识别链条中断环节:输尿管软镜在医院循环中损坏位置的分析
本研究对2017年9月至2018年6月间使用的输尿管软镜在术前和术后进行漏气测试,共收集281例完整数据。总体漏气失败率为7.8%(22/281),其中15例为术前失败(非手术损坏),仅7例发生在手术中。术中变量无显著关联,但较大结石负荷和高功率显示轻度关联。
要点
- 输尿管软镜漏气失败率为7.8%,多数损坏发生在手术室外。
- 术前漏气测试失败15例,提示运输、处理或灭菌环节损坏。
- 术中损坏7例,较大结石负荷和高功率有轻度关联。
核心结论
- 本机构输尿管软镜漏气失败率为8%(22/281),多数失败并非发生在手术中。
研究设计与样本
- 前瞻性观察性研究,2017年9月1日至2018年6月30日期间对输尿管软镜使用前后进行漏气测试,共收集281例完整数据。
主要发现
- 22例输尿管软镜漏气测试失败,总体失败率7.8%。
- 15例术前漏气测试失败,提示非手术损坏发生于运输、处理或无菌处理过程中。
- 7例失败发生在手术过程中,术中变量无显著关联,较大结石负荷和高功率显示轻度关联。
临床怎么用
- 应关注手术室外环节(运输、处理、灭菌)以减少输尿管软镜损坏。
- 对较大结石负荷和高功率手术使用软镜时需警惕术中损坏风险。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVE: To assess the timing and variables associated with damage to flexible ureteroscopes (fURS) at our institution. Flexible ureteroscopy is an important modality in the treatment of benign and malignant conditions of the upper urinary tract. While the durability and versatility of fURS have improved considerably, repair costs remain high and time out of commission diminishes workflow. After purchasing new digital fURS, we studied how and when these instruments were being damaged. MATERIALS AND METHODS: Between September 1, 2017 and June 30, 2018, we performed leak testing on fURS both before and after use. We gathered intraoperative data related to the user, the surgical indication, and the associated tools used in all cases that employed a digital or fiber optic fURS. Categorical and continuous variables were analyzed to identify risk factors for intraoperative fURS damage. RESULTS: During the study period, complete data was gathered for 281 cases. Twenty-two fURS failed leak testing indicating an overall leak failure rate of 7.8%. Of these, 15 failed leak testing preoperatively indicating nonoperative damage occurring sometime during transport, handling, or sterile processing. The other 7 failures occurred during the procedures. No intraoperative variables were significantly associated with failures. CONCLUSION: Our institutional leak failure rate is 8% (22/281). The majority of these failures did not occur during surgery. Of the 7 that occurred during surgery, larger stone burden and higher wattage showed mild association. Ongoing evaluation will target minimizing fURS damage outside of the operating room.
引用
- Higgins MM, Dugan A, Bell JR, et al. Identifying a Break in the Chain: An Analysis of Where Ureteroscope Damage Occurs in the Hospital Cycle. Urology. 2020. PMID: 32277991. doi:10.1016/j.urology.2020.02.030
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。