人肾结石自体荧光光谱用于术中结石-组织-器械分析的实验评估

Experimental Evaluation of Human Kidney Stone Spectra for Intraoperative Stone-Tissue-Instrument Analysis Using Autofluorescence.
PubMed 文献 · J Urol · 2019-01-01

研究评估了82例人结石(覆盖8种亚型)与猪尿路组织及内镜部件的荧光光谱差异,发现所有结石均有显著荧光信号,最弱结石信号比最强猪肾组织信号大3.6倍,内镜部件无荧光信号,荧光幅度与光谱曲线形式与结石类型无关。

要点

  • 所有天然结石均显示显著荧光信号,内镜部件无荧光信号。
  • 结石荧光信号幅度变化达75倍,最弱结石信号仍比最强猪肾组织信号大3.6倍。
  • 荧光幅度和光谱曲线形式与结石类型无关。

核心结论

  • 荧光光谱可用于区分所有相关结石类型与尿路组织及内镜部件。

研究设计与样本

  • 实验研究,收集82例天然人结石覆盖8种最相关亚型,与内镜部件、猪尿路组织及纯结石成分样本的荧光光谱进行比较。

主要发现

  • 所有天然结石相比猪尿路组织或内镜部件均显示显著荧光信号。
  • 荧光信号幅度变化达75倍。
  • 最弱结石信号为3.6倍于最强猪肾组织信号(均值±标准差 0.038±0.043 vs 0.00058±0.00058 任意单位)。
  • 内镜部件未观察到荧光信号。
  • 荧光幅度和光谱曲线形式与结石类型无关。

临床怎么用

  • 为未来智能激光系统实现术中实时目标区分和激光反馈控制提供光谱学依据。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

PURPOSE: The precision and safety of laser lithotripsy in the upper urinary tract is solely controlled by the surgeon. To our knowledge laser systems with integrated real-time analysis of target tissues are not available. An intelligent laser system with automated target differentiation and laser feedback control would provide tremendous improvement in patient safety and surgical precision. We evaluated the technical, medical and physical conditions for real-time analysis using fluorescence for future development of a new laser for lithotripsy. MATERIALS AND METHODS: We collected data on the fluorescence spectra of 82 native human calculi covering the 8 most relevant subtypes compared to the spectra of endoscope components, the organic porcine urinary tract and pure samples of urinary stone compositions. Data analysis was performed to determine differences in sample signal intensities and emission wavelengths. RESULTS: All native stones showed a significant fluorescence signal compared to porcine urinary tract tissue or endoscope components. The amplitude of the fluorescence signal varied by a factor of 75. The weakest signal of stone material was 3.6-fold larger than the strongest signal of pig kidney tissue (mean ± SD 0.038 ± 0.043 vs 0.00058 ± 0.00058 arbitrary units). No fluorescence signal was observed for endoscope components. Fluorescence amplitude and spectral curve form were found to be unrelated to stone type. CONCLUSIONS: Our study provides essential information on the spectral differentiation of tissue, urinary stones and relevant endoscope components. The measurements indicate that differentiation by fluorescence is possible for all relevant stone types.

来源:J Urol,查看原文。版权归原作者所有。

引用

  1. Schütz J, Miernik A, Brandenburg A, et al. Experimental Evaluation of Human Kidney Stone Spectra for Intraoperative Stone-Tissue-Instrument Analysis Using Autofluorescence. J Urol. 2019. PMID: 30077558. doi:10.1016/j.juro.2018.07.067

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。