文献速递泌尿系结石

北美首例系列:高功率超脉冲铥光纤激光输尿管镜治疗尿石症

Ureteroscopic Performance of High Power Super Pulse Thulium Fiber Laser for the Treatment of Urolithiasis: Results of the First Case Series in North America.
PubMed 文献 · Urology · 2021-02-20

常用粉末化技术(67.1%),脉冲频率最高达2400 Hz,平均手术时间59.4±31.5分钟,未见与激光使用相关的特异性并发症。

要点

  • 高功率超脉冲铥光纤激光可用于输尿管镜碎石,脉冲频率最高2400 Hz。
  • 粉末化技术使用率67.1%,平均激光时间10.8±14.1分钟。
  • 术中并发症以1级输尿管损伤为主,未见激光相关特异性并发症。

核心结论

  • 高功率超脉冲铥光纤激光在北美首例系列中治疗尿石症结果 promising。
  • 亚200微米光纤和高达2400 Hz的粉末化设置可成功使用。
  • 未见与激光使用相关的特异性并发症。

研究设计与样本

  • 多中心回顾性病历 review,经机构审查委员会批准。
  • 纳入2019年10月至2020年3月接受高功率超脉冲铥光纤激光治疗的患者。

主要发现

  • 结石位置:输尿管内32枚,下极49枚,中或上极37枚。
  • 粉末化技术常用(67.1%),脉冲频率最高达2400 Hz。
  • 平均手术时间59.4±31.5分钟;平均激光时间10.8±14.1分钟;总激光能量12.5±19.1 KJ。
  • 术中并发症:7例1级、3例2级、1例3级输尿管损伤,以及1例肾集合系统出血经激光凝固设置(1J-20Hz)充分处理。

临床怎么用

  • 可考虑使用高功率超脉冲铥光纤激光行输尿管镜碎石,采用亚200微米光纤和粉末化设置。
  • 术中需注意输尿管损伤和集合系统出血,可用激光凝固设置处理出血。

局限性

  • 回顾性多中心病历 review 设计。
  • 缺乏对照组或长期随访数据。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVE: To report the first case series of ureterorenoscopy in North America using the High Power Super Pulse Thulium Fiber Laser for the treatment of upper urinary tract stones. METHODS: After Institutional Review Board approval, a multicentric retrospective chart review of patients treated with the High Power Super Pulse Thulium Fiber Laser from October 2019 to March 2020 was conducted. Basic demographic information, pre-operative, and peri-operative data were recorded. RESULTS: Seventy-six patients were included with a mean age of 60.9 ± 13.3 years. 118 stones were treated including 32 within the ureter, 49 in the lower pole, 37 in mid or upper poles. Dusting technique was commonly used (67.1%) with pulse frequencies up to 2400 Hz. Mean operative time was 59.4 ± 31.5 minutes. Mean laser time and total laser energy were 10.8 ± 14.1 minutes and 12.5 ± 19.1 KJ, respectively. Intraoperative complications were limited to 7 grade 1, 3 grade 2, and 1 grade 3 ureteral injuries and one case of renal collecting system bleeding that was adequately managed with laser coagulation settings (1J-20Hz). CONCLUSION: This initial case series in North America of the High Power Super Pulse Thulium Fiber Laser is promising for the treatment of urolithiasis. Sub-200 μm fibers and dusting settings up to 2400 Hz were utilized successfully. No specific complications related to use of the laser were seen.

来源:Urology,查看原文。版权归原作者所有。

引用

  1. Carrera RV, Randall JH, Garcia-Gil M, et al. Ureteroscopic Performance of High Power Super Pulse Thulium Fiber Laser for the Treatment of Urolithiasis: Results of the First Case Series in North America. Urology. 2021. PMID: 33621583. doi:10.1016/j.urology.2020.12.054

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。