骶神经调节电极取出时断裂率、危险因素及结局:17年单中心回顾
该研究回顾2002至2018年464例带倒刺骶神经调节电极取出,断裂率7.5%(35例)。男性(OR 6.58)和糖尿病(OR 2.42)与断裂风险升高相关,植入至取出时间较短为保护因素(OR 0.86)。断裂多位于倒刺区(66%)且多为幽灵电极(81%),80%患者后续接受再植入,仅1例因碎片相关疼痛需再次手术。
要点
- 电极取出时断裂率为7.5%,高于厂商估计。
- 男性与糖尿病是断裂的危险因素,较短植入时间具保护作用。
- 断裂多位于倒刺区,长期并发症少见,多数患者可再植入。
核心结论
- 带倒刺骶神经调节电极取出时断裂率低(7.5%),但高于厂商估计。
研究设计与样本
- 2002至2018年单中心回顾性研究,纳入464例带倒刺电极取出,分析临床与手术因素及碎片结局。
主要发现
- 35例发生电极断裂,断裂率7.5%。
- 男性(OR 6.58,95%CI 2.54-17.01,p=0.001)和糖尿病(OR 2.42,95%CI 1.05-5.58,p=0.019)与断裂风险升高相关。
- 植入至取出时间较短为保护因素(OR 0.86,95%CI 0.76-0.98,p=0.002)。
- 35例断裂中27例有影像资料,66%(18例)断裂位于倒刺区,81%(22例)为幽灵电极。
- 80%(28/35)留有碎片患者接受再次植入,仅1例因碎片相关疼痛需再次手术。
临床怎么用
- 取出带倒刺电极时需警惕断裂风险,尤其男性及糖尿病患者。
- 多数断裂碎片为幽灵电极且长期并发症少见,可考虑再次植入。
局限性
- 单中心回顾性设计,可能存在选择偏倚与记录不全。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
PURPOSE: One risk of removal of a tined sacral neuromodulation lead is breakage, resulting in a retained lead fragment. We report lead breakage rates, risk factors and outcomes of retained fragments. MATERIALS AND METHODS: We report on 464 tined lead removals from 2002 to 2018. Clinical and surgical factors were reviewed and appropriate statistical analysis performed. Retained fragments were reviewed for fragment description and long-term complications. RESULTS: A total of 464 tined lead removals were included, with 35 lead breaks identified, for a rate of 7.5% (35). Factors associated with breakage included male gender (OR 6.58, 95% CI 2.54-17.01, p=0.001) and diabetes (OR 2.42, 95% CI 1.05-5.58, p=0.019). A shorter time since implantation was protective (OR 0.86, 95% CI 0.76-0.98, p=0.002). There was no difference in breakage rate based on age, weight, history of prior revision, bilateral leads, reason for lead removal, surgical technique or surgeon. Of the 35 breaks 27 had imaging available. Of these 66% (18) broke in the tined region and 81% (22) were ghost leads. Most patients (80%, 28 of 35) with retained lead fragments received another implant. Only 1 patient had a related complication of additional surgery for pain related to the fragment. CONCLUSIONS: There is a low rate of tined lead breakage during lead removal. However, it is higher than the manufacturer estimate. Protective factors include a shorter time between implant and lead revision. The most common location for lead breakage is in the region of the tines, most are ghost fragments and long-term complications are uncommon.
引用
- Rueb JJ, Pizarro-Berdichevsky J, Goldman HB, et al. 17-Year Single Center Retrospective Review of Rate, Risk Factors and Outcomes of Lead Breakage during Sacral Neuromodulation Lead Removal. J Urol. 2020. PMID: 31909689. doi:10.1097/JU.0000000000000740
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。