文献速递前列腺增生与排尿功能障碍男科与性医学

机器人辅助后入路膀胱颈植入AMS-800人工尿道括约肌治疗女性固有括约肌缺陷

Robot-assisted implantation of an artificial urinary sphincter, the AMS-800, via a posterior approach to the bladder neck in women with intrinsic sphincter deficiency.
PubMed 文献 · BJU Int · 2019-09-11

中位年龄64岁,术前中位尿垫重量300 g/24 h,中位手术时间244 min,无围手术期阴道或膀胱损伤。中位随访12个月时所有装置功能正常,5例无需防护(62.5%),3例需日间防护(37.5%),除1例(12.5%)因持续急迫性尿失禁要求治疗外均满意。结论认为该技术简单、可重复且安全,短期功能结果令人满意,与开放手术相当。

要点

  • 后入路机器人辅助AMS-800膀胱颈植入可避免盲性分离及阴道/膀胱损伤。
  • 8例女性患者中位随访12个月,所有装置功能正常,无围手术期阴道或膀胱损伤。
  • 5例(62.5%)无需防护,3例(37.5%)需日间防护,1例(12.5%)因持续急迫性尿失禁要求治疗。

核心结论

  • 机器人辅助后入路膀胱颈AMS-800植入在女性中是简单、可重复且安全的手术,短期功能结果满意并与开放手术相当。
  • 需要更长期比较该后入路与前入路机器人及经典开放技术的疗效和装置寿命以确认其优势。

研究设计与样本

  • 所有植入均采用后入路膀胱颈以避开盲性分离及阴道和/或膀胱损伤风险,装置于植入后5周激活,随访时间为3、6、12个月后每年一次。

主要发现

  • 术前中位尿垫重量为300 g/24 h,中位手术时间为244 min。
  • 未观察到围手术期阴道和/或膀胱损伤。
  • 中位随访12个月时所有AMS均功能正常;5例(62.5%)无需防护,3例(37.5%)需日间防护。
  • 除1例(12.5%)因持续急迫性尿失禁要求治疗外,其余患者均表示满意。

临床怎么用

  • 对于女性固有括约肌缺陷患者,可考虑机器人辅助后入路膀胱颈AMS-800植入作为避免盲性分离及阴道/膀胱损伤的术式选择。
  • 术后5周激活装置,并按3、6、12个月及每年进行随访评估装置功能与控尿情况。

局限性

  • 样本量小(8例),随访时间短(中位12个月),需更长期比较后入路与前入路机器人及开放技术的疗效和装置寿命。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVES: To describe a new technique for robot-assisted AMS-800 artificial urinary sphincter (AUS) bladder neck implantation in women. PATIENTS AND METHODS: We reviewed the medical files of patients who underwent robot-assisted AUS implantation between March 2017 and November 2018 at our centre. All of the implantations were performed using a posterior approach to the bladder neck in order to avoid blind dissection and the risk of vaginal and/or bladder injury. This strategy was viewed as an alternative to the anterior robot-assisted implantation recently described. The AUSs were activated 5 weeks after implantation. Patients were followed up at 3, 6 and 12 months, then annually. RESULTS: Eight patients, with a median age of 64 years, underwent robot-assisted AUS implantation via a posterior approach to the bladder neck. The median preoperative pad weight was 300 g/24 h. The median operating time was 244 min. No peri-operative vaginal and or bladder injuries were observed. At a median of 12 months of follow-up, all the AUSs were functional. Five patients required no protection (62.5%), three had day protection (37.5%), and all said they were satisfied except for one patient (12.5%) who requested treatment for persistent urge incontinence. CONCLUSION: Robot-assisted AUS implantation in women via a posterior approach to the bladder neck is a procedure that is simple, reproducible and safe. The short-term functional results are satisfactory and comparable to those obtained via an open approach. A more long-term comparison of the efficacy and longevity of AUSs implanted using this posterior approach is needed to confirm its benefit compared with the anterior robot-assisted approach and the classic open technique.

来源:BJU Int,查看原文。版权归原作者所有。