单孔达芬奇机器人标准哈巴狗夹肾门阻断技术
本文介绍在达芬奇单孔机器人系统下使用标准机器人哈巴狗夹进行肾门阻断和移除的可行技术。单孔机械臂因器械路径角度导致最大力低于装载和打开标准哈巴狗夹所需,通过控制性步骤序列可安全高效完成。
要点
- 单孔机器人机械臂力不足,无法直接装载和打开标准哈巴狗夹。
- 采用控制性步骤序列可完成夹的体内装载、打开与放置。
- 该技术基于两例部分肾切除术视频,可行且高效。
核心结论
- 使用达芬奇单孔机器人系统进行标准机器人哈巴狗夹的体内装载、打开和放置是可行且高效的。
研究设计与样本
- 技术分享类文章,基于两例部分肾切除术中获取的视频,使用达芬奇单孔手术系统、Scanlan Reliance 25 mm弯曲机器人哈巴狗夹和Final Cut Pro视频编辑软件。
主要发现
- 标准机器人哈巴狗夹经改造以适应多孔机器人系统,但单孔机械臂因器械路径角度导致最大力低于装载和打开夹所需。
- 文中展示了使用单孔系统进行肾门阻断和后续夹移除的控制性步骤序列。
临床怎么用
- 在单孔机器人部分肾切除术中,可参考本文演示的步骤序列安全高效地放置和移除标准哈巴狗夹,减少初学者试错。
局限性
- 证据基于两例手术的视频资料,非系统研究,未报告样本量、成功率或统计学比较。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Renal hilar clamping during a robotic partial nephrectomy is essential for a successful, nephron-sparing tumor resection. Standard robotic bulldog clamps are modified to accommodate multiport robotic surgical systems with alterations including knobs to fit robotic forceps and reduced spring tension. Despite the lower spring tension, a standard robotic bulldog clamp cannot be loaded and opened using one Single Port robotic arm. The angled path instrument cables take through multiple robotic arm articulations lowers the maximum force generated by Single Port instruments to below what is necessary to load and open a robotic bulldog clamp. OBJECTIVE: To share institutional knowledge gained regarding safe and efficient hilar clamp placement and removal using the da Vinci Single Port Surgical System with a standard robotic bulldog clamp. As the da Vinci Single Port system gains wider use, knowledge of this technique will reduce surgeon trial-and-error during first experiences using the Single Port system with bulldog clamps. MATERIAL: Video obtained during two partial nephrectomies, da Vinci Single Port Surgical System, Scanlan Reliance 25 mm curved robotic bulldog clamp, Final Cut Pro video editing software. RESULTS: A controlled sequence of steps for renal hilar clamping and subsequent clamp removal using a standard robotic bulldog clamp with the da Vinci Single Port Surgical System. CONCLUSION: The demonstrated technique for intracorporeal loading, opening, and placement of a standard robotic bulldog clamp using the da Vinci Single Port Surgical System is feasible and efficient.
引用
- Shumaker L, Rais-Bahrami S, Nix J, et al. Renal Hilar Clamping With a Standard Robotic Bulldog Clamp Using the Single Port da Vinci Robot. Urology. 2020. PMID: 32795497. doi:10.1016/j.urology.2020.07.046
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。