离体荧光共聚焦显微镜:首次应用于前列腺组织实时病理检查
研究纳入13例临床局限性前列腺癌患者的89份新鲜前列腺组织标本,比较离体荧光共聚焦显微镜(FCM)与组织病理诊断的一致性。FCM与病理诊断总体符合率为91%(κ=0.75),曲线下面积0.884(95%CI 0.840-0.920),敏感度83.33%、特异度93.53%。
要点
- FCM与组织病理诊断总体符合率91%,κ=0.75,曲线下面积0.884(95%CI 0.840-0.920)。
- FCM敏感度83.33%、特异度93.53%。
- FCM每份标本成像时间小于5分钟,作者认为可能用于术中病理标本分析。
核心结论
- FCM似乎是一种有前景的工具,可提升标本报告性能,鉴于其操作简单且显微图像生成非常迅速(每份标本小于5分钟)。
- 作者认为该技术可能用于术中病理标本分析。
研究设计与样本
- 研究纳入13例临床局限性前列腺癌患者的89份标本,患者均行机器人辅助腹腔镜根治性前列腺切除术,术中结束时用18-G活检针取新鲜前列腺组织。
- 标本随机分配给三位合作病理学家评估,采用Cohen κ检验评估观察者内和观察者间一致性,并通过受试者工作特征曲线分析评估诊断性能。
主要发现
- FCM与组织病理诊断的总体诊断一致性为实质性,正确诊断率91%(κ=0.75)。
- 曲线下面积为0.884(95%CI 0.840-0.920),敏感度83.33%,特异度93.53%。
关键数值与推荐
- FCM与组织病理诊断总体符合率91%,κ=0.75。
- 曲线下面积0.884(95%CI 0.840-0.920)。
- 敏感度83.33%,特异度93.53%。
- 每份标本成像时间小于5分钟。
临床怎么用
- 作者认为FCM似乎是一种有前景的工具,可提升标本报告性能。
- 作者认为该技术可能用于术中病理标本分析。
局限性
- 原文未报告局限性
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVE: To report the first application of ex vivo fluorescence confocal microscopy (FCM) - a novel optical technology that is capable of providing fast microscopic imaging of unfixed tissue specimens- in the urological field assessing its diagnostic accuracy for non neoplastic and cancerous prostate tissue (prostatic adenocarcinoma) compared to the 'gold standard' histopathological diagnoses. PATIENTS AND METHODS: In all, 89 specimens from 13 patients with clinically localised prostate cancer were enrolled into the study. All patients underwent robot-assisted laparoscopic radical prostatectomy with fresh prostatic tissue biopsies taken at the end of each intervention using an 18-G biopsy punch. Specimens were randomly assigned to the three collaborating pathologists for evaluation. Intra- and inter-observer agreement was tested by the means of Cohen's κ. The diagnostic performance was evaluated on receiver operating characteristic curve analysis. RESULTS: The overall diagnostic agreement between FCM and histopathological diagnoses was substantial with a 91% correct diagnosis (κ = 0.75) and an area under the curve of 0.884 (95% confidence interval 0.840-0.920), 83.33% sensitivity, and 93.53% specificity. CONCLUSION: FCM seems to be a promising tool for enhanced specimens' reporting performance, given its simple application and very rapid microscopic image generation (<5 min/specimen). This technique may potentially be used for intraoperative pathological specimens' analysis.