工作通道位置影响软性输尿管镜碎石效果:体外研究
该体外研究比较了工作通道位于9点与3点位置的两种软性输尿管镜(Flex-X2S与Flex-XC)在模拟肾盏模型中的碎石效果。结果显示不同盏位下两种通道位置各有优势,提示规划软性输尿管镜手术时应考虑工作通道位置,尤其处理下盏时。
要点
- 右后盏和左前盏:3点通道的碎石命中率与模拟激光烧灼比例高于9点通道。
- 左后盏和右前盏:9点通道的碎石命中率与模拟激光烧灼比例高于3点通道。
- 规划软性输尿管镜时应考虑工作通道位置,尤其下盏。
核心结论
- 软性输尿管镜工作通道位置影响激光碎石效果,规划手术时应考虑,尤其处理下盏时。
研究设计与样本
- 体外研究,使用K-Box模型的8个腔隙模拟左右肾盏分布,比较Flex-X2S(9点通道)与Flex-XC(3点通道)的碎石效果。
主要发现
- 右后盏和左前盏中,3点通道的STL均值±标准差分别为87%±15%和78%±35%,高于9点通道的46%±38%和43%±24%(P<0.05)。
- 左后盏和右前盏中,9点通道的STL均值±标准差分别为84%±20%和79%±30%,高于3点通道的65%±28%和44%±35%(P≤0.02)。
- 中下/下盏右后表面,3点通道的PBL为72%和77%,高于9点通道的31%和35%(P<0.01)。
- 下盏左后和右外侧面,9点通道的PBL为59%和50%,高于3点通道的34%和21%(P≤0.04)。
临床怎么用
- 规划软性输尿管镜碎石时,可根据目标肾盏位置选择工作通道方位更有利的输尿管镜。
- 处理下盏结石时,应特别关注工作通道位置对激光覆盖范围的影响。
局限性
- 研究为体外模型实验,结果需临床验证。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVE: To evaluate whether the position of the working channel affects the effectiveness of flexible ureteroscopy. MATERIALS AND METHODS: We compared the ureteroscopes Flex-X2S and Flex-XC (working channel at the 9 and 3 o'clock positions, respectively) in eight cavities of a K-Box model, simulating the distribution of the right and left intrarenal calyces. In the first and second settings, each cavity contained a 1-cm stone, lying on the bottom and fixed on the anterior sheet covering the box, respectively. In the third setting, the posterior and lateral surface of each cavity was draped with graph paper. Once the flexible ureteroscope entered each cavity, we measured and compared (i) the proportion of stone surface targeted by the laser (STL) and (ii) the proportion of graph paper burned by the laser (PBL) obtained with the two instruments. RESULTS: Higher STLs and PBLs were obtained with the 3 o'clock position than the 9 o'clock position in the right posterior and left anterior cavities (mean ± sd STL 87% ± 15% vs 46% ± 38% and 78% ± 35% vs 43% ± 24%, respectively; P < 0.05), and on the right posterior surfaces of the middle-lower/lower cavities (PBL 72% vs 31% and 77% vs 35%, respectively; P < 0.01) and on the left lateral surface of the lower cavities (PBL 45% vs 25%; P = 0048), respectively. Conversely, the 9 o'clock position provided higher STLs and PBLs than the 3 o'clock position in the left posterior and right anterior cavities (mean ± sd STL 84% ± 20% vs 65% ± 28% and 79% ± 30% vs 44% ± 35%, respectively; P ≤ 0.02), and on the left posterior and right lateral surfaces of the lower cavities (PBL 59% vs 34% and 50% vs 21%, respectively; P ≤ 0.04). CONCLUSIONS: The position of the working channel of the flexible ureteroscope should be considered when planning flexible ureteroscopy, especially when dealing with the lower pole.
引用
- Villa L, Ventimiglia E, Proietti S, et al. Does working channel position influence the effectiveness of flexible ureteroscopy? Results from an in vitro study. BJU Int. 2019. PMID: 31610080. doi:10.1111/bju.14923
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。