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVE: To evaluate the clinical profile and the organisms producing adjustable transobturator male system (ATOMS) infection in a contemporary series. METHODS: Multicenter retrospective study evaluating patients undergoing ATOMS explant for clinical signs of infection from a series of 902 patients treated in 9 academic institutions. Clinical and microbiological data were evaluated. RESULTS: Infection presented in 24 patients (2.7%). The median age was 73 ± 7yrs and the median interval from ATOMS implantation to explant 11 ± 26.5mo. Infection was diagnosed within 3-months after surgery in 7(29.2%). Scrotal port erosion was present in 6 cases (25%) and systemic symptoms of parenchymatous testicular infection in 2(8.3%). The culture of the periprosthetic fluid was positive in 20(83.3%): 12(50%) Gram-negative bacteria, 9(37.5%) Gram-positive cocci and 1(4.2%) yeast. The most frequent isolates were Enterococcus and Proteus sp. (16.7% each), followed by Pseudomona sp. and S. epidermidis (12.5% each). Methicillin resistant S. aureus was detected only in 1 case (4.2%). Despite the infection 17 patients (70.8%) were satisfied with the implant and 18(75%) received a second device (11 repeated ATOMS and 7 AUS) at a median 9.7 ± 12.6mo after explant. Limitations include retrospective design and lack of microbiological cultures in ATOMS explanted for non-infective cause. CONCLUSION: Infection of a prosthetic device is a disturbing complication. A proportion of patients with ATOMS infection is associated to scrotal port erosion and/or parenchymatous urinary tract infection. Enterococcus and Proteus sp. are the most common organisms producing ATOMS infection and this could have implications for the selection of the most appropriate surgical prophylaxis.