多触点无线供能神经刺激系统植入大鼠胫神经长期刺激的可行性
本研究在10只成年Sprague-Dawley大鼠中植入多触点无线供能神经刺激系统,评估胫神经激活特性及长期稳定性。通道1在T0时运动阈值最低(均值0.58 ± 0.10 mA),术后首次测试(2周)时运动抽动强度明显增加(75.8 ± 30.5%,通道1)。10只大鼠中8只完成3个月研究,结论认为在临床前模型中用多触点植入装置实现胫神经刺激具有长期可行性,未来需进一步研究评估该无线供能系统治疗患者下尿路症状的效果。
要点
- 10只大鼠中8只完成3个月研究,多触点植入装置可实现胫神经长期刺激。
- 通道1在T0时运动阈值最低,均值0.58 ± 0.10 mA。
- 术后2周首次测试时运动抽动强度增加75.8 ± 30.5%(通道1)。
核心结论
- 在临床前模型中,多触点植入装置实现胫神经刺激具有长期可行性。
- 未来需进一步研究评估该无线供能系统治疗患者下尿路症状的效果。
研究设计与样本
- 研究使用10只成年Sprague-Dawley大鼠,手术将植入式脉冲发生器置于下背部区域。
- 带4个活性电极的导线与胫神经平行放置,通过拇趾或足趾运动及足部肌电图确认胫神经激活阈值。
- 胫神经激活阈值每两周评估一次,持续12周。
主要发现
- 通道1在T0时运动阈值最低,均值0.58 ± 0.10 mA。
- 术后首次测试(2周)时运动抽动强度明显增加,通道1为75.8 ± 30.5%。
- 10只大鼠中8只成功完成3个月研究。
局限性
- 未来需进一步研究评估该无线供能系统治疗患者下尿路症状的效果。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVE: Implant-driven tibial nerve stimulation therapy is an effective technique for treating overactive bladder. However, the monopolar lead design in the currently available implantable devices pose long-term therapeutic challenges in terms of efficiently and selectively delivering electrical pulses to the target. Hence, the purpose of this study was to (1) characterize the tibial nerve (TN) activation properties using a multi-contact implantable system and (2) evaluate the long-term stability of using such a neural interface in a preclinical model. MATERIALS AND METHODS: Ten adult Sprague-Dawley rats were used in this study. An implantable pulse generator was surgically inserted in the lower back region. The lead wire with 4 active electrodes was placed in parallel with the TN. The threshold for activating the TN was confirmed via movement of the hallux or toes as well as the foot EMG. The TN activation threshold was assessed biweekly, over a period of 12 weeks. RESULTS: Channel 1 exhibited the lowest motor threshold at T0 (mean = 0.58 ± 0.10 mA). A notable increase in motor twitch intensity was observed during the first test session (2 weeks) following surgical implantation (75.8 ± 30.5%, channel 1). Among the 10 rats tested, 8 rats successfully completed the 3-month study. CONCLUSION: Results from this study demonstrate the long-term feasibility of achieving tibial nerve stimulation with a multi-contact implantable device in a preclinical model. Future studies are warranted to assess the effects of using such a wirelessly powered system for treating lower urinary tract symptoms in patients.
引用
- Moazzam Z, Paquette J, Duke AR, et al. Feasibility of Long-term Tibial Nerve Stimulation Using a Multi-contact and Wirelessly Powered Neurostimulation System Implanted in Rats. Urology. 2016. PMID: 27865751. doi:10.1016/j.urology.2016.11.013
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。