文献速递肾癌与肾肿瘤

微创肾切除术中血管吻合器并发症:FDA MAUDE数据库2009-2019年更新综述

Endovascular Stapler Complications During Minimally Invasive Nephrectomy: An Updated Review of the FDA MAUDE Database From 2009-2019.
PubMed 文献 · Urology · 2021-02-15

该研究回顾FDA MAUDE数据库2009年1月至2019年8月数据,分析微创根治性肾切除术中肾门结扎时血管吻合器相关并发症。共383例并发症,63%涉及Ethicon endocutters,28%为GIA,9%为TA;22例死亡(占并发症5.7%)归因于吻合器。TA吻合器无死亡或再手术,且中转开放发生率较低。

要点

  • TA吻合器无死亡或再手术,可能与切割和钉合分步进行有关。
  • Ethicon endocutters占并发症63%,GIA占28%,TA占9%。
  • 22例死亡(5.7%)归因于吻合器,制造商评估多归因于使用不当。

核心结论

  • 微创根治性肾切除术中血管吻合器相关并发症以Ethicon endocutters为主,TA吻合器无死亡或再手术且中转开放发生率较低。

研究设计与样本

  • 回顾性分析FDA MAUDE数据库2009年1月1日至2019年8月1日微创根治性肾切除术中肾门结扎吻合器并发症,按Ethicon endocutters、Endo-GIA和Endo-TA分类。

主要发现

  • 共383例吻合器相关并发症,63%涉及Ethicon endocutters,28%涉及GIA,9%涉及TA。
  • 22例死亡(占总并发症5.7%)归因于吻合器。
  • TA吻合器无死亡或再手术,且中转开放发生率低于Ethicon和GIA吻合器。
  • 除一个器械外,制造商对退回器械的评估未发现异常或归因于吻合器使用不当。

临床怎么用

  • 基于制造商评估将吻合器故障归因于人为错误,应对手术室人员进行正确使用这些器械的培训以预防潜在严重并发症。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVE: To characterize the rates of endovascular stapler complications during hilar ligation in minimally invasive radical nephrectomy over the last 10 years. MATERIAL AND METHODS: We reviewed the Food and Drug Administration Manufacturer and User Facility Device Experience database from January 1, 2009 to August 1, 2019. Staplers were categorized according to type, namely Ethicon Inc. endocutters (Johnson & Johnson, New Brunswick NJ); Endo-GIA (Medtronic, Minneapolis MN); and Endo-TA (Medtronic, Minneapolis MN). RESULTS: There were 383 cases of complications involving staplers, 63% with Ethicon endocutters; 28% with GIA; and 9% with TA. 22 deaths (5.7% of total complications) were attributed to staplers. No deaths or reoperations occurred due to TA staplers. TA staplers were also associated with a reduced incidence of conversion to open as compared to Ethicon and GIA staplers. Apart from one device, manufacturer evaluation of returned devices either showed no abnormalities or attributed fault to improper use of staplers. DISCUSSION: We characterized stapler complications during a 10-year period for minimally invasive radical nephrectomy. No deaths or reoperations occurred due to TA staplers, perhaps due to cutting and stapling occurring in separate steps. Based on manufacturer evaluation attributing stapling malfunctions to human errors, training of operating room staff on proper use of these devices is critical to prevent potentially significant complications from occurring.

来源:Urology,查看原文。版权归原作者所有。

引用

  1. Gopal N, Long B, Phillips J, et al. Endovascular Stapler Complications During Minimally Invasive Nephrectomy: An Updated Review of the FDA MAUDE Database From 2009-2019. Urology. 2021. PMID: 33600834. doi:10.1016/j.urology.2021.02.010

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。